Quick reference

Cerebrovascular

  • An ICA peak systolic velocity of 125 cm/s or greater together with an ICA/CCA velocity ratio of 2.0 or greater identifies a 50% or greater diameter stenosis under the SRU consensus criteria.

  • NASCET stenosis (%) = (1 - residual lumen diameter / normal distal ICA diameter) x 100.

  • ICA/CCA ratio = peak systolic velocity in the ICA divided by peak systolic velocity in the distal (mid) common carotid artery.

  • A normal ICA spectral waveform is low resistance, with a broad systolic peak and continuous antegrade diastolic flow.

  • PSV of 230 cm/s or greater, or end diastolic velocity of 100 cm/s or greater, in the ICA suggests 80% or greater stenosis; a thread-like residual channel indicates near occlusion.

  • Reversed ophthalmic artery flow is an indirect sign of hemodynamically significant ipsilateral ICA stenosis or occlusion.

  • Retrograde vertebral artery flow, especially after ipsilateral arm exercise, supports subclavian steal.

  • Echolucent plaque carries a higher embolic risk than densely calcified plaque; plaque surface ulceration is likewise associated with increased stroke risk.

Quick reference

Peripheral Arterial

  • ABI = highest ankle systolic pressure (dorsalis pedis or posterior tibial) divided by the higher of the two brachial systolic pressures.

  • ABI interpretation: above 1.30 non-compressible; 1.00-1.30 normal; 0.91-0.99 borderline; 0.90 or less abnormal.

  • A triphasic waveform is normal in limb arteries; monophasic, dampened flow indicates a proximal stenosis or collateral-dependent segment.

  • A segmental pressure gradient greater than 20-30 mmHg between adjacent cuffs localizes a hemodynamically significant lesion to that segment.

  • A peak systolic velocity ratio of 2.0 or greater across a stenosis corresponds to roughly 50% or greater diameter reduction, with higher ratios indicating tighter lesions.

  • Toe-brachial index above 0.70 is normal and is preferred when medial calcinosis makes ankle pressures falsely elevated.

  • Poiseuille relationship: volume flow varies with the fourth power of radius, so a 50% diameter reduction lowers flow roughly sixteen-fold at constant pressure.

  • A fall in ABI greater than 20% from baseline after treadmill exercise indicates hemodynamically significant occult disease.

Quick reference

Peripheral Venous

  • Normal venous Doppler is spontaneous, phasic with respiration, augments with distal compression, and ceases or briefly reverses with Valsalva.

  • Deep vein thrombosis criteria: non-compressibility, intraluminal echogenicity, absent spontaneous flow, and loss of respiratory phasicity.

  • Reflux thresholds: greater than 0.5 second in superficial and perforating veins; greater than 1.0 second in common femoral and popliteal veins.

  • Acute thrombus is hypoechoic with a distended vein; chronic thrombus is echogenic, the vein is small, and collaterals have formed.

  • Continuous, non-phasic femoral venous flow suggests obstruction proximal to the sampling site.

  • Blunted or absent augmentation after distal compression indicates obstruction between the compression point and the sample volume.

  • Reflux in the great saphenous vein is assessed with the patient standing and the limb dependent, sampling at the saphenofemoral junction and along the thigh.

  • Calf veins are paired (posterior tibial and peroneal); soleal and gastrocnemius muscular sinuses are not paired and require careful compressibility assessment.

Quick reference

Abdominal and Visceral Vascular

  • Infrarenal aortic diameter of 3.0 cm or more defines an aneurysm; elective repair is generally considered at 5.5 cm in men and 5.0 cm in women, or with growth over about 1 cm per year.

  • Renal artery PSV below 180 cm/s is normal; 180-200 cm/s or greater with a renal-aortic ratio of 3.5 or more suggests 60% or greater stenosis.

  • Renal-aortic ratio = renal artery peak systolic velocity divided by abdominal aortic peak systolic velocity.

  • A tardus-parvus waveform (slow upstroke, low peak velocity) distal to a stenosis indicates hemodynamically significant obstruction.

  • Normal hepatic artery resistive index is approximately 0.55-0.70; an index below 0.5 after transplantation raises concern for stenosis or thrombosis.

  • Portal hypertension is suggested by main portal vein diameter above 13 mm with mean velocity below about 15 cm/s (normal 15-30 cm/s).

  • Superior mesenteric artery PSV above 275 cm/s suggests 70% or greater stenosis; the normal fasting SMA waveform is high resistance with low diastolic flow.

  • Endograft surveillance evaluates sac diameter, endoleaks, and graft limb stenosis; any perfused endoleak with sac enlargement requires intervention planning.

Quick reference

Instrumentation

  • Doppler equation: frequency shift = (2 x transmitted frequency x velocity x cosine of Doppler angle) / speed of sound in tissue (about 1540 m/s).

  • Maintain a Doppler angle of 60 degrees or less, ideally 45-60 degrees, because cosine error grows rapidly as the angle approaches 90 degrees.

  • Nyquist limit = pulse repetition frequency / 2; velocities above this value alias.

  • Aliasing corrections: increase pulse repetition frequency or scale, lower transducer frequency, adjust baseline, increase Doppler angle, or switch to continuous wave Doppler.

  • Continuous wave Doppler has no depth resolution; pulsed Doppler with a sample volume provides range gating.

  • A higher transmitted frequency improves spatial resolution but increases attenuation and reduces penetration depth.

  • The wall filter is a high-pass filter; set too high it removes low-velocity flow such as venous or post-stenotic flow.

  • Power Doppler is angle-independent, does not alias, and detects slow flow, but provides no direction or velocity information.

Quick reference

Quality Assurance and Specialized Vascular Testing

  • Routine QA includes daily image uniformity checks, vertical and horizontal distance accuracy within about 2% or 1 mm, and an annual comprehensive system evaluation.

  • Velocity measured in a calibrated flow phantom should agree with the true value within approximately 10%.

  • Tissue-mimicking phantoms assess dead zone, axial and lateral resolution, penetration depth, and gray-scale uniformity.

  • Lindegaard ratio = middle cerebral artery mean velocity divided by ipsilateral extracranial internal carotid velocity; above 3.0 indicates vasospasm rather than hyperemia.

  • Transcranial Doppler: middle cerebral artery mean velocity above 120 cm/s suggests vasospasm; normal mean velocity is roughly 40-70 cm/s.

  • Treadmill protocol: 2 mph at 12% grade for up to 5 minutes, with ankle-brachial indices obtained before and after walking.

  • Dialysis access surveillance: access flow below 500-600 mL/min or an intra-access peak systolic velocity ratio above 2 suggests significant stenosis.

  • Bioeffects are managed with ALARA, monitoring thermal index and mechanical index, and minimizing dwell time at high output.