ABMP Medical Health Physics flashcards
ABMP Medical Health Physics · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Half-life
- Time required for the activity of a radionuclide to decrease to half its initial value; T½ = ln2/λ, where λ is the decay constant.
- Becquerel
- SI unit of radioactivity equal to one disintegration per second; 1 Ci = 3.7×10^10 Bq.
- Photoelectric effect
- Photon interaction in which the photon transfers all its energy to an inner-shell electron, which is ejected; dominant at low energies and high atomic number.
- Geiger-Mueller counter
- Gas-filled detector operating in the Geiger region; highly sensitive but cannot discriminate photon energy and has significant dead time.
- Ionization chamber
- Gas-filled detector operating in the ionization region; used as the reference instrument for radiation protection surveys and dose rate measurements.
- HPGe detector
- High-purity germanium semiconductor detector providing excellent energy resolution for gamma spectroscopy; requires cryogenic cooling.
- Absorbed dose
- Energy deposited by ionizing radiation per unit mass of material; SI unit is the gray (Gy), equal to 1 J/kg.
- Equivalent dose
- Absorbed dose averaged over a tissue or organ multiplied by the radiation weighting factor w_R; SI unit is the sievert (Sv).
- Effective dose
- Sum over tissues of equivalent dose multiplied by tissue weighting factor w_T; used to estimate stochastic risk; unit is the sievert.
- Kerma
- Kinetic energy released per unit mass in a medium by indirectly ionizing radiation; air kerma is the basis of many calibration protocols.
- Stochastic effect
- Radiation effect, such as cancer, for which probability increases with dose and for which no threshold is assumed.
- Deterministic effect
- Tissue reaction that has a threshold dose and whose severity increases with dose, such as erythema or cataracts.
- Linear-quadratic model
- Cell survival model S = e^(−αD−βD²); α/β ratio distinguishes early-responding (~10 Gy) from late-responding (~3 Gy) tissues.
- ALARA
- Principle of keeping radiation doses as low as reasonably achievable, considering economic and social factors, through time, distance, and shielding.
- Dose limit
- Regulatory upper bound on dose to workers or the public; ICRP occupational limit is 20 mSv/y averaged over 5 years.
- Medical event
- Administration of byproduct material differing from the prescribed dose by 20% or more and exceeding 0.5 Sv to an organ or 50 mSv effective dose.
- Bremsstrahlung
- Continuous X-ray spectrum produced when electrons decelerate in the anode target; maximum photon energy equals the applied kVp.
- Half-value layer
- Thickness of material that reduces photon intensity to half its original value; HVL = ln2/μ, typically 2.5–4 mm Al in diagnostic beams.
- CTDIvol
- Volume computed tomography dose index, a standardized measure of CT radiation output per unit volume; DLP equals CTDIvol times scan length.
- Tc-99m
- Technetium-99m, a 6-hour half-life radionuclide emitting 140 keV gamma rays; the most widely used diagnostic radiopharmaceutical.
- PET annihilation
- Positron emission tomography process in which a positron annihilates with an electron, producing two 511 keV photons emitted approximately 180 degrees apart.
- Dose calibrator
- Re-entrant ionization chamber used to measure the activity of radiopharmaceuticals before administration; geometry and volume affect readings.
- Percentage depth dose
- Ratio of absorbed dose at a given depth to the maximum dose along the central axis of a radiation beam, expressed as a percentage.
- Brachytherapy TG-43
- Dosimetry formalism for brachytherapy sources that calculates dose to water using radial dose and anisotropy functions.
- Primary barrier
- Shielding designed to attenuate the primary radiation beam; its thickness depends on workload, use factor, occupancy, and distance.
- Occupancy factor
- Factor T used in shielding calculations reflecting the fraction of time an area is occupied; full 1, partial 1/4, occasional 1/16.
- Tenth-value layer
- Thickness of material that reduces photon intensity to one tenth; number of TVLs equals log10 of the inverse transmission.
- Personal dosimeter
- Device such as a TLD, OSL, or film badge worn to measure occupational dose; ring badges measure extremity dose.
- Investigation level
- Dose threshold, typically 10%, 30%, or 100% of the regulatory limit, that triggers review, investigation, or reporting.
- Public dose limit
- Regulatory limit on dose to members of the public from licensed activities, generally 1 mSv per year.
- Secular equilibrium
- Condition in a decay chain where the parent half-life greatly exceeds the daughter half-life, so daughter activity approaches parent activity.
- Radiation weighting factor
- Factor w_R that adjusts absorbed dose for the biological effectiveness of the radiation type; 1 for photons, 20 for alpha particles.