Failed Bowie–Dick test on a Monday morning, plus a handpiece and a finger lesion

NZ Dental Therapy REX · Put your knowledge to work in context.

All case studies

Failed Bowie–Dick test on a Monday morning, plus a handpiece and a finger lesion

It is Monday, 8:15 am in a two-chair school dental clinic. The dental assistant runs the daily Bowie–Dick test on the class B vacuum autoclave before the first patient. The indicator card shows a pale patch in the centre - an incomplete colour change. The steriliser log shows that on Friday three loads were run, each with a passing post-sterilisation Bowie–Dick test; the pouched instruments from those loads went into the clean-store drawer and several were used on Friday afternoon. The autoclave was validated in November, five months ago, and last had a routine service in April. The first patient, a 7-year-old, is already in the waiting room with her mother. In the next surgery you notice a colleague wiping a used high-speed handpiece with an alcohol-impregnated wipe and returning it to the bracket. That colleague also has a red, weeping lesion on the side of her right index finger, covered with a plain adhesive dressing.

1. Explain what the failed Bowie–Dick test detects, what you must do before any instrument is used on this morning's patients, and whether Friday's instruments need to be quarantined or patients recalled.

Read the worked answer

The Bowie–Dick test is a dynamic air-removal test. It tells you whether the vacuum stage of a class B cycle has removed air adequately so that steam can penetrate the load. It does not measure sterility, and a pass does not guarantee sterility of a particular pack - it is a check on the process.

A failed test means you cannot have confidence that a cycle run now would sterilise anything, so the autoclave comes out of service immediately: tag it, stop processing, check the obvious causes (door seal, water quality and level, filter, drain screen, correct cycle selected, adequate warm-up), then repeat the test. If it fails again, escalate to your sterilising service or engineer and document. In the meantime use a backup autoclave or single-use items and rebook non-urgent care. The one thing you must not do is run 'just one more load'.

Friday is a different question. Those loads were released against a documented passing Bowie–Dick test, so a Monday failure does not automatically invalidate them, and a blanket recall of Friday's children would be an over-reaction with its own harms. However, the failure raises the possibility of an intermittent or developing fault. Quarantine the unused Friday packs pending advice, review Friday's records and the chemical indicators on the remaining packs, and refer the decision to your infection control lead or the dentist responsible. Any look-back should be a documented, risk-assessed decision, not a chairside guess.

Tell the waiting mother honestly that there is a delay, and why.

2. How would you address the colleague's handpiece handling and the lesion on her finger?

Read the worked answer

Handpiece. A high-speed handpiece is a semicritical item: it contacts oral tissues, and its internal lumen retracts blood, saliva and debris during use. An alcohol wipe is a low-level surface disinfectant and does nothing for the internal channels. The correct process is cleaning (including flushing and cleaning the internal lines), lubrication where the manufacturer specifies it, pouching, and steam sterilisation between every patient, with the handpiece run to dry and stored dry. Have a direct, non-punitive conversation with the colleague, reference the standard rather than the person, and if the practice continues, escalate to the clinical leader - this is a patient safety issue, not a private preference.

The finger lesion. A red, weeping lesion on the side of an index finger in a dental clinician is a herpetic whitlow until proven otherwise, and it is often occupationally acquired from a patient's lesion. A plain adhesive dressing is not adequate: it is not waterproof, not occlusive, and does not stay intact through hand hygiene and gloving. She should be assessed by occupational health or her GP, and she should not provide direct patient care while the lesion is weeping - for her own protection (auto-inoculation, secondary bacterial infection) and for patients'. She also needs to know that herpes can be transmitted to patients' mouths and to colleagues.

Document the advice given, check her hepatitis B and other immunisation status, and report the incident if the lesion is likely to be occupationally acquired.