ASCP SH: Hematology Laboratory Operations, Quality, and Instrumentation

ASCP SH · Work through your exam, one topic at a time.

All guide sections

Hematology Laboratory Operations, Quality, and Instrumentation

Laboratory operations, quality, and instrumentation

Quality in hematology spans preanalytical, analytical, and postanalytical phases. Preanalytical errors include wrong anticoagulant, underfilling citrate tubes, EDTA-induced platelet clumping, hemolysis, lipemia, and delayed processing. Correct specimen handling-fill lines, mixing, stability, and rejection criteria-prevents most errors. Analytical quality uses calibration, reagent validation, and statistical QC. Levey-Jennings charts and Westgard rules detect random and systematic error: 1:2s warning, 1:3s, 2:2s, R:4s, 4:1s, and 10:x rejection rules. Proficiency testing, linearity, reference interval verification, delta checks, and critical value reporting are required by accreditation programs. Automated analyzers use impedance, light scatter, flow cytometry, and fluorescence; WBC counts must be corrected for nucleated red cells, and platelet counts for clumping or large platelets. Coagulation analyzers use mechanical, turbidimetric, or chromogenic methods. Point-of-care testing requires the same QC rigor.

Traps: 1:2s is a warning, not rejection. R:4s indicates within-run random error; 2:2s across runs suggests systematic error. A normal result does not exclude a preanalytical problem. Carryover and dead volume affect counts. Hemolysis falsely raises potassium and LDH but not always hemoglobin.

Worked example: two QC levels differ by 4.5 SD in the same run-apply the R:4s rule and reject the run.