CMD study materials
CMD study resources

CMD flashcards

CMD · Build recall with a focused review session.

32 cards. Read the term and the definition here. The deck at the top of this page is optional.

Gray definition
The SI unit of absorbed dose, equal to one joule of energy deposited per kilogram of matter; one gray equals 100 centigray and 100 rad.
Inverse-square law
Photon fluence and dose rate decrease with the square of distance from a point source, so doubling the distance reduces dose rate to one quarter.
Percent depth dose
The ratio of absorbed dose at a given depth to the dose at the depth of maximum buildup, expressed as a percentage, for a fixed source-to-surface distance.
Tissue-phantom ratio
Dose at depth in a phantom divided by dose at a fixed reference depth, measured with constant source-to-detector distance; independent of source-to-surface distance.
Backscatter factor
The ratio of dose at the depth of maximum buildup for a finite field to the dose at the same point for a zero-area field; also called peak scatter factor.
Mayneord F factor
A conversion factor used to transform percent depth dose values measured at one source-to-surface distance into values for a different source-to-surface distance.
Equivalent square field
A square field with the same scatter characteristics as a rectangular field, calculated as twice the product of the sides divided by their sum.
Monitor unit definition
A machine unit calibrated so that one monitor unit delivers one centigray at the depth of maximum dose in water under reference conditions.
Collimator scatter factor
The ratio of machine output for a given field size to output for the reference field size, measured in air with a buildup cap.
Wedge factor
The ratio of dose with the wedge in place to dose without it at the same depth and monitor units; depends on depth and field size.
Electron practical range
The depth at which the electron percent depth dose falls to approximately 90 percent, roughly equal to beam energy in MeV divided by four.
Buildup region
The shallow depth range where photon dose rises toward maximum as secondary electrons accumulate; skin sparing depends on this effect.
Gamma index
A composite metric combining dose difference and distance-to-agreement; a point passes when the minimum combined value is less than or equal to one.
TG-218 passing criterion
Recommended IMRT verification standard of 95 percent of points passing at 3 percent dose difference and 2 millimeter distance-to-agreement with a 10 percent threshold.
Van Herk margin formula
Clinical target volume to planning target volume margin equal to 2.5 times the systematic error plus 0.7 times the random error, in standard deviations.
Patient-specific IMRT QA
Measurement of a planned fluence on a phantom or detector array before treatment to confirm that the delivered dose matches the calculated dose.
Stereotactic setup accuracy
Intracranial stereotactic treatments require targeting accuracy of about one millimeter, verified with cone-beam CT or stereoscopic imaging and six-degree couch correction.
Volumetric modulated arc therapy
A rotational delivery technique in which gantry speed, dose rate, and multileaf collimator positions vary continuously during one or more arcs around the patient.
Total body irradiation
A technique using extended source-to-surface distance, large fields, and tissue compensators to deliver a uniform midline dose, often before bone marrow transplant.
Dose-rate constant
In TG-43 formalism, the dose rate in water at one centimeter along the transverse axis per unit air kerma strength, with units of centigray per hour per unit.
Radial dose function
A TG-43 factor describing attenuation and scatter in water along the transverse axis, normalized to unity at the reference distance.
Anisotropy function
A TG-43 factor accounting for source encapsulation and self-attenuation as a function of angle relative to the transverse axis.
Air kerma strength
The product of air kerma rate at one meter and the square of that distance, expressed in units of microgray square meters per hour.
Iodine-125 half-life
A permanent implant isotope with a half-life of about 59 days and an average photon energy near 28 kiloelectron volts.
Iridium-192 HDR
A high-dose-rate brachytherapy isotope with a half-life near 74 days and an average photon energy of about 380 kiloelectron volts.
Palladium-103 properties
A permanent implant isotope with a half-life of about 17 days and an average photon energy near 21 kiloelectron volts, used for prostate implants.
ALARA principle
The radiation protection principle that occupational and public exposures should be kept as low as reasonably achievable, accounting for economic and social factors.
Half-value layer
The thickness of a specified material that reduces the intensity of a narrow beam to one half of its original value; equals 0.693 divided by the attenuation coefficient.
Primary barrier calculation
Barrier transmission equals the weekly design limit times distance squared, divided by workload times use factor times occupancy factor.
CT dose index volume
A CT dose descriptor representing the average dose over a scan volume, calculated from the weighted CT dose index divided by pitch.
Dose length product
The product of volume CT dose index and scan length in centimeters, expressed in milligray-centimeters; used to estimate effective dose.
Occupational dose limit
The annual effective dose limit for radiation workers is 50 millisieverts, with a lifetime cumulative limit of ten millisieverts times age.