ISCD CBDT flashcards
ISCD CBDT · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Areal bone mineral density
- BMD reported by DXA in g/cm², calculated as bone mineral content divided by the projected two-dimensional bone area. It underestimates density in large bones and overestimates it in small bones.
- T-score formula
- Patient BMD minus the young adult reference mean, divided by the young adult reference standard deviation. A value of −2.5 or lower defines osteoporosis by WHO criteria.
- Z-score formula
- Patient BMD minus the age-matched reference mean, divided by the age-matched standard deviation. A Z-score of −2.0 or lower is below the expected range for age.
- WHO osteoporosis threshold
- A T-score at or below −2.5 at the lumbar spine, femoral neck, total hip, or 33% radius. Values between −1.0 and −2.5 are classified as low bone mass or osteopenia.
- Least significant change
- The smallest BMD difference between two measurements that is statistically real at 95% confidence, calculated as 2.77 times the precision error. Changes smaller than this are indistinguishable from noise.
- Precision versus accuracy
- Precision is reproducibility of repeated measurements, expressed as a coefficient of variation. Accuracy is closeness to the true value and depends on scanner calibration rather than technologist technique.
- Phantom quality control
- A spine phantom scanned daily before patients, with the measured BMD plotted on a control chart. Results outside the manufacturer's limits, commonly about ±1.5%, require correction before clinical scanning.
- Invalid vertebra rule
- A lumbar vertebra is excluded from analysis if its T-score differs by more than 1.0 unit from an adjacent evaluable vertebra, or if it has fracture, focal lesion, or hardware.
- Diagnostic hip regions
- Only the femoral neck and total hip are used for diagnosis and T-score classification. Ward's triangle and the greater trochanter are not diagnostic sites and are not used to define osteoporosis.
- 33% radius site
- The distal one-third forearm site, scanned on the non-dominant arm, used when spine and hip cannot be evaluated or measured. It is the preferred site in primary hyperparathyroidism.
- Vertebral fracture assessment
- A lateral DXA spine image used to detect morphometric vertebral deformity. A reduction of 20% or more in vertebral height defines a prevalent deformity.
- FRAX tool
- A calculator estimating 10-year probability of major osteoporotic fracture and hip fracture from clinical risk factors, optionally with femoral neck BMD. Treatment is often considered at 20% and 3% thresholds.
- DXA beam principle
- Two X-ray photon energies are used, and the ratio of low- to high-energy attenuation separates bone mineral from soft tissue. This eliminates the need for a water bath.
- Pencil versus fan beam
- Pencil-beam DXA uses one detector and a narrow beam, scanning slowly. Fan-beam DXA uses a detector array, scans faster but delivers slightly more dose and demands precise centering.
- Cross-calibration
- A phantom-based regression that converts BMD values when a patient moves between scanners or after major hardware changes. T-scores are never cross-calibrated, only BMD values.
- QCT versus DXA
- Quantitative CT measures true volumetric BMD in mg/cm³ and is less affected by degenerative artifact, but uses a higher dose and lacks a WHO T-score framework.
- Quantitative ultrasound
- Calcaneal QUS measures speed of sound and broadband ultrasound attenuation. It predicts fracture risk but cannot diagnose osteoporosis by WHO criteria or be used for monitoring.
- DXA effective dose
- Roughly 1–10 µSv per site, comparable to hours to a day of background radiation and far below a chest radiograph. Occupational shielding is not routinely required.
- Spine positioning
- Supine with hips and knees flexed over a block to flatten lumbar lordosis, mid-sagittal line centered, scanning from about T12 through L5. Correct positioning prevents falsely elevated BMD.
- Femur internal rotation
- The hip is scanned with the femur internally rotated 15–25 degrees so the femoral neck lies parallel to the table, giving the longest, most reproducible neck profile.
- Artifacts raising spine BMD
- Aortic calcification, osteophytes, facet sclerosis, and overlying metal increase measured aBMD. Laminectomy and surgical resection lower it. These can mask or mimic osteoporosis.
- Body composition model
- DXA partitions the field into bone mineral, fat mass, and lean soft tissue. The three-compartment model assumes constant lean-tissue hydration, so edema and dehydration bias fat and lean estimates.
- Fragility fracture
- A fracture resulting from a fall from standing height or less. A prior hip or spine fragility fracture raises future fracture risk independently of BMD.
- Secondary osteoporosis causes
- Conditions including glucocorticoid excess, hyperparathyroidism, hyperthyroidism, vitamin D deficiency, malabsorption, hypogonadism, and multiple myeloma. Basic screening includes calcium, vitamin D, PTH, and TSH.
- Glucocorticoid-induced osteoporosis
- Caused by prednisone at 5 mg/day or more for 3 months or longer. Fracture risk rises early, and BMD monitoring is recommended within 6 months of starting and then annually.
- Denosumab discontinuation risk
- Stopping denosumab causes rapid rebound bone loss and vertebral fracture risk. Therapy requires continued dosing or transition to another agent to preserve gains.
- Anabolic agents
- Teriparatide and abaloparatide are PTH-receptor anabolics that build bone; romosozumab is a sclerostin inhibitor with anabolic and antiresorptive effects. They are used for severe or refractory osteoporosis.
- Bisphosphonate drug holiday
- A planned pause in oral bisphosphonate therapy after roughly 3–5 years in patients whose fracture risk has fallen. It limits rare risks such as atypical femoral fracture and jaw osteonecrosis.
- Monitoring interval logic
- Follow-up intervals follow the least significant change and expected rate of loss. With 1% spine precision, a 3% annual loss becomes detectable in about 1–2 years; stable patients need 2 years or more.
- Root-mean-square precision
- A precision estimate pooling the standard deviations of repeated scans across multiple subjects, preferred over a single-subject SD because it better reflects true reproducibility error.
- Control chart rules
- Quality control charts flag drift when points exceed limits, for example two consecutive phantom measurements beyond 2 SD. Drift triggers recalibration or service before clinical scanning.
- Bilateral hip baseline
- Both hips are scanned at baseline because side-to-side BMD differences can exceed the least significant change. Follow-up scans use the same hip and same region of interest.