RPFT study materials
RPFT study resources

RPFT cheat sheet

RPFT · Keep the key ideas within reach.

Quick reference

Patient Assessment and Preparation

  • Pre-test assessment covers the indication, contraindications, current medications, recent respiratory illness, smoking status, and the patient's ability to perform forced maneuvers.

  • Withhold short-acting beta-agonists for 4–6 hours, long-acting beta-agonists for 12–24 hours, and short-acting anticholinergics for 6–8 hours unless testing is intended to assess current bronchodilator therapy.

  • Record age, sex, standing height without shoes, weight, and race/ethnicity; these variables are inputs to reference equations for predicted values and lower limits of normal.

  • Absolute contraindications include recent myocardial infarction, unstable angina, pneumothorax, active hemoptysis, and recent thoracic or abdominal surgery.

  • Relative contraindications include resting tachycardia, marked hypertension, recent respiratory infection, and inability to cooperate; document the risk-benefit decision.

  • Avoid vigorous exercise for at least 30 minutes and smoking for at least 1 hour before testing because either can alter airway caliber and test results.

  • Explain and demonstrate the forced maneuver; use a sitting position, nose clip, and tight mouth seal to prevent air leak.

  • Document supplemental oxygen use, recent hospitalization, and baseline dyspnea so results can be interpreted in clinical context.

Quick reference

Equipment Quality Control and Calibration

  • Daily spirometer volume check: use a calibrated 3-L syringe; measured volume should be within ±3.5% of the syringe volume and the result logged.

  • Convert spirometry volumes from ATPS to BTPS using ambient temperature, barometric pressure, and water-vapor pressure.

  • Perform daily leak and occlusion checks; a leak falsely lowers measured volumes and flows and can invalidate results.

  • Calibrate DLCO gas analyzers daily with zero and span gases and verify linearity before patient testing.

  • Calibrate plethysmograph pressure transducers and box volume daily using known pressure and volume signals.

  • Record date, device, calibrator identity, expected value, measured value, and corrective action in the quality-control log.

  • Biological control testing by a healthy staff member helps detect system drift that syringe checks may not reveal; compare results with established limits.

  • Verify gas cylinder identity and expiration dates and confirm carbon monoxide and tracer gas concentrations before use.

Quick reference

Pulmonary Function Testing Procedures

  • An acceptable FVC trial has a rapid start, no cough in the first second, no glottic closure or leak, and a complete exhalation without early termination.

  • Back-extrapolated volume should be less than 5% of FVC or 150 mL, whichever is greater, indicating a satisfactory rapid start.

  • End-of-test criteria: a plateau with less than 25 mL change in the final second, or an exhalation lasting at least 6 seconds and generally no more than 15 seconds.

  • Reproducibility: the two largest FVC values and the two largest FEV1 values should agree within 150 mL; obtain at least three acceptable trials.

  • Bronchodilator response testing: repeat spirometry 10–15 minutes after a short-acting beta-agonist; a significant response is an increase in FEV1 or FVC of at least 12% and 200 mL.

  • Single-breath DLCO: exhale to residual volume, inhale test gas rapidly to total lung capacity, hold about 10 seconds, then exhale smoothly while discarding dead-space gas.

  • Plethysmography: with the shutter closed, the patient pants at about 1 Hz while box pressure changes are recorded to calculate thoracic gas volume.

  • MVV: the patient breathes maximally and rapidly for 12–15 seconds; the measured volume is extrapolated to liters per minute.

Quick reference

Test Interpretation and Reporting

  • Obstruction is identified by an FEV1/FVC ratio below the lower limit of normal, or below 0.70 when the fixed criterion is used.

  • Restriction is suggested by reduced FVC with a normal or high FEV1/FVC ratio and is confirmed by total lung capacity below the lower limit of normal.

  • A mixed defect shows both a reduced FEV1/FVC ratio and a reduced total lung capacity.

  • In COPD, FEV1 percent predicted defines mild (>70%), moderate (60–69%), moderately severe (50–59%), severe (35–49%), and very severe (<35%) airflow limitation.

  • Bronchodilator responsiveness is an increase in FEV1 or FVC of at least 12% and at least 200 mL from baseline.

  • A low DLCO suggests emphysema, interstitial lung disease, pulmonary vascular disease, or anemia; an elevated DLCO may occur with polycythemia or pulmonary hemorrhage.

  • Use the lower limit of normal, typically the fifth percentile of the reference population, rather than a fixed percent predicted cutoff when reference equations permit.

  • Flow-volume loop: obstructive disease produces a concave expiratory limb, while restriction produces a narrow loop with a preserved shape.

Quick reference

Safety and Infection Control

  • Perform hand hygiene before and after patient contact and after handling mouthpieces or filters; wear gloves when contact with respiratory secretions is anticipated.

  • Use a disposable mouthpiece and a bacterial/viral filter for each patient; never reuse single-use items.

  • Clean and disinfect reusable spirometry and plethysmography components according to manufacturer instructions, including adequate contact time and drying before reuse.

  • For suspected tuberculosis, use a bacterial/viral filter, schedule testing in a separate room, and follow airborne infection isolation and local public health guidance.

  • For blood or body-fluid spills, absorb the material, disinfect with an EPA-registered agent, and discard waste according to facility policy.

  • Secure oxygen cylinders and keep them away from heat or open flames; prohibit smoking in testing areas because oxygen-enriched environments increase fire risk.

  • Dispose of sharps in puncture-resistant containers and biohazard waste in labeled containers according to facility policy.

  • Clean and disinfect surfaces between patients and follow posted infection-control policies for equipment and environmental cleaning.