Common-mode rejection ratio
The ratio of differential signal gain to common-mode signal gain in an amplifier; high CMRR rejects interference such as 60 Hz power-line noise on ECG leads.
Patient leakage current limit
For cardiac-connected patient leads, the allowable leakage current is typically 10 microamperes; patient-contact applications typically allow 100 microamperes.
Defibrillator energy accuracy
Delivered energy is verified with a calibrated analyzer and resistive load; tolerance is commonly plus or minus 15 percent of the selected energy setting.
Infusion pump occlusion alarm
A pressure-threshold alarm that signals a blocked line; threshold and response time are tested separately from flow-rate accuracy, which is typically plus or minus 5 percent.
Electrosurgical return electrode monitoring
A safety system that detects inadequate pad contact or detachment, reducing the risk of burns at the return electrode during electrosurgical procedures.
SpO2 A(rms) accuracy
The root-mean-square difference between pulse oximeter readings and reference CO-oximetry values; a common specification is stated as A(rms) with the test range.
Half-value layer
The thickness of material that reduces radiation intensity by half; it is a radiographic quality-control measure of beam quality and filtration adequacy.
Total cost of ownership
The sum of acquisition, installation, maintenance, training, and disposal costs, minus salvage value, used to compare equipment options over its life.
Mean time to repair
Total repair downtime divided by the number of repairs; it measures service responsiveness and supports comparison against service-level targets.
Uptime percentage
Available time divided by scheduled time, multiplied by 100; it reflects equipment availability and is a key performance indicator for clinical engineering.
Work-order priority tiers
Emergency means immediate patient-safety impact; urgent means clinical impact within hours; routine means scheduled work with no immediate clinical effect.
In-house versus contract decision
Compare annual service-contract cost against in-house parts, labor, training, and downtime risk; choose the lower total cost with acceptable risk.
Risk-based PM interval
The scheduled maintenance interval is set by clinical criticality, failure consequence, and manufacturer guidance; higher-risk devices get shorter intervals.
PM completion rate
The number of preventive maintenance tasks completed on schedule divided by the number due in the period, multiplied by 100.
Alternate equipment management
A program that allows scheduled maintenance to follow manufacturer instructions or an approved alternative, justified by risk assessment and documented equivalence.
Recall and hazard alert process
Match the alert to inventory records, act on affected devices, document the action, and close the item within a defined timeframe.
End-of-life trigger
Conditions such as unavailable parts, unsupported software, rising failure rates, or ceased security patches that indicate a device should be replaced.
HL7 ORU message
An observation-result message type that carries clinical results from a device or system to the electronic health record using defined segments and delimiters.
FHIR Observation resource
A standardized data resource that carries a coded measurement, its value, unit, and timing, used for interoperable exchange between systems.
DICOM modality worklist
A service that sends scheduled imaging orders to the modality, reducing manual entry errors and mismatched patient or study identifiers.
Network segmentation
Separating medical devices onto isolated network segments with firewall rules, limiting lateral movement of threats and protecting device availability.
Network time protocol
A protocol that synchronizes clocks across devices and systems, enabling accurate correlation of events and log entries during investigations.
IEC 60601-1 scope
The general standard for basic safety and essential performance of medical electrical equipment; collateral standards add requirements for electromagnetic compatibility and usability.
510(k) substantial equivalence
A premarket pathway showing a new device is at least as safe and effective as a legally marketed predicate device, without full clinical trials.
Unique Device Identification
A standardized identifier on device labeling that enables traceability, inventory matching, and efficient recall management across the health system.
Medical device reporting
The requirement to report serious injury or death potentially attributable to a device to the manufacturer and the FDA within defined timeframes.
Risk priority number
In failure mode and effects analysis, the product of severity, occurrence, and detection scores; higher values rank failure modes for corrective action.
Root cause analysis
A structured review after a sentinel event that identifies system-level contributing factors in process, environment, and staffing rather than individual blame.
Alarm fatigue mitigation
Strategies such as patient-specific limits, adjusted default volumes, and alarm-management protocols that reduce nuisance alarms and clinician desensitization.
RACE fire response
A sequence for fire emergencies: rescue anyone in danger, alarm, contain the fire by closing doors, and extinguish or evacuate as directed.
ALARA principle
A radiation-safety principle meaning doses are kept as low as reasonably achievable, using time, distance, and shielding to limit occupational exposure.
PASS extinguisher use
Pull the pin, aim at the base of the fire, squeeze the handle, and sweep side to side; a standard technique for portable fire extinguisher operation.
Teach-back method
A communication technique where the learner explains the procedure in their own words, confirming understanding rather than assuming it from a yes answer.