ARRT MRI flashcards
ARRT MRI · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Normal GFR threshold
- GFR below 30 mL/min/1.73 m² is severe renal impairment; group I gadolinium is contraindicated and group II agents are used only at lowest dose with caution.
- NSF risk agents
- Nephrogenic systemic fibrosis is associated with linear gadolinium agents in severe renal failure; macrocyclic group II agents have the lowest dissociation and risk.
- Metformin and gadolinium
- For gadolinium, metformin generally may be continued; withholding 48 hours applies mainly after iodinated contrast in eGFR below 30 or AKI.
- Extravasation first steps
- Stop injection, remove IV, elevate extremity, apply cold compress, document volume and site, and monitor for compartment syndrome.
- Anaphylactoid contrast reaction
- Stop injection, call for help, assess airway, give oxygen, and administer IM epinephrine for bronchospasm or hypotension.
- Pregnancy MRI contrast
- Iodinated and gadolinium contrast cross the placenta; MRI without contrast is preferred, and gadolinium is given only when essential.
- MRI metal screening
- Ask about pacemakers, aneurysm clips, cochlear implants, ocular metal, shrapnel, and tattoos; verify implant labeling before entering the magnet room.
- Sedation monitoring
- During sedation, continuously monitor SpO2 and HR, record BP every 5 minutes, and observe respiration and level of consciousness.
- T1 weighting parameters
- Short TR under about 800 ms and short TE under about 30 ms produce T1 contrast; fat is bright and fluid is dark.
- T2 weighting parameters
- Long TR over about 2000 ms and long TE over about 80 ms produce T2 contrast; fluid is bright and fat remains bright.
- STIR fluid sensitivity
- STIR uses inversion recovery with TI about 150 ms at 1.5 T to null fat, making fluid and edema bright.
- FLAIR CSF suppression
- FLAIR applies a long inversion time near 2000 ms to null CSF, improving detection of periventricular and cortical lesions.
- MRCP technique
- MRCP uses heavily T2-weighted thick-slab or 3D respiratory-triggered sequences so stationary fluid in bile and pancreatic ducts is bright.
- TOF MRA principle
- Time-of-flight MRA relies on flow-related enhancement of unsaturated blood entering the slice; no gadolinium is required.
- Chemical shift artifact
- Fat and water protons precess at different frequencies; at the fat-water interface, misregistration occurs along the frequency-encoding direction.
- Aliasing correction
- Wrap-around artifact occurs when anatomy exceeds FOV; increase FOV, use saturation bands, or apply oversampling along the phase-encoding direction.
- Larmor equation
- Precessional frequency equals gyromagnetic ratio times field strength; for hydrogen-1, 42.58 MHz per tesla, giving 63.9 MHz at 1.5 T.
- T1 relaxation time
- T1 is the time for longitudinal magnetization to recover to 63 percent of its final value after a 90-degree radiofrequency pulse.
- T2 relaxation time
- T2 is the time for transverse magnetization to decay to 37 percent of its initial value; T2 star adds static field inhomogeneity.
- Ernst angle formula
- The Ernst angle maximizes signal for a given TR and T1 and is calculated as cosine theta equals e to the minus TR over T1.
- k-space center
- Central k-space lines encode image contrast and SNR; peripheral lines encode spatial resolution and edge detail.
- SNR voxel relation
- Signal-to-noise ratio is proportional to voxel volume times the square root of excitations; larger voxels or more averages increase SNR.
- Receive bandwidth effect
- Increasing receive bandwidth lowers SNR but shortens minimum TE and reduces chemical shift and susceptibility artifact.
- SAR determinants
- Specific absorption rate rises with field strength squared, flip angle squared, and RF duty cycle; fast spin echo sequences deposit more energy.
- Inversion time nulling
- To null a tissue, set inversion time to its T1 times the natural logarithm of 2, approximately 0.693 times T1.
- Gradient slew rate
- Slew rate is the rate of gradient amplitude change; higher slew rates permit faster encoding but increase peripheral nerve stimulation risk.
- Five gauss line
- The 5 gauss line marks the boundary where the static field may affect implanted devices and loose ferromagnetic objects; it is posted and restricted.
- MRI zones
- Zone I is public, zone II is screened, zone III is restricted, and zone IV is the magnet room with the strongest field.
- RF burn prevention
- Avoid skin-to-skin contact loops, keep cables off the patient, and use padding; RF currents can cause burns even without heating the whole body.
- Acoustic noise protection
- MRI gradients produce noise that can exceed 99 decibels; earplugs or headphones are required for patients and staff in the magnet room.
- Cryogen quench
- A quench releases helium gas, displacing oxygen; evacuate the room, open doors, and do not reenter until oxygen levels are safe.
- MR conditional implant
- An MR conditional device is safe only under specified field strength, SAR, and coil conditions; verify the exact labeling before scanning.