ASCP MB: Quality Assurance, Quality Control, Safety, and Laboratory Operations

ASCP MB · Work through your exam, one topic at a time.

All guide sections

Quality Assurance, Quality Control, Safety, and Laboratory Operations

Quality control

Each run needs positive and negative controls, a no-template control, and an internal control to monitor extraction and amplification inhibition (endogenous genes such as RNase P, or exogenous armored RNA). Results are charted on Levey-Jennings plots and judged by Westgard rules: 1-3s and 2-2s are rejection rules, R-4s signals random error, and 4-1s or 10x indicate systematic shift. Investigate shifts, trends, and drift rather than repeating until acceptable.

Validation and verification

Validation establishes performance characteristics for a laboratory-developed test; verification confirms the manufacturer's claims before clinical use. Parameters include accuracy, precision, analytical sensitivity and specificity, reportable range, reference interval, and limit of detection. Proficiency testing, alternative performance assessment, competency documentation, calibration, temperature monitoring, and turnaround-time monitoring complete the quality program. Under CLIA, laboratory-developed tests require documented validation.

Contamination control

Use unidirectional workflow: extraction, then amplification setup, then detection, in separate areas with dedicated equipment and aliquoted reagents. Amplicon contamination is prevented with dUTP/UNG, fresh bleach (about 10%), and UV where appropriate.

Safety

Most molecular work is BSL-2 with standard precautions and PPE; culture of Mycobacterium tuberculosis and certain agents requires BSL-3. Know chemical hazards: phenol causes burns, guanidine thiocyanate plus bleach releases toxic gas, ethidium bromide is a mutagen, and formaldehyde is a carcinogen. Use SDS, spill kits, and compliant waste streams.

Traps

Common errors include believing 1-3s is only a warning, swapping validation and verification, and carrying amplified product into the clean extraction area.

Worked example

A single control exceeding ±3 SD triggers the 1-3s rule: reject the run, investigate, and repeat before reporting.