ABMP MRI for Radiation Therapy flashcards
ABMP MRI for Radiation Therapy · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Larmor frequency at 3 T
- For hydrogen-1, f = 42.58 MHz/T × 3 T ≈ 127.7 MHz. This is the precession frequency and determines RF coil tuning and chemical shift magnitude.
- T1 relaxation time
- The time constant for recovery of longitudinal magnetization after an RF pulse; tissues with short T1 (fat) recover faster and appear bright on T1-weighted images.
- T2* decay
- Transverse signal decay that includes both spin-spin dephasing (T2) and static field inhomogeneity; gradient echo sequences are T2*-weighted and sensitive to susceptibility.
- SAR definition
- Specific absorption rate, RF power deposited per unit mass (W/kg). It increases with field strength squared and flip angle squared and limits fast 3 T sequences.
- Parallel imaging tradeoff
- Reducing phase-encoding lines by acceleration factor R shortens scan time but lowers SNR roughly by the square root of R and can produce aliasing if coil sensitivity is inadequate.
- MR conditional labeling
- Device is safe only under the specific conditions stated in its labeling, such as field strength, coil type, and SAR limits; outside those conditions safety is unknown.
- 5-gauss line
- The contour where the static magnetic field falls to 5 gauss; it defines the fringe field boundary that must be controlled to prevent projectile and pacemaker hazards.
- RF burn risk
- Elongated conductors such as ECG leads or implanted wires can concentrate RF energy and cause localized heating; risk rises with SAR and conductor length.
- Peripheral nerve stimulation
- Rapid gradient switching induces electric fields that can stimulate nerves, producing tingling or muscle twitching; dB/dt limits constrain fast sequences.
- Quench response
- Sudden loss of superconductivity releases cryogen; evacuate the room, monitor oxygen, and do not re-enter until oxygen levels are safe and the field is verified.
- SNR formula
- SNR increases with voxel volume and the square root of NEX and phase-encoding steps, and decreases with the square root of receive bandwidth.
- Chemical shift artifact
- Fat-water misregistration along the frequency-encoding direction; magnitude is proportional to field strength and inversely proportional to receive bandwidth.
- Susceptibility artifact
- Signal loss and geometric distortion near metal or air-tissue interfaces; worsens with longer TE, higher field strength, and echo-planar readout.
- Phase-encoding motion ghosting
- Motion during acquisition produces replicated ghosts along the phase-encoding direction; swapping phase and frequency directions relocates the artifact.
- Gradient nonlinearity distortion
- Spatially varying gradient fields cause geometric warping that is worst at the FOV periphery and off-isocenter; distortion correction is needed for radiotherapy geometry.
- MR simulation tabletop
- A flat, indexed tabletop with MR-compatible immobilization and external lasers is required so the simulation setup matches the treatment position.
- DICOM coordinate origin
- In MR DICOM images, patient coordinates are referenced to the scanner isocenter; correct slice position and orientation metadata are essential for planning geometry.
- 4D-MRI at simulation
- Respiratory-correlated MR acquisition used to characterize tumor motion for thoracic and abdominal targets; breath-hold instructions should match treatment delivery.
- Simulation scan protocol
- Large FOV, thin slices, distortion correction enabled, and avoidance of EPI-based sequences to preserve geometric accuracy for radiotherapy planning.
- Rigid registration DOF
- Rigid registration has six degrees of freedom: three translations and three rotations. It preserves distances and is standard for CT-MR fusion.
- Mutual information metric
- A similarity measure based on shared information between image intensities; it does not assume a linear intensity relationship and is standard for CT-MR fusion.
- Target registration error
- The distance between corresponding landmarks after registration; used to quantify fusion accuracy and to justify planning margins.
- Deformable registration validation
- Nonrigid transforms must be verified with phantoms and physician review of propagated contours because local errors can be large and are not captured by rigid metrics.
- DICOM registration object
- A DICOM object storing the spatial or deformable transform between image sets; it should be archived with the plan for traceability.
- Synthetic CT from MRI
- An MR-derived density map generated by bulk assignment or deep learning; it enables dose calculation when a planning CT is unavailable.
- MRI soft-tissue advantage
- Superior contrast for brain, prostate, and head-and-neck targets improves GTV delineation and reduces interobserver contouring variability compared with CT alone.
- Distortion in margins
- Residual MR geometric distortion, typically under 1-2 mm within the imaging FOV, must be quantified and included in PTV margins.
- Bone on MRI
- Cortical bone and air appear dark on most sequences, so synthetic CT must accurately represent bone density for heterogeneous dose calculation.
- Electron return effect
- In a magnetic field, secondary electrons curve back toward tissue-air interfaces, increasing surface and interface dose; dose algorithms must model it.
- Online adaptive radiotherapy
- Daily MR-guided recontouring and replanning based on the anatomy of the day, with plan QA completed before beam delivery.
- Cine MR tracking
- Real-time imaging used to track target motion for gating and beam-hold delivery; imaging-to-beam latency must be characterized during commissioning.
- Occupational dose limit
- Radiation workers are limited to 20 mSv per year averaged over five years, with no more than 50 mSv in any single year; the public limit is 1 mSv per year.