Engineer Chemical Analysis flashcards
Engineer Chemical Analysis · Build recall with a focused review session.
33 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Beer-Lambert law
- Absorbance A = ε·b·c, where ε is molar absorptivity, b is pathlength in cm, and c is molar concentration. Linear only at low concentration with monochromatic light and non-interacting analyte.
- Limit of detection
- Smallest analyte amount giving signal statistically distinguishable from blank, typically LOD = 3s0 where s0 is the standard deviation of blank or low-level replicate measurements.
- Limit of quantification
- Lowest concentration reliably measurable with acceptable precision and accuracy, commonly LOQ = 10s0. Must be validated with replicate measurements at that level.
- Traceability chain
- Unbroken sequence of calibrations, each with stated uncertainty, linking a measurement result to a certified reference material or SI unit. Any missing link breaks traceability.
- Gy's sampling theory
- Fundamental sampling error variance decreases as sample mass increases and particle size decreases. Required sample mass scales with the cube of nominal top particle size for heterogeneous material.
- Composite sampling
- Combining n increments into one sample reduces random sampling error by √n. Do not composite samples from systematically different strata or depth intervals.
- Chain of custody
- Chronological record of sample possession and handling: each transfer lists sample ID, date, time, custodian, condition, and purpose. Gaps undermine legal defensibility.
- Primary standard
- Highly pure, stable, non-hygroscopic reagent with high molar mass used to standardize a secondary solution, e.g., potassium hydrogen phthalate for base or potassium dichromate for oxidant.
- Gravimetric factor
- Ratio of analyte molar mass to weighed-form molar mass, multiplied by stoichiometric coefficients, converting mass of precipitate to mass of analyte.
- Equivalence versus endpoint
- Equivalence point is the stoichiometric completion of the reaction; endpoint is the observable signal change (indicator color). A good indicator makes these coincide.
- Karl Fischer titration
- Water-specific titration with iodine in the presence of sulfur dioxide and a base. Stoichiometric 1:1 water–iodine reaction; volumetric for high water, coulometric for trace water.
- EDTA complexometric titration
- EDTA binds most metal ions 1:1. Buffer pH controls selectivity (pH 10 for Ca/Mg, pH 5–6 for Zn). Masking agents or back-titration handle interfering metals.
- Stokes shift
- Difference between excitation and emission wavelengths in fluorescence. Emission occurs at longer wavelength because vibrational relaxation precedes photon emission.
- IR selection rule
- A molecular vibration absorbs infrared radiation only if the dipole moment changes during the vibration. Symmetric stretches of symmetric molecules are IR-inactive.
- Inner-filter effect
- At high absorbance, fluorescence intensity is no longer proportional to concentration because excitation and emitted light are absorbed. Keep A below about 0.05.
- Hollow cathode lamp
- Element-specific line source in atomic absorption spectroscopy. Emits narrow atomic lines of the analyte, giving high selectivity against background absorption.
- Internal standard
- Element or compound added at constant concentration to all standards and samples to correct for drift, matrix suppression, and transport variation in atomic and chromatographic methods.
- Polyatomic interference
- Ions formed from plasma gas and matrix, e.g., 40Ar16O on 56Fe in ICP-MS. Corrected with collision/reaction cells, interference equations, or matrix separation.
- Isotope dilution
- Adding a known amount of an enriched isotope and measuring the altered isotope ratio by ICP-MS. Provides highest accuracy because recovery cancels out.
- van Deemter equation
- H = A + B/u + C·u describes plate height as a function of linear velocity. Minimum H gives optimum velocity and maximum efficiency.
- Resolution in chromatography
- Rs = 2(tR2 − tR1)/(w1 + w2). Baseline resolution requires Rs ≥ 1.5. Increases with selectivity, efficiency, and retention factor.
- Retention factor
- k = (tR − t0)/t0, where t0 is the unretained marker time. Measures how strongly the analyte partitions into the stationary phase.
- Ion chromatography suppression
- A suppressor device reduces eluent background conductance while enhancing analyte conductance, enabling sensitive conductivity detection of anions or cations.
- Nernst equation
- E = E° − (RT/nF)·ln Q. At 25 °C this becomes E = E° − (0.05916/n)·log Q. Relates cell potential to reaction quotient.
- pH electrode slope
- Potential change per pH unit, theoretically 59.16 mV at 25 °C. Acceptable calibration slope is 95–105% of theoretical; lower slope indicates aging or contamination.
- Thermogravimetric analysis
- Measures mass change versus temperature under controlled atmosphere. Quantifies moisture, volatiles, decomposition steps, and ash residue.
- Differential scanning calorimetry
- Measures heat flow versus temperature. Determines melting point, glass transition, crystallinity, and reaction enthalpy. Calibrated with indium (mp 156.6 °C).
- Method validation
- Documented confirmation that a method meets requirements for its intended use: specificity, linearity, accuracy, precision, LOD/LOQ, robustness, and stability.
- Expanded uncertainty
- U = k·u_c, where u_c combines Type A and Type B contributions in quadrature. k = 2 gives approximately 95% coverage probability.
- Horwitz criterion
- Empirical relationship between acceptable relative standard deviation and analyte concentration. Predicts precision limits for interlaboratory studies across concentration ranges.
- ALCOA+ principles
- Data integrity criteria: Attributable, Legible, Contemporaneous, Original, Accurate, plus Complete, Consistent, Enduring, Available. Applied to all laboratory records and electronic data.
- Fume hood face velocity
- Air speed through the hood opening, typically 0.3–0.5 m/s (60–100 ft/min). Verified with a calibrated anemometer; low sash height maintains containment.
- Instrument qualification sequence
- Design Qualification, then Installation Qualification, Operational Qualification, and Performance Qualification. Each stage documented before routine use of analytical instrumentation.