ABR DMP flashcards
ABR DMP · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Half-value layer
- Thickness of material that reduces beam intensity by half; equals 0.693 divided by the linear attenuation coefficient and increases with filtration and beam energy.
- Photoelectric effect
- Photon absorption by an inner-shell electron, dominant at low energy and high atomic number; probability varies approximately as Z cubed over E cubed.
- Compton scatter
- Inelastic photon collision with an outer-shell electron; dominant interaction in soft tissue at diagnostic energies and nearly independent of atomic number.
- Inverse-square law
- Photon intensity from a point source decreases with the square of distance; doubling distance reduces intensity to one quarter of its original value.
- Bremsstrahlung production
- Radiation from electron deceleration in the target; output rises with tube current, exposure time, atomic number, and roughly the square of tube voltage.
- Grid ratio
- Ratio of lead strip height to interspace width; higher ratios improve scatter cleanup and contrast but require greater exposure and accurate centering.
- Effective dose
- Tissue-weighted sum of equivalent doses across organs; used for population comparisons and not for estimating an individual patient's risk.
- Tc-99m generator
- System producing technetium-99m from decaying molybdenum-99; eluate must be tested for Mo-99 breakthrough and aluminum contamination before patient use.
- Decay equation
- Activity at time t equals initial activity times e raised to minus 0.693 times t divided by half-life; after n half-lives activity is initial divided by 2 to the n.
- PET annihilation
- Positron combines with electron to emit two 511 keV photons about 180 degrees apart; coincidence detection of both defines the annihilation event location.
- Collimator resolution
- Spatial resolution of a gamma camera degrades as distance from the collimator face increases; sensitivity and resolution trade off with septal thickness.
- Dead-time loss
- Count loss at high count rates; paralyzable systems follow N equals n times e to the minus n tau, nonparalyzable systems follow N equals n over one plus n tau.
- Cumulated activity
- Time integral of activity in a source organ; combined with S-values in the MIRD schema to estimate absorbed dose to target organs.
- F-18 half-life
- Fluorine-18 decays with a half-life of about 110 minutes; this governs dose calibration timing and limits transport distance for FDG distribution.
- Photopeak window
- Energy acceptance range centered on the photopeak; a typical setting is a 20 percent window around 140 keV for technetium-99m imaging.
- ALARA
- Principle of keeping radiation doses as low as reasonably achievable, weighing social and economic factors; applies to occupational and public exposures.
- Occupational dose limits
- Annual limits for radiation workers: 50 mSv whole body, 150 mSv lens of eye, and 500 mSv to skin and extremities under ICRP recommendations.
- Public dose limit
- Annual effective dose limit for members of the public is 1 mSv above natural background, roughly one fiftieth of the occupational whole-body limit.
- Declared pregnancy limit
- Fetal dose limit for a declared pregnant radiation worker is 0.5 mSv in any month and 5 mSv over the entire gestation period.
- Radiation weighting factor
- Factor converting absorbed dose to equivalent dose; equals 1 for x-rays and gamma rays and 20 for alpha particles, reflecting biological effectiveness.
- Lead apron attenuation
- A 0.5 mm lead-equivalent apron attenuates scattered diagnostic x-rays by roughly 90 percent; 0.25 mm lead aprons provide less protection.
- Shielding design goals
- Dose targets for barrier design, commonly 0.1 mSv per week for controlled areas and 0.02 mSv per week for uncontrolled areas, per NCRP guidance.
- Ring dosimeter
- Extremity badge worn on the hand, typically under gloves, to monitor dose during procedures involving hands near the primary beam.
- Anatomic position
- Reference posture with the body upright, facing the observer, arms at the sides, and palms forward; right and left refer to the patient's sides.
- Right main bronchus
- Wider and more vertically oriented than the left main bronchus, making the right side a more common pathway for aspirated material.
- Kidney location
- Retroperitoneal organs lying near the T12 to L3 level; the right kidney sits slightly lower than the left because of the overlying liver.
- Vertebral segments
- The spine contains 7 cervical, 12 thoracic, 5 lumbar, 5 fused sacral, and 3 to 4 coccygeal segments; the spinal cord ends near L1 to L2.
- GI tract order
- Sequence is esophagus, stomach, duodenum, jejunum, ileum, cecum, ascending colon, transverse colon, descending colon, and rectum.
- Hounsfield unit
- CT number computed as 1000 times the difference between tissue and water attenuation coefficients divided by water's coefficient; water is 0 and air about -1000.
- CT dose index
- CTDIvol estimates dose per scan from tube output and pitch; dose-length product equals CTDIvol times scan length and supports effective dose estimation.
- AEC reproducibility
- Automatic exposure control should deliver consistent receptor exposure across repeated measurements, typically within about 10 percent of the mean value.
- Detective quantum efficiency
- DQE measures how efficiently a detector converts incident x-ray quanta into image signal; higher DQE indicates better dose efficiency and image quality.