- Half-cell potential
- The DC voltage developed at an electrode-electrolyte interface relative to a reference; Ag/AgCl is stable and low-noise, while polarizable metals drift, causing baseline artifacts in biopotential recordings.
- Nernst equation
- Relates equilibrium potential to ion concentration ratio: E = E0 - (RT/zF)ln(Q). At body temperature a tenfold monovalent gradient gives about 61 mV.
- Goldman-Hodgkin-Katz equation
- Computes resting membrane potential from all permeant ions weighted by permeability, explaining why the resting potential sits near but not exactly at the potassium equilibrium potential.
- Hodgkin-Huxley model
- Quantitative membrane model using voltage-gated sodium (m, h) and potassium (n) gating variables to reproduce action potential threshold, upstroke, repolarization, and refractory periods.
- Instrumentation amplifier CMRR
- Common-mode rejection ratio expresses differential gain divided by common-mode gain; high CMRR (80-100 dB) is essential to extract millivolt biopotentials from volt-level mains interference.
- Driven-right-leg circuit
- An active circuit that senses common-mode voltage and injects an inverted signal into the patient, reducing common-mode interference and improving biopotential amplifier performance.
- Nyquist sampling theorem
- Sampling must exceed twice the highest signal frequency to avoid aliasing; anti-aliasing filters precede the ADC, and diagnostic ECG needs at least 300 Hz sampling.
- Microshock versus macroshock
- Macroshock passes through intact skin and needs roughly 100 mA to fibrillate; microshock through intracardiac catheters can fibrillate at about 100 µA, driving type CF limits.
- Pennes bioheat equation
- Models tissue temperature as conduction plus blood perfusion cooling plus metabolic heat, used to plan RF ablation, hyperthermia, and thermal safety margins.
- Stress and strain
- Stress is force per area and strain is fractional length change; their slope in the linear region is Young's modulus, which characterizes tissue and implant stiffness.
- Viscoelastic tissue behavior
- Tissues show time-dependent response: creep under constant load and stress relaxation under constant strain, modeled with Maxwell or Kelvin-Voigt spring-dashpot combinations.
- MRI SAR limit
- Specific absorption rate quantifies RF heating; whole-body limits near 4 W/kg over six minutes constrain sequence design, alongside projectile and peripheral nerve stimulation hazards.
- Ultrasound axial resolution
- Axial resolution equals half the spatial pulse length; fewer cycles and higher frequency shorten the pulse, improving resolution at the cost of penetration depth.
- CTDIvol and DLP
- CTDIvol measures dose per rotation in mGy; DLP equals CTDIvol times scan length, and effective dose is estimated by multiplying DLP by a body-region conversion factor.
- Half-value layer
- The thickness of material that reduces beam intensity by half, equal to ln2 divided by the linear attenuation coefficient; it characterizes beam quality and filtration.
- Photoelectric versus Compton
- Photoelectric absorption dominates at low energies and high atomic number, giving contrast; Compton scatter dominates in soft tissue at diagnostic energies and degrades image contrast.
- Pacemaker lower rate limit
- The programmed minimum pacing rate; its escape interval equals 60 divided by the rate in seconds, and sensed intrinsic events reset the timing cycle in inhibit modes.
- Biphasic defibrillation waveform
- A two-phase shock, typically 120-200 J, that lowers defibrillation threshold versus monophasic shocks; impedance compensation adjusts duration to deliver consistent current.
- Steam sterilization parameters
- Autoclave cycles use saturated steam at 121 °C for 15 minutes or 134 °C for 3-4 minutes; heat-sensitive polymers instead require ethylene oxide or ionizing radiation.
- 510(k) premarket notification
- The US pathway for Class II devices demonstrating substantial equivalence to a legally marketed predicate; Class III devices generally need the stricter PMA route.
- EU MDR device classes
- Devices are classified I, IIa, IIb, or III by risk and invasiveness; all but low-risk Class I require Notified Body conformity assessment and CE marking.
- ISO 10993 biocompatibility testing
- Test selection depends on contact nature (surface, external communicating, implant) and duration; cytotoxicity, irritation, and sensitization form the initial screening battery.
- Type CF applied part
- A cardiac-floating patient connection permitting the strictest leakage limit, about 10 µA, because intracardiac current paths carry microshock fibrillation risk.
- Signal-to-noise ratio
- Ratio of signal power to noise power; in biomedical measurement it improves with electrode contact quality, averaging, filtering, and amplifier input impedance matching.
- Electrode-skin impedance
- Contact impedance at the electrode interface; high impedance raises thermal noise and common-mode pickup, so skin preparation and conductive gel are used to lower it.
- Wiener filtering
- An optimal linear filter minimizing mean-squared error for known signal and noise spectra, used to suppress noise in ECG, EEG, and other biosignals.
- Aliasing
- The folding of frequencies above half the sampling rate into lower frequencies; prevented by analog anti-aliasing filters placed before the analog-to-digital converter.
- Pulse oximetry principle
- Uses differential light absorption of oxygenated and deoxygenated hemoglobin at red and infrared wavelengths, with the pulsatile AC component isolating arterial saturation.
- Beer-Lambert law
- Attenuation is exponential in absorber concentration, path length, and extinction coefficient; underpins pulse oximetry, spectrophotometry, and optical biosensors.
- Biosensor transducer types
- Amperometric, potentiometric, optical, and piezoelectric transducers convert biorecognition events into measurable signals; glucose oxidase amperometric sensors are a common example.
- Fick principle
- Cardiac output equals oxygen consumption divided by the arteriovenous oxygen content difference; it also underlies indicator-dilution and mass-balance measurements.
- Windkessel model
- A lumped-parameter model of arterial load using resistance and compliance elements to relate aortic pressure and flow, capturing pulsatile hemodynamics beyond simple resistance.