BCN study materials
BCN study resources

BCN cheat sheet

BCN · Keep the key ideas within reach.

Quick reference

Biofeedback: Principles, Modalities, and Instrumentation

  • Biofeedback is the operant conditioning of a physiologic response using amplified real-time signals; the signal must be perceivable, the response must be controllable, and feedback must be contingent on the response.

  • Biofeedback modalities and typical units: surface EMG (microvolts), thermal (degrees Fahrenheit/Celsius at fingertip), electrodermal (microsiemens), respiratory (breaths per minute), HRV (milliseconds or bpm), EEG (microvolts by band power).

  • Skin temperature feedback exploits vasoconstriction from sympathetic arousal; training aims at warmth (typically 95-97 F at the fingertip), a sign of peripheral vasodilation.

  • Electrodermal activity reflects eccrine sweat gland activity driven by sympathetic cholinergic fibers; skin conductance level rises with arousal and is measured in microsiemens.

  • Surface EMG electrode placement over the muscle belly with interelectrode distance of a few centimeters; 50/60 Hz notch filtering and proper skin prep reduce artifact; raw signal is rectified and averaged (RMS or moving average).

  • Respiratory training targets slow paced breathing near 6 breaths per minute (about 0.1 Hz), which increases respiratory sinus arrhythmia and baroreflex gain.

  • Therapist scope of practice: biofeedback certification requires knowledge of instrumentation, but clinical interpretation and treatment planning remain within the clinician's licensed scope; never diagnose outside licensure.

  • Artifact sources to rule out before interpreting a signal: 60 Hz electrical interference, movement, poor electrode contact, cardiac artifact in EMG, and respiration artifact in HRV or EEG.

Quick reference

Neurofeedback: EEG Bands, Protocols, and Practice

  • Standard EEG bands: delta <4 Hz, theta 4-8 Hz, alpha 8-13 Hz, beta 13-30 Hz, and high beta/gamma above 30 Hz; band power is quantified by FFT or by digital filtering.

  • Arousal model: theta and alpha generally index underarousal or relaxed states, beta indexes alert focus, and excess high beta or EMG artifact can mimic tension; sensorimotor rhythm (SMR) is 12-15 Hz over sensorimotor cortex.

  • Common protocols: alpha/theta training (eyes closed, posterior sites) for relaxation; SMR uptraining with theta/beta downtraining for attention and hyperactivity; beta uptraining for alertness.

  • Theta/beta ratio (TBR) is computed as theta power divided by beta power at a specified site; higher TBR is associated with inattention in some populations but is not diagnostic.

  • Reward/inhibit thresholds are set so the client succeeds roughly 70-80% of epochs; thresholds are adjusted adaptively as performance changes.

  • Electrode sites follow the 10-20 system (Fz, Cz, Pz, C3, C4, T3/T4, O1/O2); referential, sequential, and Laplacian montages differ in spatial specificity and artifact sensitivity.

  • Neurofeedback requires artifact control: eye blinks raise frontal delta/theta, jaw clenching raises temporal EMG, and drowsiness shifts spectra toward theta and alpha.

  • QEEG is a statistical comparison of a client's EEG to a normative database; it informs protocol selection but is not a stand-alone diagnosis, and clinician scope of practice governs its use.

Quick reference

HRV Biofeedback: Physiology, Metrics, and Protocol

  • Heart rate variability (HRV) is the beat-to-beat variation in R-R intervals; higher vagally mediated HRV generally indicates greater parasympathetic (vagal) tone.

  • Time-domain metrics: SDNN reflects overall variability; RMSSD and pNN50 reflect short-term vagal activity; these are computed from R-R (N-N) interval series.

  • Frequency-domain bands: VLF <0.04 Hz, LF 0.04-0.15 Hz, HF 0.15-0.40 Hz; HF power reflects respiratory sinus arrhythmia and vagal tone; LF/HF is a contested index of autonomic balance.

  • Resonance frequency is the breathing rate (typically 4.5-6.5 breaths/min, near 0.1 Hz) that maximizes HRV amplitude for an individual; it is found by paced-breathing sweeps and then used for training.

  • During resonance breathing, HR and respiration become phase-synchronized; the goal is a smooth, high-amplitude sinusoidal HR pattern, not merely a slow rate.

  • Breathing at resonance increases baroreflex gain and RSA amplitude; typical training sessions use 10-20 minutes of paced breathing with real-time HRV display.

  • HRV is confounded by age, posture, medications (beta-blockers), caffeine, hydration, and time of day; interpret trends within-person rather than against a single cutoff.

  • The LF band overlaps with baroreflex activity (~0.1 Hz), so peak LF power near 0.1 Hz during slow breathing is expected and is not evidence of sympathetic dominance.

Quick reference

Pelvic Muscle Dysfunction Biofeedback: Assessment and Training

  • Pelvic floor muscle (PFM) biofeedback uses surface EMG (vaginal or anal sensor, or surface electrodes) to teach contraction and relaxation; typical units are microvolts.

  • A correct PFM contraction shows increased EMG amplitude with an abdominal and gluteal muscle quieting; accessory muscle recruitment (abdominal, gluteal, hip adductor) indicates substitution.

  • Baseline resting tone is recorded before contraction; elevated resting EMG suggests hypertonicity or guarding, and training then emphasizes downtraining and relaxation rather than strengthening.

  • Typical contraction parameters: hold 5-10 seconds, rest 10 seconds or more between contractions, with a work-to-rest ratio that avoids fatigue; the exact prescription is individualized.

  • Incontinence (stress and urge) and pelvic pain are the common indications; pelvic pain and hypertonic dysfunction call for relaxation/downtraining, while stress incontinence often calls for strength and endurance training.

  • Patient education includes correct identification of the PFM (the 'stop the flow of urine' cue is for identification only, not for regular exercise), relaxed breathing, and avoiding breath-holding or Valsalva.

  • Outcome is tracked with EMG amplitude, endurance (hold time), coordination, and validated symptom questionnaires; progress is judged against the patient's own baseline.

  • Contraindications and cautions to screen before internal sensors: active vaginal or urinary infection, recent surgery, pregnancy-related precautions, and patient discomfort; obtain informed consent and follow infection-control procedures.