ABC-BIO flashcards
ABC-BIO · Build recall with a focused review session.
33 cards. Read the term and the definition here. The deck at the top of this page is optional.
- LLOQ acceptance criteria
- The lowest calibrator with signal-to-noise of at least 5, accuracy within plus or minus 20 percent of nominal, and precision CV no greater than 20 percent across runs.
- 4PL calibration curve
- A four-parameter logistic model fitting signal against log concentration, with parameters for minimum, maximum, slope, and inflection point; standard for immunoassays.
- Stable-isotope internal standard
- A labeled analog of the analyte differing only in mass; it co-elutes and corrects for extraction losses and ionization variability in LC-MS/MS quantification.
- Matrix effect assessment
- Comparison of post-extraction spiked blank response to neat solvent response at the same concentration; ion suppression or enhancement is expressed as a percentage ratio.
- Carryover acceptance
- Residual analyte signal in a blank injected immediately after a high-concentration sample must not exceed 20 percent of the LLOQ response.
- Dilution integrity test
- A QC sample above the upper limit of quantification is diluted and re-assayed; measured accuracy must fall within plus or minus 15 percent of nominal.
- Selectivity in bioanalysis
- Demonstrated by assaying blank matrix from at least six individual lots; interference at the analyte retention time must be under 20 percent of LLOQ.
- Extraction recovery
- Ratio of analyte response from an extracted spiked sample to response from a post-extraction spiked blank at equal concentration, expressed as a percentage.
- Calibration standard acceptance
- At least 75 percent of standards and 67 percent of each concentration level must be within plus or minus 15 percent, or 20 percent at LLOQ.
- Solid-phase extraction principle
- Analytes partition between a liquid sample and a solid sorbent; conditioning, loading, washing, and elution steps yield selective cleanup and concentration.
- Protein precipitation
- Addition of organic solvent, acid, or salt denatures matrix proteins, releasing bound analyte; fast but yields dirtier extracts than SPE.
- ELISA sandwich format
- Capture antibody immobilizes antigen, a detection antibody conjugated to enzyme generates signal proportional to antigen concentration; suited to large analytes.
- Surrogate analyte strategy
- When the true analyte is unstable or unavailable, a stable structural analog is quantified and results are reported relative to the surrogate.
- Run acceptance batch rule
- A batch passes only if at least 67 percent of QCs and 50 percent at each level are within plus or minus 15 percent of nominal.
- Internal standard response variability
- Peak-area ratios of analyte to internal standard are tracked per injection; IS response outside a defined window flags extraction or injection failure.
- Beer-Lambert law
- Absorbance equals molar absorptivity times path length times concentration; linearity fails at high absorbance or with polychromatic or stray light.
- Chromatographic resolution formula
- Resolution equals two times the difference in retention times divided by the sum of the two peak widths at baseline.
- Baseline resolution criterion
- A resolution value of at least 1.5 separates two peaks with essentially complete return to baseline between them.
- Van Deemter equation
- Plate height equals the eddy diffusion term plus longitudinal diffusion over velocity plus mass transfer times velocity; minimum height defines optimum flow.
- Retention factor k
- k equals retention time minus void time, divided by void time; it expresses how much longer a solute is retained than an unretained marker.
- Selectivity factor alpha
- Alpha equals the retention factor of the later-eluting peak divided by that of the earlier peak; values above one indicate separation is possible.
- Henderson-Hasselbalch equation
- pH equals pKa plus the logarithm of the ratio of conjugate base to acid concentration; used to predict ionization and buffer behavior.
- Buffer capacity maximum
- Buffer capacity peaks when pH equals the pKa of the buffering species, where equal acid and base forms resist pH change most effectively.
- Gravimetric factor
- The ratio of the molar mass of the sought analyte to that of the weighed precipitate, multiplied by stoichiometric coefficients, converts precipitate mass to analyte mass.
- Standard addition method
- Known analyte increments are added to aliquots of sample; extrapolating the linear response to zero signal yields the original concentration, correcting for matrix effects.
- Signal averaging improvement
- For random noise, signal-to-noise improves with the square root of the number of averaged scans; systematic drift is not reduced by averaging.
- Internal standard in GC
- A compound added at constant amount before analysis; ratioing analyte peak area to internal standard area corrects for injection volume variability.
- Equivalence point versus endpoint
- The equivalence point is the stoichiometric completion of the reaction; the endpoint is the observed signal change, ideally matched by indicator selection.
- Standard deviation of the mean
- The standard error equals the sample standard deviation divided by the square root of n; it quantifies uncertainty in a mean of replicate measurements.
- Limit of detection
- The concentration giving signal three times the standard deviation of the blank, corresponding to the lowest reliably detectable amount.
- Theoretical plates N
- N equals sixteen times the square of retention time divided by peak width squared; higher plate counts indicate more efficient separation.
- Gradient elution purpose
- Mobile phase composition changes during the run, increasing elution strength to resolve analytes with widely differing polarities in one separation.
- Karl Fischer titration
- A coulometric or volumetric method measuring water content via iodine consumption in the presence of sulfur dioxide and a base.