Quick reference

Decontamination

  • Decontamination is the first step of reprocessing: it removes bioburden and soil so later cleaning, inspection, and sterilization can work; it does not sterilize instruments.

  • Standard PPE for the decontamination area includes fluid-resistant gown, gloves, mask, eye protection, and shoe covers; sharps handling requires puncture-resistant utility gloves.

  • Traffic flow is one-way: dirty-to-clean. Soiled items enter at the decontamination side; clean items exit on the clean side; doors between areas stay closed.

  • Manual cleaning steps in order: presoak or pretreat, then wash with detergent at correct dilution and temperature, then rinse, then dry.

  • Enzymatic detergent breaks down protein, fat, and carbohydrate soil; it is single-use in most facilities and loses activity if reused or left standing.

  • Water quality matters: final rinse water should be treated (deionized or reverse osmosis) to prevent mineral and pyrogen deposits that interfere with sterilization.

  • Lumened instruments need brushing with a lumen brush sized to the channel, flushing with detergent, rinsing, and drying with forced air before assembly.

  • Powered equipment must be disassembled per manufacturer instructions before cleaning; batteries are removed and contacts cleaned separately.

  • Mechanical washer-disinfectors are monitored with wash-cycle verification: check detergent dose, water temperature, spray-arm rotation, and final rinse; log each cycle.

Quick reference

Inspection, Assembly, and Packaging

  • Inspection checks function, cleanliness, and integrity: hinges move freely, ratchets hold, jaws align, insulation is unbroken, and no corrosion, cracks, or retained soil remain.

  • A lumen is inspected by looking through it against a light source; a borescope is used for long or narrow channels where direct vision is impossible.

  • Instrument sets are assembled against a count sheet or pick list; each tray is verified item by item, and missing or extra items are resolved before packaging.

  • Packaging must be compatible with the chosen sterilization method: steam needs steam-permeable wrap; low-temperature methods require materials validated for that process.

  • Sterile barrier systems must allow air removal and sterilant penetration, provide a sterile barrier to microorganisms, and resist tears and punctures during handling.

  • Chemical indicators are placed inside every package; an external indicator is used on packages going into steam so processed and unprocessed items can be distinguished.

  • Package labeling includes contents, sterilizer ID, cycle or load number, date of sterilization, and expiration or event-related date; labeling happens before sterilization.

  • Package weight and density limits follow the sterilizer manufacturer's written instructions; overloading blocks steam penetration and causes wet packs.

  • Tamper-evident seals must be continuous; a package with a broken seal, tear, or visible moisture is considered non-sterile and is reprocessed.

Quick reference

Sterilization

  • Steam sterilization parameters: 250°F (121°C) for 30 minutes or 270°F (132°C) for 4 minutes in a gravity displacement cycle; prevacuum cycles are shorter.

  • Steam quality matters: saturated steam with less than 3% moisture is required; superheated or wet steam causes failed cycles and wet packs.

  • Biological indicators for steam use Geobacillus stearothermophilus spores; they are incubated and read per manufacturer instructions after the cycle.

  • Ethylene oxide (EO) is a low-temperature method for heat- and moisture-sensitive items; it requires aeration to remove residual gas before items are released.

  • Hydrogen peroxide gas plasma is a low-temperature, non-toxic-residue method; it requires items and packaging validated for that specific system.

  • Monitoring has three levels: physical (gauges, printouts), chemical (indicators), and biological (spore tests); all three must pass for release of an implant load.

  • Implant loads are held until a biological indicator is negative; if the BI is positive, the load is recalled and the sterilizer is taken out of service.

  • Sterilizer loading: place items so steam can circulate, do not stack trays tightly, and place rigid containers flat with lids per manufacturer instructions; never overload.

  • Each sterilizer is tested at least weekly with a biological indicator and daily with a Bowie-Dick test for prevacuum sterilizers, per facility policy and standards.

Quick reference

Sterile Storage and Distribution

  • Sterile storage areas are clean, dry, temperature-controlled, and separated from the decontamination area; relative humidity and temperature are monitored and logged.

  • Sterile items are stored at least 8 to 10 inches above the floor, 2 inches from walls, and 18 inches below the ceiling to allow cleaning and air circulation.

  • Event-related shelf life uses package integrity and storage conditions rather than a fixed date; a package remains sterile until an event compromises it.

  • A package is considered compromised by tears, holes, water stains, crushed corners, broken seals, or being dropped; it is reprocessed, not used.

  • Distribution uses first-in, first-out (FIFO) rotation to prevent expired or deteriorated products from reaching the point of use.

  • Inventory is tracked by par levels and minimum/maximum quantities; stockouts are prevented by reorder points tied to usage data.

  • Transport of sterile items uses closed carts or covered containers; soiled and clean items never share the same cart or route.

  • Case carts are assembled from physician preference cards; missing items are documented and communicated to the OR before the case starts.

  • Sterile storage requires that shelving be smooth and cleanable; cardboard boxes are not used because they harbor dust and moisture.

Quick reference

Professional Knowledge and Skills

  • Regulatory roles: FDA clears sterilizers and biological indicators; OSHA enforces bloodborne pathogen and PPE rules; AAMI publishes technical standards; Joint Commission surveys compliance.

  • Standard Precautions treat all blood and body fluids as infectious; hand hygiene is performed before and after glove use and after contact with contaminated items.

  • Sharps injuries are managed by washing the site with soap and water, reporting immediately, and following the facility exposure control plan for evaluation and prophylaxis.

  • Quality assurance uses process monitors, audits, and documentation; every cycle is recorded with load contents, parameters, operator, and results.

  • Recall procedure: identify the affected load, quarantine it, notify users, document the event, and perform root cause analysis before resuming.

  • Chemical and biological indicator results are documented and retained per facility policy and accreditation requirements; records must be traceable to the load.

  • Inventory management uses par levels, usage reports, and periodic cycle counts; expired or recalled products are removed and documented.

  • Staff competency is verified at hire and annually; training records for sterilization, decontamination, and packaging are maintained.

  • Safety data sheets (SDS) are available for every chemical used; staff are trained on hazards, spill response, and waste handling.