The 90Y microsphere dose that disagrees with the treatment plan
A 67-year-old man with unresectable hepatocellular carcinoma (single 6.2 cm lesion, Child-Pugh A) is treated with resin 90Y microspheres via lobar infusion. The dosimetry team used a partition model on a pre-treatment 99mTc-MAA SPECT/CT and predicted a tumor absorbed dose of 142 Gy and a normal-liver dose of 28 Gy. At the post-treatment 90Y PET/CT, the measured tumor dose is 61 Gy and the measured normal-liver dose is 71 Gy. The activity administered was 1.4 GBq. The lung shunt fraction was 4%. The referring oncologist is asking whether to plan a second treatment in 6 weeks. The physicist is asked to explain the discrepancy before any decision is made.
1. What are the most likely technical and biological causes of the tumor/normal-liver dose discrepancy, and how would you systematically investigate it?
Read the worked answer
The most important point is that 99mTc-MAA is a surrogate, not the therapy itself. Systematically check:
- Catheter position and flow. MAA particles are ~10-90 um and distribute by first-pass capillary trapping. If the microcatheter was more proximal at MAA time than at treatment (or vice versa), the lobar distribution differs. Compare the angiographic position and the extrahepatic/segmental distribution on both scans.
- Arteriovenous shunting. A 4% lung shunt is acceptable, but a higher tumor-to-normal shunt at treatment can divert activity away from tumor and into normal parenchyma. Recompute the lung shunt from the post-treatment PET.
- Reconstruction and quantification differences. MAA SPECT and 90Y PET have very different resolution, scatter, and partial-volume behavior. A 6.2 cm lesion is large enough that partial volume is modest, but the normal-liver VOI near the tumor will be contaminated by spill-in. Confirm the same VOI definitions, same reconstruction (e.g., TOF-OSEM with resolution recovery), and same segmentation method on both datasets.
- Assumed versus measured activity. Verify the actual calibrated activity in the vial and the residual in the delivery system; a 1.4 GBq nominal dose can differ from delivered dose.
- Partition model assumptions. The partition model requires a tumor-to-normal ratio (T/N). If the T/N was estimated from a region that included necrotic or hypovascular tumor, the tumor dose is overestimated and the normal-liver dose underestimated. This is consistent with the observed direction of the error (predicted tumor high, predicted normal low).
- Tumor biology. Necrotic core, prior locoregional therapy, or altered arterial supply can make MAA uptake unrepresentative.
In practice, the MAA-based prediction is the weaker dataset; the post-treatment PET is the ground truth for what was delivered. The discrepancy itself is the finding to act on.
2. Given the post-treatment measurement, what is your recommendation regarding a second treatment, and what would you require before one is delivered?
Read the worked answer
The post-treatment PET shows a normal-liver dose of 71 Gy, which is well above the commonly cited safety threshold for lobar resin microsphere therapy (roughly 40-50 Gy to the normal liver in a single session, and lower for whole-liver or compromised reserve). The tumor received only 61 Gy, which is below the typical tumoricidal range for HCC. This is the opposite of the intended plan and is a safety signal.
Recommendation: do not commit to a second lobar treatment in 6 weeks on the current data. Before any second treatment:
- Reassess liver function (bilirubin, albumin, INR, Child-Pugh, MELD) and imaging for radiation-induced liver disease or biliary injury.
- Recompute the cumulative normal-liver dose, not just the second-session dose. Cumulative normal-liver tolerance is the binding constraint.
- Repeat dosimetry with a better predictive dataset. Options include a repeat MAA with careful catheter matching, or a scout 90Y PET-based approach, and explicit measurement of the T/N ratio with necrotic core excluded from the tumor VOI.
- Consider a segmental rather than lobar approach if the tumor is confined to one segment, to spare normal parenchyma.
- Document the discrepancy and the rationale in the chart and discuss it with the multidisciplinary team.
The judgment being tested: the post-treatment scan is not just a QA artifact; it changes the risk-benefit calculation for the next fraction. The correct answer is to slow down, not to repeat the same plan with a dose adjustment based on a model that has already been shown to be wrong in this patient.