Maharashtra HSC Medical Lab Technology (Vocational): Laboratory Instrumentation, Glassware and Reagents

Maharashtra HSC Medical Lab Technology (Vocational) · Work through your exam, one topic at a time.

All guide sections

Laboratory Instrumentation, Glassware and Reagents

What this domain tests

This domain covers spectrophotometry, centrifugation, pipetting, glassware and reagent handling - the practical basis of accurate measurement.

Core rules

  • A spectrophotometer follows Beer-Lambert law: absorbance is proportional to concentration and path length. Zero the instrument with a reagent blank before reading standards and samples.
  • Use clean, scratch-free cuvettes; wipe the outside; avoid bubbles and fingerprints. Match cuvettes and use the correct wavelength.
  • Balance the centrifuge by placing tubes of equal weight opposite each other. Do not open the lid until the rotor stops.
  • Calibrate micropipettes and balances regularly; use the correct tip and never pipette by mouth.
  • Class A volumetric glassware is calibrated for accuracy. Clean glassware with detergent, rinse with distilled water and dry; do not use soda-lime glass for strong alkali.
  • Store reagents at the recommended temperature, label opening dates, and never use expired or contaminated reagents.

Worked example

A glucose sample has absorbance 0.30; a 100 mg/dL standard has absorbance 0.60. Reasoning: concentration is proportional to absorbance, so sample concentration = (0.30/0.60) × 100 = 50 mg/dL. The blank must be subtracted from both readings first.

Traps

  • Reading without a blank or with a dirty cuvette.
  • Using an unmatched cuvette or the wrong wavelength.
  • Centrifuging unbalanced tubes.
  • Assuming a pipette is accurate without calibration.