BONENT CHT flashcards
BONENT CHT · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Target spKt/V
- For thrice-weekly hemodialysis the minimally adequate single-pool Kt/V is 1.2, which corresponds to a urea reduction ratio of at least 65 percent.
- Urea reduction ratio formula
- URR equals pre-dialysis BUN minus post-dialysis BUN, divided by pre-dialysis BUN, times 100; it estimates urea clearance as a percentage.
- Dialysis disequilibrium syndrome
- Cerebral edema caused by overly rapid urea removal during dialysis; risk is greatest in the first treatments and is reduced by slower blood flow and shorter early sessions.
- Intradialytic hypotension criteria
- A systolic pressure drop of at least 20 mmHg, or mean arterial pressure drop of at least 10 mmHg, accompanied by symptoms such as cramps, nausea, or dizziness.
- Dry weight
- The lowest weight a patient tolerates at the end of a dialysis session without hypovolemic symptoms or evidence of fluid overload; it sets the ultrafiltration goal.
- Fistula rule of sixes
- A mature arteriovenous fistula has flow of at least 600 mL/min, diameter of at least 6 mm, depth no more than 6 mm, and is usable about 6 weeks after surgery.
- Carbon tank function
- Granular activated carbon removes chlorine and chloramine from feed water by adsorption; total chlorine must be tested before each shift and kept at or below 0.1 mg/L.
- Reverse osmosis rejection
- RO membranes reject roughly 90 to 99 percent of dissolved ions, bacteria, and endotoxins; declining percent rejection indicates membrane damage or fouling.
- Product water bacterial limit
- AAMI and ISO standards limit water used to prepare dialysate to no more than 100 CFU/mL with endotoxin below 0.25 EU/mL.
- Water softener mechanism
- An ion-exchange resin swaps sodium for calcium and magnesium, preventing scale; hardness is checked daily and the resin is regenerated with brine.
- Deionization limitation
- Deionization removes dissolved ions but not bacteria or endotoxin, and an exhausted bed can release adsorbed ions and organics into the product water.
- Ultrafiltration coefficient Kuf
- Kuf is the milliliters of fluid removed per hour per mmHg of transmembrane pressure; UF rate equals Kuf multiplied by TMP.
- High TMP alarm
- A rising transmembrane pressure at a fixed UF goal usually signals clotting fibers, a kinked line, or inadequate dialysate flow; inspect the dialyzer and circuit.
- Conductivity normal range
- Dialysate conductivity is typically 13 to 15 mS/cm at 25 degrees Celsius, roughly equal to a sodium concentration of 140 mEq/L, and is verified with an independent meter.
- Air detector alarm action
- The blood pump stops automatically; clamp the lines and do not return any blood to the patient until the air has been removed from the circuit.
- Blood leak detector
- A photometric sensor in the dialysate path detects blood; it stops dialysate flow and ultrafiltration, and blood is not reinfused until the leak is excluded.
- Total cell volume criterion
- For dialyzer reuse, total cell volume must remain at least 80 percent of the original value; below that threshold the dialyzer is discarded.
- Hepatitis B isolation rule
- Patients who are hepatitis B surface antigen positive are dialyzed in a separate room with dedicated equipment because hepatitis B is the most transmissible bloodborne virus in dialysis units.
- Hepatitis C precautions
- Hepatitis C and HIV do not require isolation or dedicated machines; standard precautions with dedicated patient equipment and routine disinfection are sufficient.
- Blood spill disinfection
- Routine blood spills are decontaminated with a 1:100 household bleach dilution (about 500 to 600 ppm chlorine); gross contamination uses a 1:10 dilution.
- Needlestick post-exposure steps
- Wash with soap and water, report immediately, and start HIV post-exposure prophylaxis ideally within 2 hours and no later than 72 hours.
- Hand hygiene indications
- Perform hand hygiene before patient contact, after glove removal, after touching blood or contaminated equipment, and whenever moving between patients.
- Technician scope of practice
- The certified hemodialysis technician delivers treatments under RN or nephrologist supervision and does not independently diagnose, prescribe, or change the dialysis prescription.
- Unsafe order decision rule
- If an order or machine condition seems unsafe, stop the action, clarify with the supervising RN or physician, document the concern, and escalate through the chain of command.
- Late entry documentation
- Record the time the event actually occurred, add the time the note was written, label it as a late entry, and never alter or backdate the original record.
- Patient confidentiality rule
- Protected health information is shared only for treatment, payment, or health care operations, or when the patient authorizes release in writing.
- BONENT recertification cycle
- BONENT certifications are time-limited, renewed on a four-year cycle through continuing education or re-examination as required to maintain active credential status.
- Conditions for Coverage
- CMS regulations at 42 CFR Part 494 establish minimum health, safety, and quality requirements for ESRD facilities and are enforced through state surveys.
- OSHA bloodborne pathogens standard
- 29 CFR 1910.1030 requires a written exposure control plan, engineering controls, personal protective equipment, training, and free hepatitis B vaccination for at-risk employees.
- AAMI RD47 and RD52
- AAMI RD47 addresses dialyzer reprocessing and RD52 addresses dialysate quality; the RD62 and ISO 23500 series cover water treatment for dialysis.
- Teach-back method
- The learner explains or demonstrates the instruction in their own words; this confirms comprehension and is preferred for patients with low health literacy.
- Plan of correction
- A facility's written response to survey deficiencies that states the corrective action, the person responsible, and the date the correction will be completed.