ABMP MRI Physics: Image Reconstruction, Sampling, and Artifacts

ABMP MRI Physics · Work through your exam, one topic at a time.

All guide sections

Image Reconstruction, Sampling, and Artifacts

Core concepts

Reconstruction applies the inverse Fourier transform to k-space samples. The Nyquist criterion governs sampling: readout sampling must satisfy bandwidth, and phase encoding must cover enough k-space to avoid aliasing. Undersampling causes wraparound (aliasing).

Common artifacts and causes

  • Aliasing/wrap: FOV too small; fix by enlarging FOV or oversampling.
  • Chemical shift: fat–water frequency difference causes misregistration; increases with field strength; fix with fat saturation or higher bandwidth.
  • Ghosting: motion or phase-encode inconsistency; fix with flow compensation, gating.
  • Truncation (Gibbs): ringing from abrupt k-space edges.
  • Susceptibility: signal loss near metal/air; worse with GRE and long TE.
  • Zipper: RF interference; herringbone: vibration or gradient issues.

Traps

Questions often ask which direction an artifact appears. Aliasing occurs along the phase-encode direction unless oversampling is used; chemical shift occurs along the frequency-encode direction.

Rule of thumb

If an artifact moves when you change phase-encode direction, it is likely a sampling or motion artifact. If it is fixed to anatomy, suspect susceptibility or chemical shift.