NHPC BMLT: Quality Control and Laboratory Management

NHPC BMLT · Work through your exam, one topic at a time.

All guide sections

Quality Control and Laboratory Management

Quality control and laboratory management tests whether you can prove a result is trustworthy and run a laboratory safely. Expect internal quality control, external quality assessment, Westgard rules, documentation, biosafety and audit.

Core rules

  • Accuracy is closeness to the true value; precision is closeness of repeated results. Bias is systematic error; imprecision is random error.
  • Plot daily controls on a Levey–Jennings chart. Warning rule 1-2s; rejection rules include 1-3s, 2-2s, R-4s, 4-1s and 10x.
  • Calibration establishes the relationship between signal and concentration; quality control verifies it daily. Never adjust a control range to fit a failing run.
  • External quality assessment or proficiency testing compares your results with peer laboratories and is not the same as internal quality control.
  • Documentation: standard operating procedures, records of maintenance, corrective action and staff competency.
  • Biosafety: standard precautions, hand hygiene, PPE, safe sharps disposal, spill kits and correct waste segregation.

Worked example

A chemistry control gives 1-3s on the morning run. The rule is violated because one point lies beyond three standard deviations. Stop reporting patient results, repeat the control with a fresh aliquot, and check reagents and calibration. If the repeat is in control, the first value was probably random error; if not, recalibrate and re-run the patient samples from that period. Document the investigation and the corrective action.

Traps

Releasing results during a rejected QC run; accepting external quality assessment results without investigating an unsatisfactory score; and signing off competency for staff who have never been directly observed.