HPTC RT Exam: Treatment Planning and Simulation

HPTC RT Exam · Work through your exam, one topic at a time.

All guide sections

Treatment Planning and Simulation

Treatment Planning and Simulation tests the chain from CT scan to an approved, deliverable plan: immobilization and indexing, image acquisition, volume definition, dose calculation, and plan evaluation against DVH criteria.

Simulation and image acquisition

  • Reproducibility starts with immobilization indexed to the couch (wing board, vac-lok, thermoplastic mask) and reference marks placed at the intended isocenter.
  • CT simulation: thin slices (2-3 mm routine, 1 mm for intracranial SRS), scan extent about 5 cm beyond the PTV, arms positioned out of the beam path, and a CT-to-density table validated with a phantom.
  • 4DCT or breath-hold scanning for thoracic and upper-abdominal targets; the motion envelope from all phases defines the internal target volume.

Volumes and margins (ICRU 62/83)

  • GTV is visible tumor; CTV adds microscopic spread; ITV adds internal motion; PTV adds setup and machine uncertainty.
  • Margin recipes such as van Herk (2.5 sigma + 0.7 sigma) convert measured systematic and random errors into a PTV margin. A default 5 mm is not a recipe.

Dose calculation

  • Type A pencil beam fails in heterogeneous tissue: it overestimates dose beyond lung and underestimates inside lung because it does not model lateral scatter.
  • Type B (superposition/convolution, e.g. AAA) and Type C (Acuros, Monte Carlo) handle density heterogeneity better.
  • Use a calculation grid of 3 mm or less for small fields, and confirm whether the algorithm reports dose to medium or dose to water.

Worked example

A 10 x 20 cm rectangular field needs an equivalent square for depth-dose lookup: 4A/P = 4(200)/(60) = 13.3 cm. Use the PDD for the 13.3 cm square, not for the 20 cm side length. When evaluating a 60 Gy prescription, report D95 (must reach at least 57 Gy) and the near-maximum dose (D2 or D0.03 cc) for serial structures; a plan that looks uniform on a color wash can still violate these numbers.

Traps

  • Placing respiratory motion in the CTV instead of the ITV.
  • Using the physical field size for a heavily MLC-blocked field.
  • Accepting a DVH that looks acceptable while ignoring deliverability (leaf travel, jaw settings, gantry-couch collision).
  • Forgetting that a plan without a signed approval is not treatable.