Tc-99m physical half-life
Tc-99m decays with a 6.02-hour physical half-life and emits a 140.5 keV gamma photon, which is well suited to gamma camera imaging.
Mo-99 breakthrough limit
Generator eluate must contain no more than 0.15 kBq of Mo-99 per MBq of Tc-99m (0.15 µCi/mCi); exceeding this limit prevents patient administration.
Effective half-life formula
Effective half-life combines physical and biological decay: 1/Teff = 1/Tphysical + 1/Tbiological. It determines how long activity remains in the patient.
Tc-99m MAA particle size
Tc-99m macroaggregated albumin particles range from 10 to 90 micrometers and lodge in pulmonary capillaries; typical doses contain 200,000 to 500,000 particles.
High-probability PE pattern
A high-probability ventilation/perfusion scan shows two or more large segmental perfusion defects with normal ventilation (mismatched defects), suggesting pulmonary embolism.
Meckel scan radiopharmaceutical
Tc-99m pertechnetate is used for Meckel diverticulum imaging because it is taken up by ectopic gastric mucosa; pentagastrin or H2 blockers can improve sensitivity.
Acute cholecystitis HIDA finding
On a hepatobiliary (HIDA) scan, nonvisualization of the gallbladder with normal hepatic uptake and biliary transit by 60 minutes is positive for acute cholecystitis.
GI bleed radiopharmaceuticals
Tc-99m sulfur colloid clears rapidly via the liver and detects about 0.1 mL/min bleeding; Tc-99m labeled RBCs allow longer imaging for intermittent bleeding.
Normal 24-hour thyroid uptake
Normal 24-hour thyroid uptake of I-123 is approximately 10 to 30 percent. Graves disease shows elevated uptake, while subacute thyroiditis shows low uptake.
Occupational dose limit
The occupational effective dose limit for radiation workers is 50 mSv (5 rem) per year; the public limit is 1 mSv per year.
Bone scan timing and dose
Adult Tc-99m MDP bone scans use about 740 to 925 MBq (20 to 25 mCi) intravenously, with imaging delayed 2 to 4 hours after injection.
Sentinel node injection site
For breast lymphoscintigraphy, Tc-99m sulfur colloid is injected peritumorally or periareolarly; imaging and a gamma probe identify the first draining node.
Radioactive decay equation
Activity remaining after time t is A = A0 × e^(−0.693t/T½), where T½ is the physical half-life of the radionuclide.
Collimator selection rule
Use a low-energy collimator for photons below about 150 keV, medium-energy for 150 to 300 keV, and high-energy above 300 keV to limit septal penetration.
Photopeak energy window
A typical gamma camera energy window is ±10 percent around the photopeak; for Tc-99m at 140.5 keV this is roughly 126 to 154 keV.
Hounsfield unit formula
CT number in Hounsfield units equals 1000 × (µtissue − µwater)/µwater, so water is 0, air about −1000, and dense bone about +1000.
Window width versus level
Window width controls displayed contrast; a narrow width increases contrast. Window level sets the center gray value of the displayed range.
Pitch definition
Pitch equals table travel per gantry rotation divided by total collimated beam width. Pitch above 1 reduces dose and scan time.
CTDIvol meaning
CTDIvol estimates radiation dose per unit volume in milligray for a given scan protocol and is displayed on the scanner dose report.
Dose-length product formula
DLP equals CTDIvol multiplied by scan length in centimeters, expressed in mGy·cm; multiplying DLP by a body region factor estimates effective dose.
Bone window settings
Bone windows use a wide width near 2000 HU and a high level near 400 to 500 HU to display high-attenuation osseous structures without clipping.
Tube current modulation
Automatic exposure control varies tube current with patient attenuation along the scan, lowering dose while maintaining a target image noise level.
Iterative reconstruction effect
Iterative reconstruction reduces image noise compared with filtered back projection, permitting lower dose, but can change image texture and low-contrast detectability.
Beam hardening artifact
Beam hardening produces dark streaks or cupping because lower-energy photons are preferentially absorbed; higher kVp and metal artifact reduction help reduce it.
Photon starvation artifact
Photon starvation occurs when very few photons reach detectors through high-attenuation regions such as shoulders, producing prominent streak artifacts across the image.
Multidetector row advantage
Multiple detector rows allow thinner acquisition slices and faster table speed, improving z-axis resolution and enabling high-quality multiplanar reconstructions.
Bolus tracking purpose
Bolus tracking monitors contrast arrival in a target vessel and triggers the scan at a preset threshold, optimizing arterial or venous phase timing.
Iodinated contrast risk factors
Reduced renal function (low eGFR), prior contrast reaction, and dehydration increase risk with iodinated contrast; renal function should be checked before elective contrast CT.
CT effective dose estimate
Effective dose in CT is estimated by multiplying DLP by a region-specific conversion factor, typically about 0.015 mSv/mGy·cm for the body.
Reconstruction slice thickness
Reconstructed slice thickness can be thinner than the acquired detector width; overlapping thin reconstructions improve multiplanar and three-dimensional image quality.
CT number of water
Water is assigned 0 HU by definition, providing the reference point for the Hounsfield scale used to characterize tissue attenuation.