Spill of F-18 FDG in the Hot Lab
At 07:40, during preparation for a PET list, a technologist drops a shielded syringe containing approximately 370 MBq [10 mCi] of F-18 FDG onto the hot-lab floor. The syringe barrel cracks and liquid spreads over roughly a 30 cm × 30 cm area near the dose-calibrator. The technologist's gloved hand and forearm are contaminated; she is wearing a lab coat but no sleeve protection. Two other staff are in the room. The department's survey meter is in the PET suite being used for a patient. The next patient is due at 08:00. The hot lab has a floor drain and a sink.
1. What are your immediate priorities in the first minutes, and in what order?
Read the worked answer
Priority order (people → containment → contamination control → recovery):
- Stop and secure: announce the spill, stop work, and have everyone leave the immediate spill area without spreading contamination. Do not let anyone walk through it.
- Decontaminate the person first: the technologist's hand/forearm - remove contaminated gloves and lab coat carefully (inside-out), wash skin with lukewarm water and mild soap, and survey (get a meter from the PET suite or use a backup). Repeat washing until survey shows acceptable levels; do not abrade skin.
- Contain the spill: if safe, cover the liquid with absorbent pads/paper towels from the edges inward to prevent spreading; mark the area and restrict access. Do not use the floor drain or sink - F-18 must not be flushed; it would contaminate the drainage system.
- Control access: post signs, keep the hot lab closed, and divert the 08:00 patient (notify the scheduler/physician).
- Notify the RSO and document time, activity, location, and personnel involved.
Key point: Person decontamination and preventing spread come before cleaning the floor, and the drain/sink are off-limits.
2. How do you decontaminate the floor and verify it is clean, given the survey meter is initially unavailable?
Read the worked answer
Decontamination:
- Don double gloves, shoe covers, and a gown; work from the outside of the spill inward.
- Use absorbent pads to blot (don't wipe/spread), then clean with detergent and water in small sections, changing pads/gloves frequently and placing all waste in a radioactive waste bag.
- Do not use the drain or sink. Collect all liquid and cleaning materials as radioactive waste.
- F-18 has a short half-life (110 min), so contamination decays quickly, but you still decontaminate and verify rather than wait.
Verification without the primary meter:
- Retrieve the survey meter from the PET suite as soon as possible, or use a backup/calibrated meter. You cannot declare the area clean without a survey.
- Survey with the probe at a fixed distance (e.g., 1 cm) over a grid of the area; compare to background and to the department's contamination action level (e.g., a defined multiple of background or a fixed Bq/cm² criterion).
- Re-survey after each cleaning cycle until readings are at/below the action level; record results and the final clearance.
Do not release the room based on visual inspection or elapsed time alone.
3. What are the dose/contamination implications for the technologist's skin, and what follow-up and reporting are required?
Read the worked answer
Skin dose/contamination implications:
- F-18 emits positrons (511 keV annihilation photons); external dose from contamination is significant, and internalization (through broken skin, mouth, or inhalation) is the main concern.
- Skin contamination is assessed by survey and, if needed, skin dose estimation. The skin dose from a small contaminated area is generally managed by prompt decontamination; the priority is to prevent internal contamination and to confirm the skin is clean by survey.
- If contamination is not removable, or if internalization is suspected, escalate for bioassay/internal dose assessment and medical evaluation.
Follow-up and reporting:
- Document the incident: activity, location, personnel, decontamination steps, survey results, and any residual contamination.
- Notify the RSO and complete the required incident/occurrence report per the facility's regulatory obligations; some frameworks require reporting of spills above defined activity thresholds or any personnel contamination.
- Review the event (root cause: unshielded handling, no secondary containment, missing meter) and implement corrective actions - e.g., secondary containment trays, a dedicated hot-lab meter, and spill drills.