ARRT Radiation Therapy study materials
ARRT Radiation Therapy study resources

ARRT Radiation Therapy cheat sheet

ARRT Radiation Therapy · Keep the key ideas within reach.

Quick reference

Patient Care and Education

  • Radiation dermatitis timeline: faint erythema at 10–15 fractions (~20–30 Gy), dry desquamation near 30–40 Gy, moist desquamation typically above 40 Gy once basal cells are depleted.

  • Alopecia begins 2–3 weeks after the first 2 Gy fraction; hair regrows 2–3 months after treatment ends, but scalp doses above ~45–50 Gy may cause permanent loss.

  • Oral mucositis appears 10–14 days after the start of radiation, peaks at 2–3 weeks, and resolves 2–4 weeks after completion unless dose exceeds mucosal tolerance.

  • Fatigue is the most common side effect of radiation therapy; it typically begins within the first 1–2 weeks and may persist 4–6 weeks or longer after treatment ends.

  • Informed consent must document diagnosis, treatment intent (curative vs palliative), technique, expected benefits, material risks, alternatives including no treatment, and the right to withdraw.

  • Two patient identifiers (name plus date of birth or medical record number) are verified before every simulation and treatment per National Patient Safety Goal requirements.

  • Weekly weights are recorded during treatment; loss greater than 5% of pre-treatment weight triggers dietitian referral and consideration of enteral supplementation.

  • Skin care rule: cleanse with lukewarm water and mild soap, pat dry, and avoid heat, sun, friction, adhesive tape, and deodorant on the treated field unless institutional protocol states otherwise.

Quick reference

Safety

  • NCRP Report 116 annual occupational limits: 50 mSv (5 rem) total effective dose, 150 mSv (15 rem) to the lens of the eye, 500 mSv (50 rem) to skin and extremities.

  • Public dose limit from a licensed facility is 1 mSv (100 mrem) per year and 0.02 mSv (2 mrem) in any single hour.

  • Inverse square law: I1/I2 = (d2/d1)²; doubling the distance from a point source reduces exposure to one quarter.

  • Half-value layer HVL = 0.693/μ, and transmitted intensity follows I = I0e^(−μx); ten half-value layers reduce intensity by a factor of about 1000.

  • Posting thresholds measured at 30 cm from the source: 'Caution: Radiation Area' above 0.05 mSv/h (5 mrem/h); 'High Radiation Area' above 1 mSv/h (100 mrem/h).

  • ALARA controls ranked as time, distance, and shielding; distance is the most practical and effective single control for staff exposure.

  • Linac safety interlocks: door interlock terminates beam on entry, and beam-off/beam-on indicators plus audio-visual alarms are checked daily before patient treatment.

  • A patient with a permanent brachytherapy implant may be released only if the calculated dose to any other person is unlikely to exceed 5 mSv (500 mrem) and written radiation-safety instructions are provided.

Quick reference

Treatment Planning

  • Volume chain: GTV (visible tumor) → CTV (GTV plus microscopic disease) → ITV (CTV plus internal motion) → PTV (CTV or ITV plus setup margin), with organs at risk contoured separately.

  • Van Herk margin recipe: PTV margin = 2.5Σ + 0.7σ, where Σ is the systematic error SD and σ the random error SD, giving at least 95% CTV coverage in 90% of patients.

  • QUANTEC central nervous system constraints: spinal cord maximum ≤50 Gy (45 Gy to full cord cross-section), brainstem ≤54 Gy, optic chiasm/nerves ≤50 Gy, lens ≤6–10 Gy.

  • Parotid constraint: mean dose below 26 Gy, or at least one gland below 20 Gy, reduces the incidence of grade 2 or worse xerostomia.

  • Thoracic constraints: lung V20 ≤30–35% and mean lung dose ≤20 Gy; heart V25 ≤10% and mean heart dose as low as reasonably achievable in breast planning.

  • Electron beam rule of thumb: practical range Rp ≈ E/2 cm in water and the 90% depth dose lies near E/3 to E/4 cm, so energy selection is based on target depth.

  • Equivalent square field size S = 2ab/(a+b); it is used to look up PDD, TMR, Sc, and Sp for rectangular fields.

  • ICRU-83 reporting: prescribe to the median dose (D50%), report near-maximum D2% and near-minimum D98%, and keep PTV coverage at 95% of prescription in 95% of the volume.

  • CT simulation practice: 2–3 mm slice thickness, reproducible immobilization, and 4DCT or respiratory gating when target motion exceeds about 1 cm.

Quick reference

Radiation Treatment Delivery

  • TG-142 output constancy tolerances: ±3% daily, ±2% monthly, and ±1% annual relative to the baseline calibration.

  • Daily mechanical checks: lasers and optical distance indicator within ±2 mm, collimator and gantry angle indicators within ±1 degree.

  • Winston-Lutz test verifies coincidence of radiation field center and mechanical isocenter to within 1 mm at multiple gantry, collimator, and couch positions for SRS/SBRT programs.

  • SBRT delivery requires image-guided localization within 1 mm, 6D couch correction before each fraction, and motion management (compression or gating) when excursion exceeds 5 mm.

  • VMAT delivers dose during continuous gantry rotation by simultaneously modulating multileaf collimator positions, dose rate, and gantry speed across one or more arcs.

  • Half-beam blocking places the collimated field edge at the machine midline so that adjacent fields (for example cranial and spinal fields) match without divergence overlap at depth.

  • Total body irradiation uses extended SSD (typically 300–400 cm), AP/PA parallel-opposed fields, 5–10 cGy/min dose rate to limit pulmonary toxicity, and lung compensators; 12 Gy in 6 fractions is a common regimen.

  • Deep inspiration breath hold for left breast irradiation increases heart-to-chest-wall distance and lowers mean cardiac dose compared with free breathing.

  • Orthogonal kV or CBCT verification tolerance is typically 2–3 mm for bony anatomy and about 2 mm for soft-tissue targets before treatment.

Quick reference

Dosimetry and Quality Assurance

  • TG-51 photon reference conditions: 10 cm depth in water, 100 cm SSD or SAD, 10 × 10 cm field, 0.6 cc Farmer chamber; Dw = M·N_Dw·P_ion·P_tp·P_elec·P_pol.

  • Isocentric MU calculation: MU = prescribed dose ÷ (calibrated output × TMR(d,r) × Sc(rc) × Sp(r) × wedge/tray factor × inverse-square correction).

  • Converting percent depth dose to TMR: TMR(d,r) = (%DD/100) × ((SSD+d)/(SSD+dmax))² × (Sp(dmax)/Sp(d)).

  • Gamma index combines dose difference and distance-to-agreement; 3%/3 mm with at least 95% of points passing is the standard IMRT QA threshold, and 2%/2 mm is often used for SBRT.

  • Isotope data: I-125 half-life 59.4 days with ~28 keV average photons; Pd-103 17 days at ~21 keV; Cs-131 9.7 days; Ir-192 73.8 days at ~380 keV average.

  • TG-43 brachytherapy formalism: D(r,θ) = Sk·Λ·[G(r,θ)/G(r0,θ0)]·g(r)·F(r,θ), with Λ the dose-rate constant, g the radial function, and F the anisotropy function.

  • In-vivo diode dosimetry action level is commonly ±5% between measured and calculated dose at the measurement point, prompting investigation when exceeded.

  • Chamber correction factors applied during calibration: P_ion for recombination, P_tp for temperature and pressure, P_pol for polarity, and P_elec for electrometer calibration.