Half-value layer
Thickness of material that reduces beam intensity by half; equals 0.693 divided by the linear attenuation coefficient and increases with filtration and beam energy.
Photoelectric effect
Photon absorption by an inner-shell electron, dominant at low energy and high atomic number; probability varies approximately as Z cubed over E cubed.
Compton scatter
Inelastic photon collision with an outer-shell electron; dominant interaction in soft tissue at diagnostic energies and nearly independent of atomic number.
Inverse-square law
Photon intensity from a point source decreases with the square of distance; doubling distance reduces intensity to one quarter of its original value.
Bremsstrahlung production
Radiation from electron deceleration in the target; output rises with tube current, exposure time, atomic number, and roughly the square of tube voltage.
Grid ratio
Ratio of lead strip height to interspace width; higher ratios improve scatter cleanup and contrast but require greater exposure and accurate centering.
Effective dose
Tissue-weighted sum of equivalent doses across organs; used for population comparisons and not for estimating an individual patient's risk.
Tc-99m generator
System producing technetium-99m from decaying molybdenum-99; eluate must be tested for Mo-99 breakthrough and aluminum contamination before patient use.
Decay equation
Activity at time t equals initial activity times e raised to minus 0.693 times t divided by half-life; after n half-lives activity is initial divided by 2 to the n.
PET annihilation
Positron combines with electron to emit two 511 keV photons about 180 degrees apart; coincidence detection of both defines the annihilation event location.
Collimator resolution
Spatial resolution of a gamma camera degrades as distance from the collimator face increases; sensitivity and resolution trade off with septal thickness.
Dead-time loss
Count loss at high count rates; paralyzable systems follow N equals n times e to the minus n tau, nonparalyzable systems follow N equals n over one plus n tau.
Cumulated activity
Time integral of activity in a source organ; combined with S-values in the MIRD schema to estimate absorbed dose to target organs.
F-18 half-life
Fluorine-18 decays with a half-life of about 110 minutes; this governs dose calibration timing and limits transport distance for FDG distribution.
Photopeak window
Energy acceptance range centered on the photopeak; a typical setting is a 20 percent window around 140 keV for technetium-99m imaging.
ALARA
Principle of keeping radiation doses as low as reasonably achievable, weighing social and economic factors; applies to occupational and public exposures.
Occupational dose limits
Annual limits for radiation workers: 50 mSv whole body, 150 mSv lens of eye, and 500 mSv to skin and extremities under ICRP recommendations.
Public dose limit
Annual effective dose limit for members of the public is 1 mSv above natural background, roughly one fiftieth of the occupational whole-body limit.
Declared pregnancy limit
Fetal dose limit for a declared pregnant radiation worker is 0.5 mSv in any month and 5 mSv over the entire gestation period.
Radiation weighting factor
Factor converting absorbed dose to equivalent dose; equals 1 for x-rays and gamma rays and 20 for alpha particles, reflecting biological effectiveness.
Lead apron attenuation
A 0.5 mm lead-equivalent apron attenuates scattered diagnostic x-rays by roughly 90 percent; 0.25 mm lead aprons provide less protection.
Shielding design goals
Dose targets for barrier design, commonly 0.1 mSv per week for controlled areas and 0.02 mSv per week for uncontrolled areas, per NCRP guidance.
Ring dosimeter
Extremity badge worn on the hand, typically under gloves, to monitor dose during procedures involving hands near the primary beam.
Anatomic position
Reference posture with the body upright, facing the observer, arms at the sides, and palms forward; right and left refer to the patient's sides.
Right main bronchus
Wider and more vertically oriented than the left main bronchus, making the right side a more common pathway for aspirated material.
Kidney location
Retroperitoneal organs lying near the T12 to L3 level; the right kidney sits slightly lower than the left because of the overlying liver.
Vertebral segments
The spine contains 7 cervical, 12 thoracic, 5 lumbar, 5 fused sacral, and 3 to 4 coccygeal segments; the spinal cord ends near L1 to L2.
GI tract order
Sequence is esophagus, stomach, duodenum, jejunum, ileum, cecum, ascending colon, transverse colon, descending colon, and rectum.
Hounsfield unit
CT number computed as 1000 times the difference between tissue and water attenuation coefficients divided by water's coefficient; water is 0 and air about -1000.
CT dose index
CTDIvol estimates dose per scan from tube output and pitch; dose-length product equals CTDIvol times scan length and supports effective dose estimation.
AEC reproducibility
Automatic exposure control should deliver consistent receptor exposure across repeated measurements, typically within about 10 percent of the mean value.
Detective quantum efficiency
DQE measures how efficiently a detector converts incident x-ray quanta into image signal; higher DQE indicates better dose efficiency and image quality.