Quick reference

Image Production and Evaluation

  • DSA subtraction requires a mask image acquired before contrast arrival; patient motion between mask and contrast frames produces misregistration artifact appearing as dark or light halos at vessel edges.

  • Pixel shift (remasking) is the post-processing correction for misregistration; it realigns the mask to the contrast image without re-exposing the patient.

  • Spatial resolution of a digital angiographic system is limited by pixel size at the image intensifier or flat-panel detector; larger field of view (FOV) spreads the same matrix over more area and reduces spatial resolution.

  • Contrast resolution is governed by bit depth: a 10-bit image displays 1024 gray levels, a 12-bit image 4096 gray levels.

  • Frame rate selection trades temporal resolution against dose: 30 frames/sec gives smoother motion depiction than 3 frames/sec but delivers proportionally more exposure per second.

  • Roadmapping overlays a previously acquired contrast-filled vessel image on live fluoroscopy to guide catheter and wire navigation.

  • Magnification modes improve spatial resolution by using a smaller input area of the detector but increase patient dose because the same output brightness requires higher exposure.

  • Quantum mottle appears as grainy noise when too few photons reach the detector; it is reduced by increasing exposure (mAs) or increasing detector dose, not by window/level adjustment.

  • Vessel stenosis quantification by DSA is performed by comparing the diameter of the narrowed segment with the adjacent normal reference diameter, expressed as a percentage reduction.

Quick reference

Patient Care and Education

  • Informed consent for angiography must document the procedure, its benefits, material risks (bleeding, infection, contrast reaction, vessel injury, radiation risk), and alternatives.

  • Contrast-induced nephropathy risk is increased by pre-existing renal insufficiency, diabetes mellitus, dehydration, and large contrast volume; pre-procedure serum creatinine and eGFR are the standard screening labs.

  • Patients with a documented prior anaphylactoid reaction to iodinated contrast require premedication with a corticosteroid (e.g., prednisone) plus an antihistamine, typically beginning 13 hours before contrast.

  • Metformin is commonly withheld after iodinated contrast in patients with renal impairment because of lactic acidosis risk; reinstitution follows renal function reassessment per institutional protocol.

  • Post-angiography access-site care includes bed rest, pressure dressing, and monitoring the puncture site for hematoma, bruit, or expanding mass; a new pulsatile mass suggests pseudoaneurysm.

  • Pulse checks (distal pulses, capillary refill, temperature, color, sensation) are documented before and after the procedure on the accessed extremity.

  • Renal protective measures include hydration with isotonic saline and using the lowest contrast volume consistent with diagnostic and therapeutic needs.

  • Patient education before the procedure includes fasting instructions, expected sensations (warmth or flushing with contrast injection), and the need to remain still during image acquisition.

Quick reference

Procedures

  • Seldinger technique: needle puncture, guide wire insertion, needle removal, catheter introduction over the wire; the wire must be secured at all times to prevent loss into the vessel.

  • Diagnostic catheter shapes are selected by vessel anatomy: a cobra or visceral curve for selective abdominal branches, a Simmons or reverse curve for vessels arising at sharp downward angles.

  • Angioplasty balloon diameter should match the diameter of the adjacent normal vessel; underexpansion leaves residual stenosis and oversizing risks dissection or rupture.

  • Stent placement requires adequate proximal and distal landing zones in relatively normal vessel; stents are deployed over a guide wire and typically post-dilated to ensure full expansion.

  • Embolization agents include coils (precise, permanent, proximal occlusion), gelatin sponge (temporary, intermediate vessel), and liquid agents such as NBCA or ethylene vinyl alcohol copolymer (distal penetration).

  • Thrombolysis for arterial occlusion delivers a fibrinolytic agent directly into the thrombus via a multi-side-hole infusion catheter, with fibrinogen and hemoglobin monitored during infusion.

  • IVC filter indications include contraindication to or failure of anticoagulation in a patient with documented deep vein thrombosis or pulmonary embolism.

  • Transjugular intrahepatic portosystemic shunt (TIPS) creates a channel between the hepatic vein and portal vein to decompress portal hypertension; it is used for variceal bleeding or refractory ascites.

  • Percutaneous nephrostomy access is obtained through a posterior calyx, typically at the 12th rib margin, to avoid pleural and vascular injury.

Quick reference

Radiation Protection and Equipment Operation

  • The inverse square law states that dose rate decreases with the square of the distance from the source; doubling the distance from the patient to the tube reduces dose to one quarter.

  • Fluoroscopic dose rate is proportional to the square of the tube potential (kVp); reducing kVp lowers dose but also lowers penetration and contrast.

  • Pulsed fluoroscopy reduces patient dose in proportion to the pulse rate: 15 pulses/sec delivers roughly half the dose of 30 pulses/sec for the same exposure per pulse.

  • Last-image-hold freezes the final fluoroscopic frame and eliminates additional exposure during review; it is a primary dose-reduction practice.

  • The image intensifier or flat-panel detector should be positioned as close to the patient as possible, and the x-ray tube as far from the patient as possible, to reduce skin dose.

  • Magnification modes increase dose because a smaller detector area is used, requiring higher exposure to maintain brightness; they should be used only when resolution is required.

  • Collimation to the region of interest reduces patient dose and improves image contrast by limiting scattered radiation reaching the detector.

  • Occupational protection: lead apron (0.5 mm Pb equivalent), thyroid shield, and dosimeter worn at collar level outside the apron; personnel should maximize distance and use shielding.

  • Air kerma (mGy) and kerma-area product (Gy·cm²) are the standard dose metrics displayed on modern interventional systems for dose tracking.