ABMP MRI Physics study materials
ABMP MRI Physics study resources

ABMP MRI Physics cheat sheet

ABMP MRI Physics · Keep the key ideas within reach.

Quick reference

Basic MR Physics and Nuclear Magnetism

  • Nuclear spin arises when a nucleus has an odd number of protons or neutrons; 1H (single proton) has spin 1/2 and is the dominant imaging nucleus.

  • Gyromagnetic ratio γ relates Larmor frequency to field: ω0 = γB0; for 1H, γ/2π ≈ 42.58 MHz/T.

  • Net magnetization M0 grows as spins populate the lower energy state; Boltzmann excess is small (~ppm at clinical fields).

  • Equilibrium M0 is proportional to proton density and to B0 (Curie behavior, M0 ∝ B0/T at thermal equilibrium).

  • Bloch equation: dM/dt = γM × B − (Mx i + My j)/T2 − (Mz − M0)k/T1.

  • A 90° pulse rotates M0 into the transverse plane; flip angle θ = ∫γB1(t)dt.

  • Resonance condition: RF at ω0 transfers energy; off-resonance reduces the effective flip angle.

  • T2* combines reversible field inhomogeneity with true T2: 1/T2* = 1/T2 + 1/T2′.

Quick reference

Relaxation, Contrast Mechanisms, and Pulse Sequence Timing

  • T1 recovery: Mz(t) = M0(1 − e^(−t/T1)); after a 90° pulse, Mz returns toward M0 with time constant T1.

  • T2 decay: Mxy(t) = Mxy(0)e^(−t/T2); T2 is always ≤ T1 for tissue.

  • Signal equation (gradient echo, flip α): S ∝ ρ sinα (1 − e^(−TR/T1)) e^(−TE/T2*) / (1 − cosα e^(−TR/T1)).

  • Spin-echo signal: S ∝ ρ (1 − e^(−TR/T1)) e^(−TE/T2); long TR + short TE = PD weighting, long TR + long TE = T2 weighting.

  • Short TR and short TE favor T1 weighting; TR controls T1 contrast, TE controls T2 contrast.

  • Inversion recovery: Mz(t) = M0(1 − 2e^(−TI/T1)); null point when TI = T1 ln2 (≈0.693 T1).

  • STIR nulls fat (TI ≈ 0.69 × T1_fat, ~150 ms at 1.5T); FLAIR uses a long TI to null CSF.

  • Ernst angle (gradient echo) maximizes signal at α = arccos(e^(−TR/T1)).

Quick reference

Spatial Encoding, Gradients, and Image Formation

  • Slice selection: RF bandwidth Δf and slice-select gradient Gss set slice thickness Δz = Δf/(γGss).

  • Frequency encoding applies a readout gradient so position maps linearly to Larmor frequency along x.

  • Phase encoding applies a brief gradient whose area sets ky; number of phase-encoding steps sets matrix size in y.

  • k-space: FOV = 1/Δk and resolution = 1/(k_max); readout sampling interval Δt with gradient G gives Δk = γGΔt.

  • Gradient amplitude determines maximum spatial frequency (resolution); gradient duration/area determines ky spacing.

  • Two-dimensional Fourier reconstruction maps k-space samples to image via inverse 2D FT.

  • Slice profile is the FT of the RF envelope; sinc RF gives a rectangular (idealized) slice profile.

  • Chemical shift in pixels = Δf / (bandwidth per pixel); fat–water offset ≈ 3.5 ppm.

Quick reference

Radiofrequency Coils, Hardware, and Signal Reception

  • Coil SNR improves with proximity to tissue; surface coils give high SNR near the coil but limited field of view.

  • Quadrature (circularly polarized) coils improve SNR by √2 versus linear coils and reduce transmit power needs.

  • Body/volume coils provide uniform B1 for transmit; phased-array receive coils combine element signals for high SNR over large FOV.

  • Coil noise dominance: when sample noise exceeds coil noise, SNR scales with B0; at low fields coil noise may dominate.

  • B1 field homogeneity affects flip-angle uniformity; B1 mapping and parallel transmit address high-field inhomogeneity.

  • Receive bandwidth: wider bandwidth lowers SNR (more noise) but reduces chemical-shift artifact.

  • Gradient hardware limits slew rate (dB/dt) and maximum amplitude; peripheral nerve stimulation thresholds relate to dB/dt.

  • RF power deposition (SAR) scales approximately with B0² and with flip angle squared for a given sequence.

Quick reference

Image Reconstruction, Sampling, and Artifacts

  • Nyquist criterion: sampling interval must satisfy Δt ≤ 1/(bandwidth) to avoid aliasing; undersampling k-space produces wrap.

  • Aliasing (wrap) occurs when FOV is smaller than the object; remedies include larger FOV, oversampling, or saturation bands.

  • Truncation (Gibbs) artifact arises from finite k-space sampling; manifests as ringing near sharp edges.

  • Motion artifact appears as ghosting along the phase-encode direction because motion modulates ky lines.

  • Chemical-shift artifact shifts fat relative to water along the frequency-encode direction.

  • Susceptibility artifact produces signal loss and geometric distortion near metal/air-tissue interfaces, worse with gradient echo.

  • Zero-filling interpolates but does not add true resolution; it reduces Gibbs ringing appearance.

  • Partial Fourier and parallel imaging reduce scan time at the cost of SNR and potential unfolding artifacts.

Quick reference

Fast, Parallel, and Advanced Acquisition Techniques

  • Echo-planar imaging (EPI) fills k-space rapidly after a single excitation; susceptible to geometric distortion and T2* blurring.

  • Parallel imaging (SENSE/GRAPPA) uses coil sensitivity to skip phase-encode lines; acceleration reduces SNR by roughly √R.

  • SENSE unfolds in image space using coil sensitivity maps; GRAPPA reconstructs missing lines in k-space.

  • Turbo/fast spin echo uses multiple 180° refocusing pulses per TR to fill multiple k-space lines.

  • Balanced SSFP gives high SNR with T2/T1 contrast; sensitive to off-resonance banding.

  • Compressed sensing exploits sparsity and incoherent sampling to reconstruct from undersampled data.

  • Simultaneous multislice (SMS) excites multiple slices at once and separates them with coil sensitivity encoding.

  • Non-Cartesian trajectories (radial, spiral) offer motion robustness and efficient sampling but require gridding reconstruction.

Quick reference

Safety, Bioeffects, and RF Heating

  • Specific absorption rate (SAR) = deposited RF power per unit mass (W/kg); regulatory limits exist for whole-body and localized exposure.

  • SAR scales approximately with B0², flip angle², and duty cycle; reducing flip angle or TR increases safety margin.

  • RF heating risk increases near conductive implants and leads; long resonant leads can concentrate induced currents.

  • Static field limits: the main field attracts ferromagnetic objects; projectile risk is the primary hazard.

  • Gradient-induced electric fields can cause peripheral nerve stimulation; dB/dt limits are sequence-dependent.

  • Acoustic noise arises from gradient coil vibration; hearing protection is required.

  • Pregnancy and implanted-device status must be verified; device MR-conditional status specifies allowed field and SAR conditions.

Quick reference

Quality Assurance, Acceptance Testing, and Artifact Troubleshooting

  • Daily QA typically checks SNR, geometric accuracy, slice thickness, and image uniformity with a phantom.

  • SNR can be measured as mean signal divided by standard deviation of background noise (with correction for magnitude bias).

  • Geometric accuracy is verified by comparing known phantom distances to measured distances.

  • Ghosting/phase stability is assessed by measuring signal in a background region across phase-encode direction.

  • Coil element failure shows as signal dropout or nonuniformity in a specific region; compare element images.

  • Center frequency and transmit gain drift indicate hardware or coil tuning issues.

  • Artifact troubleshooting: identify direction (frequency vs phase encode), then match to sampling, motion, or hardware cause.

  • Acceptance testing establishes baseline values against manufacturer or protocol specifications before clinical use.

Quick reference

Quantitative MRI and Emerging Methods

  • T1 mapping: variable flip angle or inversion-recovery methods fit signal to recovery models; B1 correction is required.

  • T2 mapping uses multi-echo spin echo with exponential decay fitting; T2* mapping uses multi-echo gradient echo.

  • ADC maps from diffusion-weighted imaging quantify water mobility; high cellularity lowers ADC.

  • DWI signal: S = S0 e^(−b·ADC); b-value depends on gradient amplitude, duration, and diffusion time.

  • MR elastography derives tissue stiffness from shear-wave propagation measured with phase-contrast sequences.

  • Quantitative susceptibility mapping (QSM) estimates tissue susceptibility from phase data after dipole inversion.

  • Perfusion (DCE/DSC) quantifies flow and volume from tracer kinetics; requires arterial input function modeling.

  • Machine-learning reconstruction and synthetic contrasts are emerging; validation against reference standards is required.

Quick reference

Basic MR Physics and Nuclear Magnetism

  • Larmor frequency: f0 = (γ/2π)B0; at 1.5 T, 1H precesses at ~63.9 MHz; at 3 T, ~127.7 MHz.

  • Spin angular momentum and magnetic moment are related by μ = γS; only nuclei with net spin show MR signal.

  • RF excitation tips M0 by angle θ; a 180° pulse inverts Mz to −M0.

  • Free induction decay (FID) is the signal after a single RF pulse as Mxy dephases.

  • Bloch equation describes precession and relaxation of the macroscopic magnetization vector.

  • T1 and T2 are intrinsic tissue time constants; T2* includes additional dephasing from field inhomogeneity.

  • Boltzmann distribution: the small population difference between spin states produces the measurable net magnetization.

Quick reference

Relaxation, Contrast Mechanisms, and Pulse Sequence Timing

  • T1 is spin-lattice (longitudinal) recovery; T2 is spin-spin (transverse) decay.

  • Signal in spin echo: S = k·ρ·(1 − e^(−TR/T1))·e^(−TE/T2).

  • Inversion recovery nulls a tissue when TI = T1·ln2; STIR nulls fat, FLAIR nulls CSF.

  • Ernst angle α = arccos(e^(−TR/T1)) maximizes gradient-echo signal for given TR and T1.

  • Long TR reduces T1 weighting; long TE increases T2 weighting.

  • Proton density weighting uses long TR and short TE to minimize T1 and T2 effects.

  • Contrast agents shorten T1 (and T2*) to increase signal on T1-weighted images.

Quick reference

Spatial Encoding, Gradients, and Image Formation

  • Slice thickness Δz = Δf/(γGss); thinner slices need larger gradient or narrower RF bandwidth.

  • Frequency encoding maps position to frequency via a readout gradient; phase encoding uses gradient area to set ky.

  • FOV = 1/Δk; resolution = 1/(k_max); both depend on gradient amplitude and sampling.

  • k-space center holds low-frequency contrast; periphery holds high-frequency edge detail.

  • Inverse 2D Fourier transform reconstructs the image from k-space samples.

  • Slice profile equals the Fourier transform of the RF pulse envelope.

  • Chemical shift (pixels) = Δf/(bandwidth per pixel); fat–water separation ~3.5 ppm.

Quick reference

Radiofrequency Coils, Hardware, and Signal Reception

  • Surface coils give high SNR near tissue but nonuniform sensitivity; phased arrays combine elements for large FOV.

  • Quadrature coils improve SNR by √2 relative to linear coils.

  • Body coils transmit uniform B1; local receive coils improve SNR for reception.

  • Receive bandwidth affects SNR: wider bandwidth admits more noise but reduces chemical-shift artifact.

  • B1 inhomogeneity worsens with field strength; B1 mapping and parallel transmit mitigate it.

  • SAR scales approximately with B0² and flip angle²; gradient slew rate limits dB/dt.

  • Coil noise vs sample noise determines whether SNR scales with B0 in a given regime.

Quick reference

Image Reconstruction, Sampling, and Artifacts

  • Aliasing/wrap occurs when FOV is smaller than the object; oversampling or saturation bands prevent it.

  • Gibbs (truncation) ringing results from finite k-space sampling near sharp edges.

  • Motion ghosts appear along the phase-encode direction because ky lines are modulated inconsistently.

  • Chemical-shift artifact displaces fat along the frequency-encode direction.

  • Susceptibility artifact causes signal loss and distortion near metal and air–tissue interfaces.

  • Zero-filling interpolates but does not improve true resolution.

  • Parallel imaging and partial Fourier reduce scan time at the cost of SNR and possible unfolding artifacts.

Quick reference

Fast, Parallel, and Advanced Acquisition Techniques

  • EPI fills k-space after one excitation; prone to geometric distortion and T2* blurring.

  • Parallel imaging (SENSE/GRAPPA) uses coil sensitivities to skip phase-encode lines; SNR falls roughly with √R.

  • Turbo spin echo uses multiple 180° refocusing pulses per TR to acquire multiple lines.

  • Balanced SSFP gives T2/T1 contrast with high SNR; sensitive to off-resonance banding.

  • Compressed sensing exploits sparsity and incoherent sampling to reconstruct undersampled data.

  • Simultaneous multislice excites multiple slices and separates them using coil sensitivity encoding.

  • Non-Cartesian radial/spiral trajectories improve motion robustness but require gridding reconstruction.

Quick reference

Safety, Bioeffects, and RF Heating

  • SAR is RF power absorbed per unit mass (W/kg); limits differ for whole-body and localized exposure.

  • SAR scales approximately with B0², flip angle², and duty cycle.

  • Conductive implants and leads can concentrate RF currents, increasing local heating risk.

  • The static field attracts ferromagnetic objects; projectile hazard is the primary safety concern.

  • Gradient-induced electric fields can cause peripheral nerve stimulation; dB/dt limits apply.

  • Acoustic noise from gradient vibration requires hearing protection.

  • Device MR-conditional status specifies allowed field strength and SAR conditions.

Quick reference

Quality Assurance, Acceptance Testing, and Artifact Troubleshooting

  • Daily QA checks SNR, geometric accuracy, slice thickness, and uniformity with a phantom.

  • SNR is measured as mean signal divided by background noise standard deviation, with magnitude-bias correction.

  • Geometric accuracy compares known phantom distances with measured distances.

  • Ghosting is assessed by signal in a background region along the phase-encode direction.

  • Coil element failure appears as regional signal dropout or nonuniformity.

  • Center frequency and transmit gain drift indicate tuning or hardware issues.

  • Artifact troubleshooting starts with direction (frequency vs phase encode) to identify the cause.

Quick reference

Quantitative MRI and Emerging Methods

  • T1 mapping uses variable flip angle or inversion recovery with B1 correction.

  • T2 mapping uses multi-echo spin echo; T2* mapping uses multi-echo gradient echo.

  • DWI signal: S = S0·e^(−b·ADC); higher b-values emphasize restricted diffusion.

  • ADC maps quantify water mobility; high cellularity lowers ADC.

  • MR elastography measures tissue stiffness from shear-wave propagation.

  • QSM estimates tissue susceptibility from phase data after dipole inversion.

  • DCE/DSC perfusion quantifies flow and volume using tracer kinetics and an arterial input function.