PRC RadTech flashcards
PRC RadTech · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Bremsstrahlung
- Continuous x-ray spectrum produced when electrons decelerate in the target nucleus field; maximum photon energy equals kVp.
- Characteristic radiation
- Discrete x-rays emitted when a projectile electron ejects an inner-shell electron and an outer-shell electron fills the vacancy.
- Half-value layer
- Thickness of material that reduces x-ray intensity to half; increases with kVp and filtration; measures beam quality.
- Inverse square law
- Intensity is inversely proportional to the square of distance: I1/I2 = (d2/d1)^2; doubling distance reduces intensity to one quarter.
- Photoelectric effect
- Complete absorption of an x-ray photon by an inner-shell electron; dominates at low kVp and high atomic number; produces high contrast.
- Anode heel effect
- X-ray intensity is lower toward the anode side of the beam; place thicker body parts toward the cathode for uniform exposure.
- Focusing cup
- Negatively charged metal cup around the filament that confines the electron beam to a small area on the anode.
- Line-focus principle
- Angled anode target creates a smaller effective focal spot than actual focal spot, improving resolution while spreading heat.
- Automatic exposure control
- Ionization chambers measure radiation reaching the receptor and terminate exposure at a preset level, controlling receptor exposure.
- Tube rating chart
- Graph showing maximum kVp, mA, and exposure time combinations for a focal spot; prevents anode overheating and tube damage.
- PA chest radiograph
- Upright projection at 72 inches with scapulae rotated clear of lungs; reduces heart magnification and thyroid dose.
- KUB abdomen
- Supine AP abdomen radiograph at 40 inches showing kidneys, ureters, and bladder; no contrast medium is used.
- Odontoid view
- Open-mouth AP projection of the upper cervical spine showing C1 and C2, including the dens and lateral masses.
- Anatomic position
- Upright, facing forward, arms at sides with palms forward; reference position for describing body planes and directions.
- Epiphyseal plate
- Cartilaginous growth plate between epiphysis and metaphysis in immature long bones; appears as a radiolucent line.
- Barium sulfate
- Positive oral contrast agent used for GI studies; contraindicated if bowel perforation is suspected because it causes peritonitis.
- 15 percent rule
- Increase kVp by 15% to halve mAs, or decrease kVp by 15% to double mAs, maintaining receptor exposure.
- Magnification
- Size distortion calculated as SID divided by SOD; reduced by increasing SID or decreasing object-to-image receptor distance.
- Grid ratio
- Ratio of height to interspace width of grid strips; higher ratio removes more scatter and requires more mAs.
- Quantum mottle
- Noise from too few x-ray photons reaching the receptor; reduced by increasing mAs or using a faster receptor.
- Exposure indicator
- Digital value reflecting receptor exposure; deviates from manufacturer's target when technique is outside optimal range.
- Modulation transfer function
- Measure of how well an imaging system preserves spatial resolution at different spatial frequencies; higher MTF indicates sharper image detail.
- ALARA
- Principle of keeping radiation dose as low as reasonably achievable, considering economic and social factors.
- Occupational dose limit
- Effective dose limit for radiation workers is 50 mSv per year; lens limit 150 mSv and extremity limit 500 mSv.
- Controlled area
- Area where occupational exposure may occur, posted with radiation warning signs and access limited to authorized personnel.
- Lead apron equivalence
- Protective apron must be at least 0.25 mm lead equivalent; 0.5 mm is recommended for fluoroscopy and high-dose procedures.
- Declared pregnant worker
- Worker who voluntarily declares pregnancy; fetal dose limit is 0.5 mSv per month and 5 mSv over gestation.
- Informed consent
- Patient's voluntary agreement after receiving information about procedure, risks, benefits, and alternatives; must be documented.
- Patient confidentiality
- Duty to protect patient health information from unauthorized disclosure, consistent with privacy laws and professional codes.
- Scope of practice
- Legal boundaries of the radiologic technologist role: perform imaging and related procedures under supervision, not diagnose.
- Beneficence
- Ethical principle of acting for the patient's benefit and promoting good outcomes while minimizing harm.
- Professional boundary
- Limits that protect the therapeutic relationship; avoid personal, financial, or romantic involvement that could exploit patients.