Therapeutic Radiographer flashcards
Therapeutic Radiographer · Build recall with a focused review session.
33 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Inverse square law
- Dose rate from a point source varies inversely with the square of distance: D1/D2 = (d2/d1)². Doubling distance quarters the dose rate, which underpins distance-based protection.
- Half-value layer
- Thickness of a material that reduces beam intensity by half. HVL = ln2/μ = 0.693/μ, where μ is the linear attenuation coefficient. Used to specify beam quality and shielding.
- Klystron versus magnetron
- Both generate microwave power for linear accelerators. Klystrons amplify a low-power signal to high peak power; magnetrons oscillate directly and produce lower peak power. Both feed the waveguide.
- Flattening filter function
- A conical absorber in the linac head that makes the photon fluence uniform across the field. Removing it increases dose rate but requires compensation in planning, as in flattening-filter-free beams.
- Cobalt-60 gamma energies
- Cobalt-60 decays with a 5.26-year half-life, emitting gamma rays at 1.17 and 1.33 MeV. The average energy used in planning is about 1.25 MeV.
- Electron practical range
- The depth in water at which the electron depth-dose curve falls to near zero. It approximates energy in MeV divided by 2, so a 12 MeV beam reaches about 6 cm.
- ALARA principle
- As low as reasonably achievable: optimisation of protection by keeping doses below limits while considering economic and social factors. Applied through time, distance, and shielding.
- Occupational dose limits
- For radiation workers: 20 mSv per year averaged over 5 years, with no year exceeding 50 mSv. Eye lens limit is 20 mSv/year; skin and extremities 500 mSv/year.
- Effective dose formula
- Effective dose E equals the sum of tissue equivalent doses multiplied by tissue weighting factors: E = Σ wT·HT. It allows comparison of stochastic risk across exposure types.
- Radiation weighting factors
- Factors converting absorbed dose to equivalent dose: 1 for photons and electrons, 2 for protons, 20 for alpha particles, and energy-dependent values of 2.5 to 20 for neutrons.
- Pregnant worker fetal dose
- After a worker declares pregnancy, the dose to the fetus should not exceed 1 mSv for the remainder of the pregnancy. Working conditions may be adjusted to meet this constraint.
- Controlled area threshold
- An area where occupational exposure is possible and dose rates exceed 7.5 μSv/h, or where annual dose could exceed 6 mSv. Requires local rules, signage, and dose monitoring.
- Informed consent elements
- Valid consent requires patient capacity, sufficient information about the procedure's purpose, benefits, risks and alternatives, and voluntary agreement without coercion. Documented before treatment starts.
- Radiotherapy skin reaction timeline
- Erythema typically appears in week 2 to 3, followed by dry desquamation and possible moist desquamation in skin folds. Managed with gentle washing, emollients, and avoidance of friction.
- Thermoplastic mask use
- A moulded mask that immobilises the head and neck and reproduces set-up daily. It is indexed to the couch and marked with reference points for isocentre alignment.
- Oral mucositis management
- Mucositis from head and neck radiotherapy peaks around weeks 4 to 5. Care includes saline or bicarbonate rinses, oral hygiene, analgesia, and nutritional support when eating is impaired.
- Teach-back communication
- A method where the patient explains information in their own words, confirming understanding. It identifies gaps and supports informed consent and safe self-care during treatment.
- Four Rs of radiobiology
- Repair of sublethal damage, redistribution through the cell cycle, repopulation of surviving cells, and reoxygenation of hypoxic tumour cells. These processes shape fractionation response.
- Linear-quadratic model
- Surviving fraction S = exp(−(αD + βD²)). The α/β ratio is about 10 Gy for tumours and early effects, and about 3 Gy for late-reacting normal tissues.
- Biologically effective dose
- BED = nd(1 + d/(α/β)) for n fractions of dose d. It compares schedules with different fraction sizes and is converted to EQD2 by dividing by (1 + 2/(α/β)).
- Oxygen enhancement ratio
- Ratio of dose needed under hypoxia to dose under full oxygenation for the same effect. It is about 2.5 to 3 for photons, reflecting the radiosensitising role of oxygen.
- Radiosensitivity ranking
- Most radiosensitive tissues include lymphoid tissue, bone marrow, germ cells, and intestinal epithelium. Least sensitive include neurons, muscle, and mature connective tissue.
- ICRU target volumes
- GTV is visible tumour, CTV adds microscopic spread, and PTV adds margin for set-up and organ motion. OARs and PRVs are also defined for planning and reporting.
- PTV margin formula
- Margin approximately equals 2.5 times the systematic error standard deviation plus 0.7 times the random error standard deviation, accounting for both set-up and organ motion uncertainty.
- DVH interpretation
- A dose-volume histogram shows the fraction of a structure receiving at least a given dose. Key metrics include D95 for target coverage and V20 for lung.
- Inverse planning concept
- Planning starts from dose-volume objectives and iteratively optimises fluence or machine parameters to meet them. It underlies IMRT and VMAT, unlike manual forward planning.
- Image guidance tolerance
- Acceptable set-up correction depends on site: about 3 mm for intracranial targets and 5 mm for pelvis. Exceeding tolerance triggers repositioning and repeat imaging before treatment.
- VMAT delivery characteristics
- Volumetric arc therapy delivers continuous gantry rotation with dynamic multileaf collimation and variable dose rate. Treatment is typically 2 to 4 minutes per fraction.
- Portal imaging purpose
- Electronic portal imaging verifies field shape and position against reference images. It can also support in vivo dosimetry and transit dose measurement.
- Six-degree-of-freedom couch
- A treatment couch that corrects three translational and three rotational set-up errors. Rotational correction is essential for intracranial and spinal stereotactic treatments.
- Daily linac QA checks
- Output constancy within 3%, laser alignment, door interlock, beam-on indicator, and audiovisual monitoring are verified each treatment day before patients are treated.
- TG-51 reference dosimetry
- A protocol for calibrating clinical beams in terms of absorbed dose to water using an ion chamber with a calibration factor traceable to a standards laboratory.
- MLC picket fence test
- A test imaging adjacent leaf pairs at set positions to reveal gap or overlap errors. Leaf position accuracy should be within about 1 mm for stereotactic treatments.