A failed autoclave load and a needlestick on the same morning
At a primary health centre, 18 stainless-steel dressing instruments have been processed in a single autoclave load before a busy dressing round. The written log records 121°C and 15 psi for 15 minutes, but the operator mentions that pressure dropped to 10 psi for about 4 minutes midway and the cycle was extended by 5 minutes. The packs were wrapped in cloth, tied tightly, and loaded so that the chamber was completely full. The chemical indicator strip inside one pack changed colour, but the strip in another pack did not. The autoclave's pressure gauge was calibrated eight months ago. The staff nurse is asked whether the load can be released for use. Later the same morning, a nurse sustains a needlestick from a needle used to draw blood from a patient whose HIV status is unknown.
1. Should the load be released? Give the decision and the corrective actions before any instrument is used.
Read the worked answer
Do not release any of the 18 instruments. The cycle is invalid because pressure fell below the required level, even though the time was extended, and because one chemical indicator did not change. A failed indicator means steam penetration or exposure was inadequate for that pack; since the load was tightly packed and tied, the failure may affect several packs, not just one. Extending time does not correct a pressure drop because the required temperature may not have been maintained. The correct action is to quarantine the whole load, reprocess all instruments after cleaning, drying and inspection, and use alternative sterile stock or refer patients if no safe stock exists. Correct the loading: packs should be loosely wrapped, placed upright with space between them, and the chamber not overfilled. Check the door seal, pressure gauge and air removal. Record the failed cycle, inform the in-charge, and arrange maintenance and calibration. Do not use the one pack with a passing strip as evidence that the rest are sterile.
2. Outline the immediate and follow-up management of the needlestick injury, including what must not be delayed.
Read the worked answer
Immediately wash the wound with soap and running water; do not squeeze, suck or scrub the site. Apply an antiseptic such as povidone-iodine or alcohol and cover it. Report to the officer in charge at once and start post-exposure prophylaxis assessment without waiting for the source patient's result. Delay is the main danger: HIV post-exposure prophylaxis should begin as soon as possible, ideally within two hours and no later than 72 hours, following the local protocol. Take a baseline sample from the nurse with consent and arrange testing of the source patient with counselling and consent, plus hepatitis B and C assessment. Check the nurse's hepatitis B vaccination status. Document the time, instrument, depth, source patient, and action taken. Arrange follow-up testing at the recommended intervals and counsellor support. If the source is known HIV-positive or has risk factors, the protocol may require a full course of antiretroviral prophylaxis; this decision belongs to the medical officer, not the nurse alone.
3. How should the used needle, contaminated gloves and the blood spill be managed?
Read the worked answer
Do not recap, bend or break the needle. Drop it directly into a puncture-proof, leak-proof sharps container kept at the point of use, and replace the container when it is three-quarters full. Used gloves and contaminated dressings go into yellow infectious waste, while the sharps container follows the sharps stream; do not put gloves into the sharps box. For the blood spill, wear gloves and an apron, cover the spill with absorbent material, then clean the area with detergent and water and disinfect with sodium hypochlorite solution prepared according to local protocol, usually around 1% available chlorine for blood spills. Allow contact time before wiping. Discard cleaning materials as infectious waste and wash hands afterwards. Record the spill and the needlestick in the incident register. These steps protect staff and patients and also reveal system problems, such as an overfilled autoclave or a sharps container placed too far from the procedure trolley.