The Trembling Tracing: Artifact vs. Real Rhythm
You are called to a med-surg floor to run a 12-lead on a 67-year-old man with a new cough and shortness of breath. He is sitting upright, shivering under two blankets, and has a fine tremor in both hands. He has an IV in his right forearm with a pump running, and he is on 2 L nasal cannula. The first tracing comes out with a wavy, irregular baseline in leads I and aVL; the limb leads show an undulating sawtooth-like pattern at about 300 per minute that disappears when you look at leads V1–V6. The nurse says, "That looks like atrial flutter - should we call a code?" The patient's radial pulse is 88, regular, and he is alert.
1. How do you decide whether the sawtooth pattern is atrial flutter or artifact, and what do you check first?
Read the worked answer
Check the pattern's distribution and behavior.
- The sawtooth appears only in the limb leads and vanishes in the precordial leads. A true atrial arrhythmia is a cardiac electrical signal and will show up wherever the atria are "seen" - it does not conveniently disappear when you change lead groups.
- Rate and regularity: the undulation is ~300/min, but the radial pulse is 88 and regular. That mismatch (no corresponding ventricular response) argues against a genuine flutter with conduction.
- Patient factors: shivering plus a fine hand tremor is a classic source of somatic tremor artifact, which is most prominent in limb leads because the electrodes sit on moving arms.
First checks: confirm electrode contact and placement, ask the patient to relax his arms and place them flat and still, and warm him if possible. Then repeat the tracing. If the pattern disappears when he stops shivering, it was artifact.
2. List at least four concrete steps to reduce artifact on the repeat tracing, and explain why each helps.
Read the worked answer
- Warm the patient / add a blanket - shivering is involuntary muscle activity; warming reduces it at the source.
- Reposition arms flat and supported - resting the forearms on the bed reduces tremor transmission to limb electrodes.
- Prepare skin properly: clip hair if needed, gently abrade, use fresh electrodes - high skin impedance amplifies noise; good contact improves signal-to-noise.
- Move the IV pump/limb leads away from the running pump and route cables away from power cords - 60 Hz electrical interference from equipment produces a thick, fuzzy baseline; separation and cable routing reduce it.
- Ensure the patient is not talking or coughing during the tracing - ask him to breathe normally and stay still for the ~10 seconds of acquisition.
- Verify limb lead placement (arm vs. leg) - reversed or misplaced limb electrodes create patterns that mimic pathology; correct placement prevents false findings.
3. The nurse wants to call a code. What do you say, and what do you document?
Read the worked answer
What to say: "His pulse is 88 and regular and he's alert - this looks like tremor artifact in the limb leads, not flutter. Let me repeat the tracing with him warm and still before we escalate. If the pattern persists in the precordial leads or his rhythm changes, we call the provider immediately."
Be respectful but clear: you are not diagnosing, you are pointing to objective findings (regular pulse, pattern limited to limb leads) that argue for repeating the study first. Escalation is still appropriate if the repeat is abnormal or the patient deteriorates.
Document: patient condition (shivering, tremor, IV pump), the artifact observed and which leads, the interventions taken (warming, repositioning, skin prep, repeat tracing), the repeat findings, and that the provider was notified of the final tracing. Keep both tracings labeled - the comparison is the evidence.