ABMP MRI for Radiation Therapy study plan
ABMP MRI for Radiation Therapy · Give each week a clear focus.
8 weeks · About 6 hours per week
MRI Physics and Instrumentation
Explain how proton precession, Larmor frequency, and the net magnetization vector give rise to the MR signal, writing the relationships from memory
Diagram the transmit and receive RF chain and the gradient coil set, labeling each component's function
Compare spin-echo and gradient-echo sequences by deriving how TR, TE, and flip angle change contrast weighting
Describe how slice selection, phase encoding, and frequency encoding build k-space, and predict the effect of changing each gradient
Your checkpoint
A one-page hand-drawn pulse-sequence and hardware diagram with a 10-sentence written explanation of contrast weighting, self-graded against a physics reference
MRI Physics and Instrumentation & Image Quality, Artifacts, and Quality Assurance
Distinguish T1, T2, T2*, and proton density contrast and predict which tissues brighten under each weighting
Classify the major artifacts (motion, wrap/aliasing, chemical shift, susceptibility, truncation, zipper) by their cause and their corrective strategy
Calculate signal-to-noise tradeoffs from changes in voxel size, bandwidth, averages, and field strength
List the acceptance and routine QA tests for the scanner (geometric accuracy, SNR, uniformity, ghosting) and state the pass criteria logic
Your checkpoint
An artifact-and-QA table with at least 8 artifacts, each with cause, appearance, mitigation, and the QA metric that would catch it
MRI Safety and Patient Management
Apply the static field, RF, gradient, and SAR safety limits and state the zone definitions and who may enter each
Explain the rationale and screening workflow for implanted devices, foreign bodies, and claustrophobia management
Describe quench, acoustic noise, cryogen, and pregnancy considerations and the response steps for each emergency
Differentiate MR contrast agent classes by mechanism and state the renal and pregnancy cautions
Your checkpoint
A written patient-screening checklist and a one-page emergency response flow covering quench, projectile, and patient-burn scenarios
MR Simulation for Radiation Therapy
Map the MR-simulation workflow from patient set-up through positioning aids, coil selection, and isocenter marking
Compare immobilization and positioning conventions for brain, head-and-neck, pelvis, and spine simulators
Select sequences and geometry (FOV, slice thickness, coverage) appropriate to each treatment site and justify the choices
Identify the DICOM tags and scan parameters that must be verified before exporting a simulation dataset
Your checkpoint
A site-by-site MR-simulation protocol sheet for four treatment sites listing sequence, coil, positioning, and export checks
MR Simulation for Radiation Therapy & Multimodality Image Registration and Fusion
Explain rigid versus deformable registration, the degrees of freedom, and when each is appropriate
Describe the registration workflow from image import through fusion, review, and acceptance, including resampling and interpolation effects
Identify registration failure modes (FOV mismatch, contrast differences, distortion, patient motion) and how to detect them
State the criteria used to judge registration accuracy and who signs off on a fusion
Your checkpoint
A worked registration review form applied to one anonymized CT-MR pair, with a written accuracy judgment and rationale
Treatment Planning Considerations with MRI
Explain how MR-derived electron density surrogates and bulk-density overrides are assigned in dose calculation
Describe the role of MR in target and OAR delineation for at least three disease sites
Identify geometric distortion sources and their dosimetric consequences, and state mitigation strategies
Compare MR-only and CT-MR planning workflows by their quality-assurance requirements
Your checkpoint
A site-specific delineation and planning-considerations summary for three disease sites, including distortion mitigation and density-assignment method
Treatment Delivery and On-Board MRI Guidance
Describe on-board MRI-guided delivery workflow, including daily set-up, plan adaptation triggers, and gating
Explain the safety and QA considerations unique to an in-room magnetic field during treatment
Compare gating, tracking, and adapt-to-shape strategies for motion management
Trace the steps of an online adaptive session from imaging through reoptimization and delivery
Your checkpoint
A step-by-step online adaptive workflow diagram with decision points and the QA check at each step
Radiation Protection and Professional Practice
Apply radiation protection principles (ALARA, time-distance-shielding) to the MR-simulation and treatment environment
Outline the professional and regulatory responsibilities of the MR technologist and the therapy team
Integrate all eight domains into a timed practice-exam review, flagging weak areas
Build a personal remediation plan for any domain scoring below target
Your checkpoint
A completed timed practice exam with a scored domain-by-domain analysis and a written remediation plan
Milestones
Week 2: Score at least 80% on a self-written 20-question physics and artifact quiz without notes
Week 4: Deliver the four-site simulation protocol sheet and pass a verbal walkthrough of the MR-sim workflow
Week 6: Correctly justify density assignment and distortion mitigation for all three chosen sites in a mock case review
Week 8: Achieve at least 75% overall on the timed practice exam with no single domain below 65%