All guide sections

Physics, Instrumentation and Radiation Safety

The physics, instrumentation and radiation safety domain tests whether you can connect X-ray tube physics to image quality and to the dose a patient receives. Expect calculations using the inverse square law, half-value layer and CT dose indices, plus practical MRI and ultrasound safety.

Photon production and beam quality

  • Bremsstrahlung produces the continuous spectrum; characteristic radiation produces discrete K-shell peaks.
  • kVp controls beam energy and penetration; mAs controls photon number and dose approximately linearly.
  • Filtration removes low-energy photons, raises the half-value layer and lowers skin dose without changing the useful beam much.

Dose metrics and instrumentation

  • Absorbed dose is in gray (Gy); equivalent and effective doses are in sievert (Sv).
  • CTDIvol is scanner output measured in a phantom; DLP = CTDIvol × scan length. Effective dose is approximately DLP × k, with k about 0.014-0.015 mSv/(mGy·cm) for adult abdomen.
  • Pitch below 1 overlaps slices and raises dose for the same anatomy.

Worked example

An abdominal CT has DLP 600 mGy·cm. Effective dose = 600 × 0.015 = 9 mSv, roughly 3-4 years of natural background (about 2.4 mSv/year). If pitch is reduced from 1.2 to 0.8, dose rises by 1.2/0.8 = 1.5-fold to about 13.5 mSv, so only reduce pitch when fine detail genuinely requires it.

Traps

  • Treating CTDIvol as the patient's dose rather than a phantom index.
  • Forgetting that a lead apron does not shield against the primary beam.
  • Quoting MRI zone IV as safe for ferromagnetic implants; it is the scanner room and the highest-risk zone.