CRT (CREST): Air Travel Planning With a Pulse-Dose Concentrator

CRT · Put your knowledge to work in context.

All case studies

Air Travel Planning With a Pulse-Dose Concentrator

A 68-year-old woman with COPD (FEV1 38% predicted, FVC 62%) is seen in clinic before a planned trip. She lives at sea level and uses continuous-flow oxygen at 2 L/min by nasal cannula around the clock. Her resting SpO2 is 92% and she desaturates to 87% during a 30-meter walk on that flow. Her itinerary is a 6-hour nonstop flight, a 2-hour layover, and 5 days at a destination 1,200 meters above sea level. Her durable medical equipment company has supplied an FAA-approved portable oxygen concentrator that delivers pulse-dose oxygen only. The manufacturer's label states that setting 3 is equivalent to 3 L/min continuous flow, and that each of her two batteries lasts 4 hours at setting 2 and 2.5 hours at setting 3. The clinic note suggests she simply turn the device to setting 3 during the flight and otherwise keep her usual routine.

1. Evaluate whether the proposed plan provides adequate oxygen for this itinerary. Identify each point of failure.

Read the worked answer

Several failures exist.

First, pulse-dose delivery relies on a pressure trigger. At altitude the patient will breathe faster and shallower, and the concentrator may fail to trigger or may deliver a smaller bolus as respiratory rate rises. Manufacturer claims of equivalence to continuous flow are bolus estimates, and they do not hold for a patient who already desaturates with exertion.

Second, cabin altitude on commercial aircraft is roughly 2,400 meters (8,000 feet). Inspired oxygen partial pressure falls from about 150 mmHg to around 118 mmHg, and arterial oxygen in a healthy traveler falls to roughly 55 to 60 mmHg. A patient with an FEV1 of 38% and exertional desaturation will fall further.

Third, battery duration is inadequate. At setting 3 she has 5 hours of power for a flight of 6 hours plus 2 hours of ground and layover time, and most airlines require battery capacity for 150% of flight duration. Setting 2 gives 8 hours, which is still marginal.

Fourth, the destination at 1,200 meters adds additional altitude exposure not addressed in the note, and the plan says nothing about supply during the layover.

2. What test should be performed before the flight, and how should the result be interpreted and applied?

Read the worked answer

Order a hypoxia altitude simulation test, in which the patient breathes 15% oxygen for about 20 minutes to simulate a cabin altitude of 8,000 feet, with SpO2 monitoring and arterial blood gas if indicated. A reasonable alternative when the test is unavailable is a walk test with oximetry on her current flow, but the altitude simulation test is preferred because it reproduces the hypoxic challenge.

Interpretation: an SpO2 falling below 85%, or a PaO2 below 50 mmHg, indicates that in-flight oxygen is required. The test is also used to titrate the in-flight flow to keep SpO2 at or above 90%.

Apply the result as a prescription: for the flight, increase continuous flow above her baseline, titrated to the test. A common starting estimate is baseline plus 2 L/min, or roughly 1 L/min added per 1,000 feet of altitude gain, but the simulation test supersedes any rule of thumb. Repeat the calculation for the destination altitude of 1,200 meters. Document the numbers in the chart so the airline and equipment supplier can see the rationale.

3. If the pulse-dose concentrator cannot meet her needs, what equipment and prescription changes do you recommend?

Read the worked answer

Switch to a continuous-flow source. Compressed gas cylinders or liquid oxygen can deliver reliable flow regardless of respiratory pattern. Liquid oxygen is efficient for higher flows over long periods, and cylinders can be used as backup.

Calculate the requirement rather than guessing. At 3 L/min for 8 hours of travel and ground time, she needs 1,440 liters, and an E cylinder holds about 680 liters, so multiple cylinders or a larger M cylinder (about 3,000 liters) is required. Add a reserve so total capacity covers 150% of the flight time, plus the layover and ground transport.

Request medical clearance through the airline, provide a physician's letter stating the FAA-approved device, flow rate, and battery capacity, and confirm the supplier at the destination can provide the same flow. Advise against relying on a pulse-dose device in flight without a documented altitude simulation test, and arrange backup power and a backup oxygen source.