IBHRE CCDS key terms
IBHRE CCDS · Get comfortable with the vocabulary.
- Sensing threshold
- Related: Sensitivity, Undersensing
The smallest intrinsic intracardiac electrogram amplitude that a device reliably detects as a cardiac event at the programmed sensitivity setting, measured in millivolts during testing.
- Sensitivity
- Related: Sensing threshold, Oversensing
The programmed voltage level above which a device counts an electrical signal as a sensed event; lowering the number makes the device more sensitive to small signals.
- Pacing threshold
- Related: Strength-duration curve, Capture
The minimum stimulus energy, usually expressed in volts or milliamps at a given pulse width, that consistently produces cardiac depolarization when the heart is paced.
- Strength-duration curve
- Related: Pacing threshold, Pulse width
A plot of the minimum current or voltage needed for capture at each pulse width, showing that shorter pulses require higher output to depolarize myocardium.
- Pulse width
- Related: Pacing threshold, Strength-duration curve
The duration of each pacing stimulus, typically programmed in milliseconds; longer widths lower threshold but consume more battery energy.
- Capture
- Related: Pacing threshold, Loss of capture
Depolarization of the myocardium produced by a pacing stimulus, confirmed on ECG by a P wave or QRS complex following the pacing artifact.
- Loss of capture
- Related: Capture, Pacing threshold
A pacing artifact that fails to produce a resulting P wave or QRS complex, indicating lead dislodgement, elevated threshold, insulation failure, or depleted output.
- Undersensing
- Related: Sensitivity, Oversensing
Failure of the device to detect an intrinsic cardiac event because the signal falls below the programmed sensitivity, which can lead to inappropriate pacing during the cardiac cycle.
- Oversensing
- Related: Undersensing, Far-field sensing
Detection of electrical signals that are not true cardiac depolarizations, such as T waves, myopotentials, or electromagnetic interference, causing inappropriate inhibition or counting.
- Far-field sensing
- Related: Oversensing, Crosstalk
Detection by an atrial or ventricular channel of depolarization originating in the opposite chamber, because the signal is large enough to exceed that channel's sensitivity setting.
- Crosstalk
- Related: Far-field sensing, Blanking period
An atrial lead sensing the ventricular pacing stimulus, or a ventricular lead sensing the atrial stimulus, which can trigger inappropriate inhibition or mode switching.
- Blanking period
- Related: Crosstalk, Refractory period
A brief programmed interval after a paced or sensed event during which the device ignores input on a channel to prevent oversensing of its own output or the opposite chamber's signal.
- Refractory period
- Related: Blanking period, Mode switching
A time window after a sensed or paced event during which the device does not respond to further signals on that channel, though it may still count them for rate purposes.
- Lower rate limit
- Related: Upper tracking rate, Hysteresis
The slowest pacing rate the device will deliver in a given mode, programmed to match the patient's expected physiologic needs at rest.
- Upper tracking rate
- Related: Lower rate limit, Mode switching
The fastest rate at which a dual-chamber device will pace the ventricle in response to sensed atrial events, protecting against rapid ventricular pacing during atrial tachyarrhythmias.
- Mode switching
- Related: Upper tracking rate, Refractory period
An automatic change from a tracking mode such as DDD to a non-tracking mode such as DDI when the device detects an atrial tachyarrhythmia, preventing rapid ventricular pacing.
- Hysteresis
- Related: Lower rate limit, Rate response
A programmable feature that allows the paced rate to fall below the lower rate limit after a sensed event, promoting intrinsic conduction before pacing resumes.
- Rate response
- Related: Hysteresis, Chronotropic incompetence
A sensor-driven feature that raises the pacing rate during exertion or stress, typically using minute ventilation, accelerometer, or impedance-based sensors.
- Chronotropic incompetence
- Related: Rate response
The inability of the sinus node to increase heart rate appropriately with metabolic demand, a common indication for rate-responsive pacing.
- AV delay
- Related: Ventricular safety pacing, Mode switching
The programmed interval between an atrial event and the subsequent ventricular pacing stimulus, adjusted to optimize filling and avoid unnecessary ventricular pacing.
- Ventricular safety pacing
- Related: Crosstalk, AV delay
A protective feature that delivers a ventricular pacing stimulus shortly after an atrial event when crosstalk is suspected, preventing inappropriate inhibition of ventricular output.
- Defibrillation threshold testing
- Related: Shock impedance, Tachycardia detection zones
Intraprocedural delivery of a test shock to confirm that the ICD can reliably terminate induced ventricular fibrillation at the programmed energy setting.
- Tachycardia detection zones
- Related: Defibrillation threshold testing, Antitachycardia pacing
Programmed rate intervals that classify sensed ventricular events as ventricular tachycardia or ventricular fibrillation, determining which therapy the device delivers.
- Antitachycardia pacing
- Related: Tachycardia detection zones, Shock impedance
Delivery of rapid pacing bursts through the ICD lead to terminate ventricular tachycardia without a shock, reducing patient discomfort and battery drain.
- Shock impedance
- Related: Defibrillation threshold testing, Lead impedance
The electrical resistance measured across the defibrillation pathway, used to confirm lead integrity and adequate coil-to-can contact before and after shock delivery.
- Lead impedance
- Related: Shock impedance, Lead dislodgement
The resistance measured along a pacing lead, typically 300 to 1000 ohms; values outside this range suggest fracture, insulation breach, or dislodgement.
- Lead dislodgement
- Related: Lead impedance, Pacing threshold
Movement of a lead tip away from its intended myocardial site, producing loss of capture, rising thresholds, or altered sensing on the affected channel.
- Pocket hematoma
- Related: Device erosion, Pocket infection
Collection of blood around the device pocket after implantation, which can cause wound breakdown, infection, or device erosion if not recognized and managed.
- Pocket infection
- Related: Pocket hematoma, Device erosion
Bacterial infection of the device pocket, often requiring system extraction and antibiotic therapy; signs include erythema, drainage, and fever.
- Device erosion
- Related: Pocket infection, Pocket hematoma
Gradual wearing of the device or lead through the skin, typically from pressure, thin tissue, or infection, requiring surgical revision or system removal.
- Venous access
- Related: Lead dislodgement, Pneumothorax
The route used to introduce pacing or ICD leads, commonly the subclavian, cephalic, or axillary vein, chosen based on anatomy and risk of lead crush.
- Pneumothorax
- Related: Venous access
Air in the pleural space, a recognized complication of subclavian venous access during device implantation, presenting with chest pain and dyspnea.
- Magnet application
- Related: Electromagnetic interference, Mode switching
Placing a magnet over a device to trigger a preset behavior, such as asynchronous pacing in pacemakers or suspension of tachycardia therapy in ICDs, for procedures or troubleshooting.
- Electromagnetic interference
- Related: Magnet application, Oversensing
External electrical noise from sources such as electrosurgery, MRI, or welding that can be oversensed by a device, causing inhibition, inappropriate pacing, or spurious shocks.
- MRI conditional device
- Related: Electromagnetic interference, Magnet application
A pacemaker or ICD system approved for scanning under specified magnetic field, lead configuration, and programming conditions, with post-scan interrogation and reprogramming required.
- Remote monitoring
- Related: Lead impedance, Tachycardia detection zones
Transmission of device diagnostics and arrhythmia data from home to a clinical center, enabling early detection of lead problems and clinical events between visits.
- Interrogation
- Related: Remote monitoring, Magnet application
Retrieval of stored device data and programmed parameters using a programmer, forming the basis for follow-up assessment and reprogramming decisions.