CAQ-CVTS study materials
CAQ-CVTS study resources

CAQ-CVTS cheat sheet

CAQ-CVTS · Keep the key ideas within reach.

Quick reference

Vascular Physics and Instrumentation

  • Doppler equation: f_d = (2 · f_0 · v · cos θ) / c, where c ≈ 1540 m/s in soft tissue; the factor 2 accounts for the round trip to and from the reflector.

  • Cos 60° = 0.50, so a 60° Doppler angle halves the measured shift; errors in angle estimation grow steeply beyond 60°, which is why 60° or less is the standard threshold.

  • Velocity from shift: v = (f_d · c) / (2 · f_0 · cos θ); with a 5 MHz probe, 1 kHz of shift at 60° corresponds to roughly 15.4 cm/s.

  • Nyquist limit = PRF/2; aliasing occurs when the Doppler shift exceeds it, and it is corrected by increasing PRF, lowering transmit frequency, increasing the angle toward 60°, or using continuous wave.

  • Continuous wave Doppler has no PRF and therefore no aliasing, but it lacks depth discrimination; pulsed wave Doppler provides range gating at the cost of a Nyquist limit.

  • Resistive index RI = (PSV − EDV)/PSV; values above 0.70–0.75 in native renal arteries suggest increased parenchymal or downstream resistance.

  • Pulsatility index PI = (PSV − EDV)/TAMV, using time-averaged mean velocity rather than peak velocity.

  • Spectral broadening indicates a wide range of velocities within the sample volume and is seen with turbulence, sampling over a stenosis, or a large sample gate.

  • Wall filter removes low-frequency, high-amplitude signals from vessel wall motion; setting it too high can erase low-velocity flow such as in a nearly occluded vessel.

Quick reference

Cerebrovascular Testing and Interpretation

  • ICA criteria for ≥70% stenosis include PSV >230–250 cm/s, EDV >100 cm/s, and ICA/CCA peak systolic ratio >3.2–4.0, often with spectral broadening and post-stenotic turbulence.

  • A PSV below roughly 50 cm/s with a damped, tardus-parvus waveform suggests near-occlusion or distal severe disease rather than a mild lesion.

  • The vertebral artery is normally lower resistance than the CCA; loss of diastolic flow or a high-resistance vertebral signal can indicate subclavian steal or distal stenosis.

  • Ophthalmic artery flow direction reversal suggests severe ipsilateral ICA disease or external carotid collateral supply; assess with compression or indirect testing only when safe.

  • A normal CCA waveform is low resistance with continuous diastolic flow; an externalized, high-resistance CCA signal suggests distal ICA occlusion or severe disease.

  • Temporal window insonation at 50–60 mm depth typically identifies the MCA at 30–60 mm; mean flow velocities above 120 cm/s suggest vasospasm in the appropriate clinical setting.

  • Ratio-based interpretation reduces angle and equipment variability; always document the angle, sample depth, and the reference vessel used for the ratio.

Quick reference

Peripheral Arterial Testing and Interpretation

  • Ankle-brachial index ABI = highest ankle systolic pressure / highest brachial systolic pressure; normal is 1.00–1.40, and values above 1.40 suggest non-compressible calcified vessels.

  • ABI 0.90 or below is abnormal; 0.50–0.90 suggests claudication, and below 0.40 suggests critical limb ischemia when symptoms match.

  • Segmental pressures drop by more than 20–30 mmHg between adjacent cuffs to localize disease; a gradient across a cuff level indicates stenosis at or above that level.

  • Pulse volume recordings show a damped, rounded, or absent waveform distal to a hemodynamically significant stenosis; loss of the dicrotic notch is an early sign.

  • Toe-brachial index TBI above 0.70 is normal; TBI is preferred in diabetics and dialysis patients with medial calcinosis because toe vessels are less often calcified.

  • Exercise testing with post-exercise ankle pressures that fall by more than 20 mmHg, or fail to recover within 5 minutes, supports hemodynamically significant disease.

  • A uniphasic or biphasic waveform indicates damping; triphasic flow with a sharp systolic upstroke and reversal is normal in a resting limb artery.

  • Reactive hyperemia and cuff-induced hyperemia can unmask borderline disease; document the protocol and timing because values change rapidly after cuff release.

Quick reference

Peripheral Venous Testing and Interpretation

  • Normal lower extremity venous flow is phasic with respiration and augments with distal compression; absence of phasicity or augmentation suggests obstruction or valve failure.

  • Reflux is defined as reversed flow lasting more than 0.5 seconds in the common femoral and femoral veins, and more than 1.0 second in the popliteal and calf veins.

  • Perform the exam with the patient standing or reverse Trendelenburg for reflux, and supine or semi-upright with limb dependency for acute DVT assessment.

  • Acute DVT findings: hypoechoic, non-compressible vein with expanded caliber, loss of spontaneous flow, and continuous or absent phasicity.

  • Chronic DVT findings: echogenic, narrowed, or recanalized vein with collateral channels, thickened walls, and persistent partial compressibility.

  • Compression is the primary diagnostic maneuver; color Doppler alone can miss non-occlusive thrombus, so compress in short-axis increments along the vein.

  • Superficial venous thrombosis is assessed with compression and reflux mapping; document distance from the saphenofemoral junction because proximity affects management.

  • Calf vein evaluation is limited by depth and anatomy; isolated calf DVT is more common in symptomatic patients and warrants repeat study if initially negative.

Quick reference

Abdominal and Visceral Vascular Testing

  • Normal renal artery PSV is typically below 180–200 cm/s; values above 180–200 cm/s with post-stenotic turbulence suggest stenosis, and the renal-aortic ratio above 3.5 supports it.

  • The parvus-tardus waveform in the renal parenchyma, with a prolonged acceleration time above 0.07 seconds, suggests upstream stenosis or reduced perfusion.

  • The abdominal aorta is measured outer wall to outer wall in anteroposterior and transverse planes; an infrarenal diameter of 3.0 cm or more defines an aneurysm.

  • Aortic endograft surveillance requires measurement of the sac, detection of endoleaks with color and spectral Doppler, and assessment of graft limb patency.

  • Hepatic artery evaluation after transplant focuses on the resistive index, normally 0.55–0.80; a very low RI can reflect stenosis or arterioportal fistula.

  • Portal vein flow is normally hepatopetal with velocities around 15–30 cm/s; reversed flow suggests portal hypertension or a shunt.

  • Mesenteric ischemia evaluation includes peak systolic velocities in the SMA and celiac artery; elevated velocities with post-stenotic turbulence support stenosis.

Quick reference

Dialysis Access and Preoperative Vascular Mapping

  • Preoperative mapping assesses arterial inflow with segmental pressures and duplex, venous diameter and compressibility, and central venous patency by history or imaging.

  • A minimum venous diameter of about 2.0–2.5 mm with a tourniquet, and arterial diameter of about 2.0 mm, supports a usable autogenous fistula.

  • A mature arteriovenous fistula typically has flow above 500–600 mL/min; brachial artery flow volume is calculated as TAMV × cross-sectional area × 60.

  • Access stenosis is suggested by a greater than 50% diameter reduction on grayscale, a peak systolic velocity ratio above 2:1, and a doubling of the velocity at the stenosis versus the adjacent segment.

  • A thrill on palpation and a low-resistance, high-diastolic waveform in the feeding artery indicate a functioning access; loss of diastolic flow suggests outflow obstruction.

  • Steal syndrome is suggested by reversed flow in the artery distal to the anastomosis and symptoms of hand ischemia; document direction of flow in the distal radial or ulnar artery.

  • Pseudoaneurysm is diagnosed by a yin-yang color pattern and a neck with to-and-fro spectral flow; document size and neck dimensions.

Quick reference

Patient Care, Safety, and Professional Standards

  • ALARA applies to all Doppler modes: power, dwell time, and frame rate determine exposure, and Doppler exposes tissue to higher intensities than B-mode.

  • The mechanical index MI estimates cavitation risk and should be kept below 1.9; the thermal index TI should be kept at or below 1 where practical.

  • The recommended maximum output for ophthalmic and neonatal exams is lower; follow the on-screen MI and TI and reduce output when the indication allows.

  • Standard precautions require gloves and hand hygiene for all patients; add gown, mask, and eye protection when splash or body fluid exposure is anticipated.

  • Informed consent, patient identification with two identifiers, and documentation of the indication, findings, limitations, and comparison studies are required standards.

  • Ergonomic injury prevention includes adjusting bed height, using arm support, alternating scanning hands, and taking micro-breaks during long venous or mapping studies.

  • Professional scope and credentialing are governed by the interpreting physician and institutional policy; technologists should not issue final diagnoses outside their scope.