Low Platelets on Every Feline Sample

VLE · Put your knowledge to work in context.

All case studies

Low Platelets on Every Feline Sample

During a junior laboratory techniques course, five EDTA blood samples from healthy cats are run on an impedance hematology analyzer. All five return platelet counts between 40 and 70 x 10^9/L; the analyzer flags platelet clumps on only one. The instructor's manual estimate on a smear from the first sample gives 280 x 10^9/L. A diluent background count run at 1:00 pm shows WBC 0.3 x 10^9/L, RBC 0.06 x 10^12/L, and platelets 28 x 10^9/L. Morning quality control was within limits at 8:00 am. The reagent pack was changed two weeks ago, the waste container is nearly full, and the peristaltic pump tubing is 18 months old; the analyzer has no service contract and the last preventive maintenance was 11 months ago. Samples were collected at 8:00 am and run at 1:00 pm; one sample is visibly clotted. Students have two hours left and ask whether they should record the analyzer values or their manual counts.

1. What is your differential diagnosis for the low platelet counts? Rank analyzer, preanalytical, and biologic causes, and say what data support each.

Read the worked answer

Four categories. Analyzer: the elevated background (platelets 28, WBC 0.3, RBC 0.06) indicates contaminated diluent, a dirty flow cell, or a partially clogged aperture; low platelet counts on every sample, including healthy cats, point to the instrument. Preanalytical: feline platelets clump in EDTA, counts fall with time (five hours here), and one sample is clotted; however, clumping would not usually affect all five healthy cats uniformly, and the analyzer flagged only one. Biologic: cats can have low platelets, but five healthy cats is implausible. Method: impedance counters can misclassify large platelets and microcytes; the manual estimate of 280 on sample one contradicts the analyzer.

The clotted sample is invalid for any cell count and must be recollected or excluded. The key discriminator is the background count: a properly functioning analyzer should give near-zero background. Until the background is clean, every platelet result is suspect regardless of the flag. Morning QC passing at 8:00 am does not exonerate the instrument at 1:00 pm; a clog or reagent problem can develop during the day.

2. Give an ordered troubleshooting plan for the next two hours, including decision points and what you would release to the students' reports.

Read the worked answer

Order: (1) stop reporting analyzer platelet counts and label them pending verification; (2) repeat the background in duplicate with a new diluent container; if still elevated, the fault is in the fluidics or aperture; (3) clean the aperture and flow cell per manufacturer instructions, backflush, and check the waste container, because a full waste container can cause backpressure and erroneous counts; (4) inspect the peristaltic tubing for wear; 18 months is old, and a worn tube can aspirate variable volumes, so replace it if collapsed or cracked; (5) repeat the background and confirm it is within manufacturer limits before proceeding; (6) run both QC levels; (7) re-run the samples, and for each make a manual estimate on a fresh smear from a non-clotted sample; (8) compare analyzer and manual values. If they agree within the laboratory's acceptance limit, release with a comment; if not, use the manual estimate and report it.

The clotted sample must be recollected. Do not release values just because students need to finish; teach them that the two hours are for resolution, not for inventing numbers.

3. What preventive maintenance and training changes would you put in place, and how would you justify them with a limited budget?

Read the worked answer

Preventive maintenance: daily background and QC, weekly aperture cleaning, monthly fluidics check, tubing replacement on a defined interval (for example every 12 months or per manufacturer), a waste alarm, and a maintenance log. Justify with cost: a service call is expensive, but invalid results and repeated recollections cost more in staff time and client trust; a single misreported platelet count in a bleeding patient is a safety event. Budget ask: a service contract covering two preventive visits and priority response, plus a spare tubing kit and a validated manual counting method as backup.

Training: students perform and document daily background and QC as part of the rotation, with sign-off. Competency check: each student demonstrates a manual platelet estimate and explains how it differs from the analyzer. Place a troubleshooting decision tree at the bench. Track downtime and repeat rates for six months and use the data to justify renewal. If the budget is truly limited, prioritize the service contract over new instruments, because the current analyzer is usable if maintained.