Quick reference

Specimen Procurement, Handling, and Fixation

  • 10% neutral buffered formalin (NBF) contains 4% formaldehyde buffered to pH 7.0-7.4; it cross-links proteins and penetrates tissue at roughly 1 mm per hour at room temperature.

  • Minimum fixative-to-tissue volume ratio is 10:1; blocks are sliced 2-3 mm thick so fixative reaches the center within a practical time frame.

  • Routine NBF fixation is 6-24 hours for small biopsies and 24-48 hours for large specimens; fixation time is documented when IHC or molecular testing is planned.

  • Cold ischemia time runs from devascularization in the patient to immersion in fixative; prolonged warm ischemia degrades DNA, RNA, and phosphorylated epitopes.

  • Alcohol fixatives such as 95% ethanol are used for cytology smears and preserve glycogen, but they coagulate proteins and cause more shrinkage than formalin.

  • EDTA decalcification at neutral pH chelates calcium slowly and preserves antigenicity and nucleic acids better than strong acids such as nitric or hydrochloric acid.

  • Acid hematin (formalin pigment) forms in acidic or unbuffered formalin, especially in blood-rich tissue, and is removed with saturated alcoholic picric acid.

  • Mercury-based fixatives (B5, Zenker) require post-fixation in alcohol and pigment removal with iodine followed by sodium thiosulfate.

  • 10% NBF is markedly hypertonic relative to plasma; for electron microscopy use 2-3% glutaraldehyde in buffer at approximately 300-400 mOsm/kg.

Quick reference

Processing, Embedding, and Microtomy

  • Standard processing sequence: dehydration in graded ethanol (70-100%), clearing in xylene, then infiltration with paraffin at 56-60 degrees C.

  • Incomplete dehydration leaves the clearing agent milky and causes poor paraffin infiltration, section crumbling, and holes in the tissue.

  • Routine H&E sections are cut at 3-5 micrometres; small biopsies and kidney at 2-3 micrometres; IHC and FISH sections at 4-5 micrometres.

  • The flotation bath is set 5-10 degrees C below the paraffin melting point (about 40-45 degrees C); overheated water over-expands and distorts sections.

  • Clearance angle between the knife bevel and block face is 3-8 degrees; too little produces chatter, too much causes uneven or skipped sections.

  • Cryostat chamber temperature for routine frozen sections is approximately minus 20 to minus 25 degrees C, with sections cut at 4-6 micrometres.

  • Decalcification endpoint is confirmed by the ammonium oxalate chemical test or by radiography before the specimen is processed.

  • Blades are changed frequently; tungsten carbide blades are used for hard tissue and glass or diamond knives for resin and electron microscopy sections.

  • Section adhesives such as Mayer's egg albumin, poly-L-lysine, or APES are required for IHC, bone, and other sections prone to detachment.

Quick reference

Routine (H&E) and Special Staining

  • Hematoxylin is a basic (cationic) dye oxidized to hematein and complexed with aluminum; the complex stains nucleic acids blue-purple.

  • Eosin is an acidic (anionic) dye applied at slightly acidic pH, staining cytoplasm, collagen, and red cells pink after hematoxylin bluing.

  • Differentiation with 1% acid alcohol removes excess hematoxylin; bluing in Scott's tap water or lithium carbonate converts red alum-hematin to blue.

  • Progressive hematoxylin (Gill, Mayer) stains to endpoint, while regressive hematoxylin (Harris) deliberately overstains and is then differentiated.

  • PAS relies on periodic acid oxidizing 1,2-glycol groups to aldehydes that react with Schiff reagent, giving magenta staining of glycogen, mucin, and basement membrane.

  • Every special stain requires a positive control slide; PAS also uses a diastase-digested section to separate glycogen from other PAS-positive substances.

  • Prussian blue (hydrochloric acid plus potassium ferrocyanide) demonstrates hemosiderin; Masson trichrome stains collagen blue or green and muscle red.

  • Silver stains require oxidation and a reduction step: GMS uses chromic acid oxidation and stains fungal walls black against a green counterstain.

  • Gram stain sequence is crystal violet, iodine mordant, alcohol decolorizer, and safranin counterstain, leaving gram-negative organisms pink.

Quick reference

Immunohistochemistry and Molecular Techniques

  • Fixation in 10% NBF for 6-24 hours is preferred; both under-fixation and over-fixation reduce antigen detection and alter staining intensity.

  • HIER uses citrate pH 6.0 or Tris-EDTA pH 8.0-9.0 with a pressure cooker, steamer, or microwave; PIER uses enzymatic digestion with trypsin or pepsin at 37 degrees C.

  • Endogenous peroxidase is blocked with 3% hydrogen peroxide before primary antibody; endogenous biotin is blocked when avidin-biotin detection systems are used.

  • Each run includes a positive tissue control, a negative reagent (isotype) control, and an internal control element where one exists in the section.

  • DAB produces an insoluble brown precipitate resistant to alcohol, whereas AEC and Fast Red are alcohol soluble and require aqueous mounting media.

  • ASCO/CAP HER2 interpretation: IHC 3+ is positive; FISH is positive when HER2/CEP17 is 2.0 or greater or HER2 copies are 6.0 or greater per nucleus.

  • Estrogen receptor is reported positive when 1% or more of invasive tumor nuclei stain; the same threshold applies to progesterone receptor.

  • Molecular testing from FFPE depends on controlled cold ischemia and fixation time; 10 micrometre curls are cut with clean technique and an H&E slide guides tumor enrichment.

Quick reference

Laboratory Operations, Quality Assurance, and Safety

  • OSHA formaldehyde permissible exposure limit is 0.75 ppm as an 8-hour time-weighted average, with a 2 ppm short-term exposure limit.

  • Xylene is flammable and neurotoxic; it is dispensed in a fume hood with nitrile gloves, eye protection, and no ignition sources nearby.

  • Spilled formalin is absorbed with pads and neutralized with ammonium hydroxide or sodium bisulfite before disposal as hazardous waste.

  • A 16-section safety data sheet must be accessible for every hazardous chemical present in the laboratory.

  • Daily quality control includes documented temperature logs for processors, embedding centers, flotation baths, refrigerators (2-8 degrees C), and freezers.

  • Each slide is labeled with two patient identifiers plus specimen source and collection date, and is verified against the requisition before staining.

  • CAP/CLIA proficiency testing is required for predictive marker immunohistochemistry such as ER, PR, and HER2, with documented review of results.

  • Equipment maintenance records, reagent lot verification, and documented corrective action for failed QC are core components of the quality management system.