LC Technology flashcards
LC Technology · Build recall with a focused review session.
40 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Retention factor k
- k = (tR − t0)/t0, where tR is the analyte retention time and t0 the unretained marker time. Practical values are 1–10.
- Resolution equation
- Rs = 2(tR2 − tR1)/(w1 + w2). Baseline resolution is generally Rs ≥ 1.5; Rs of 1.0 gives about 2% overlap.
- Van Deemter equation
- H = A + B/u + C·u describes plate height versus linear velocity. The A term is eddy diffusion, B longitudinal diffusion, C mass transfer resistance.
- Plate number N
- N = 16(tR/w)² or 5.54(tR/w1/2)², measuring column efficiency. Higher N narrows peaks and improves resolution by the square root.
- Selectivity factor α
- α = k2/k1, the ratio of retention factors of two adjacent peaks. Selectivity is the most efficient lever for improving resolution.
- Reversed-phase retention
- Retention increases with analyte hydrophobicity and decreases as organic content rises. Ionizable analytes shift retention with mobile-phase pH.
- HILIC retention mechanism
- HILIC retains polar analytes on a polar stationary phase with high organic mobile phase. It requires at least 80% organic at start and aqueous-rich injection solvent.
- Gradient LSS model
- log k = log kw − S·φ relates retention to organic fraction φ. Gradient peak capacity depends on gradient time, flow, and column dead volume.
- Degassing purpose
- Removing dissolved gas from mobile phase prevents bubbles that cause pump noise, baseline spikes, and retention-time drift. Methods include helium sparge and in-line vacuum.
- Autosampler carryover
- Residual analyte from a prior injection contaminates the next run. Needle wash with strong then weak solvent and blank injections quantify and reduce it.
- PDA peak purity
- Photodiode array detectors compare spectra across a peak. Spectral match and absorbance ratio flag co-eluting impurities that share retention time.
- ESI versus APCI
- ESI ionizes polar and ionic analytes from solution; APCI suits nonpolar, low-polarity compounds. Mobile-phase pH and composition strongly affect ESI response.
- Refractive index detector limits
- The RI detector is universal but incompatible with gradient elution and sensitive to temperature. It offers lower sensitivity than UV or fluorescence detection.
- Column oven stability
- Temperature control to ±0.1–0.5 °C stabilizes retention and selectivity. Drift in oven temperature produces baseline drift and shifting retention times.
- Extra-column volume
- Tubing and fittings after the column add dead volume that broadens bands. Low-dead-volume connections preserve resolution, especially on narrow-bore columns.
- Mobile-phase buffer choice
- Phosphate buffers cover pH 2–8 but are nonvolatile. Ammonium formate and acetate are volatile and preferred for MS-compatible methods.
- Method transfer scaling
- Keep linear velocity constant when changing column diameter by scaling flow with d². Keep gradient volume constant when changing length or flow.
- Forced degradation study
- Acid, base, oxidation, heat, and light stress demonstrate specificity. Mass balance and peak purity confirm degradants are separated from the main peak.
- System suitability criteria
- Typical acceptance limits: resolution ≥ 1.5–2.0, tailing factor ≤ 2.0, replicate injection RSD ≤ 2.0%, and plates above a method-defined minimum.
- ICH Q2 validation parameters
- Accuracy, precision, specificity, linearity, range, detection and quantitation limits, and robustness, each with predefined acceptance criteria and statistical basis.
- Injection volume rule
- Injection volume should be no more than 1–2% of column volume. A weaker injection solvent than the mobile phase prevents peak distortion.
- Protein precipitation
- Adding at least three volumes of acetonitrile or methanol to sample precipitates protein. Centrifugation and filtration follow before injection.
- SPE sequence
- Condition, equilibrate, load, wash, and elute. Sorbent chemistry must match analyte polarity, and controlled flow prevents channeling.
- Internal standard choice
- A stable-isotope-labeled analog is preferred for LC-MS because it co-elutes and corrects recovery and ionization. A structural analog is the fallback.
- Carryover acceptance
- A blank injected after a high-concentration sample should show carryover below the method LOQ, commonly ≤20% of LOQ response.
- Sample stability testing
- Bench-top, refrigerated, freeze-thaw, and autosampler residence conditions are evaluated. Autosampler temperature affects volatile or degradable analytes.
- LOD and LOQ formulas
- LOD = 3.3σ/S and LOQ = 10σ/S, where σ is the standard deviation of the response and S the slope. Signal-to-noise of 3:1 and 10:1 also apply.
- Calibration model selection
- External standards suit clean matrices. Internal standards correct variable recovery and injection. Standard addition suits matrix-matched quantitation.
- Linearity assessment
- Use at least five or six calibration levels. Correlation coefficient alone is insufficient; residual plots reveal curvature and heteroscedasticity.
- Data integrity ALCOA+
- Records must be attributable, legible, contemporaneous, original, and accurate. Audit trails, unique accounts, and controlled electronic records support this.
- High backpressure diagnosis
- Disconnect the column and measure system pressure to locate the restriction. Clogged frits, guard columns, and inline filters are common causes.
- Ghost peak causes
- Carryover, contaminated mobile phase, or prior injections produce peaks absent from the blank. A blank gradient and fresh solvent isolate the source.
- Peak tailing causes
- Silanol interactions, a partially blocked frit, or column voiding produce tailing. Check pH, guard column condition, and fitting integrity.
- Baseline drift causes
- Inadequate degassing, lamp aging, incomplete equilibration, and temperature instability cause drift. A UV baseline drift often indicates a failing lamp.
- Pump seal maintenance
- Replace pump seals at manufacturer-defined intervals and verify check-valve function. Noisy pump traces often indicate a failing check valve or air.
- Peroxide-forming solvents
- THF, diethyl ether, dioxane, and isopropanol form peroxides over time. Date containers on receipt and test or discard before expiration.
- HPLC waste classification
- Solvent-rich HPLC waste is hazardous when ignitable or toxic. Segregate aqueous and organic streams and label with accumulation start date.
- Fume hood face velocity
- Hoods are typically maintained at 0.3–0.5 m/s (60–100 ft/min). Verify with an anemometer and confirm sash height markers before use.
- Compressed gas cylinder safety
- Cylinders must be secured at one-third and two-thirds height. Regulators are gas-specific; hydrogen service requires compatible equipment.
- Chemical hygiene plan
- The plan documents SOPs, engineering controls, PPE, training records, SDS access, and waste procedures for each laboratory method.