ABR DMP study materials
ABR DMP study resources

ABR DMP key terms

ABR DMP · Get comfortable with the vocabulary.

Air kerma

Kinetic energy released per unit mass in air by uncharged particles such as photons, expressed in gray. It is the standard quantity for specifying output of radiographic and fluoroscopic x-ray tubes.

Related: Absorbed dose, Dose equivalent
Absorbed dose

Energy deposited by ionizing radiation per unit mass of material, measured in gray (joule per kilogram). It describes energy actually imparted to tissue rather than energy merely present in the beam.

Related: Air kerma, Dose equivalent
Dose equivalent

A quantity that weights absorbed dose by radiation type to reflect biological harm, expressed in sievert. In modern usage the operational and protection quantities (equivalent dose, effective dose) supersede it.

Related: Effective dose, Absorbed dose
Effective dose

Tissue-weighted sum of equivalent doses across organs, expressed in sievert, estimating whole-body stochastic risk from a nonuniform exposure. It allows comparison of risk across different imaging procedures.

Related: Dose equivalent, Diagnostic reference level
ALARA

The principle of keeping radiation exposures as low as reasonably achievable, accounting for economic and social factors. It underpins shielding design, protocol optimization, and occupational dose monitoring programs.

Related: Diagnostic reference level, Half-value layer
Diagnostic reference level

A dose or activity benchmark, usually set near the third quartile of a survey of facilities, used to flag unusually high exposures for adult imaging. Values are advisory for optimization, not legal dose limits.

Related: ALARA, Effective dose
Half-value layer

Thickness of a specified material that reduces the intensity of a radiation beam to half its original value. It characterizes beam quality and is used in shielding calculations and filtration specifications.

Related: Air kerma, ALARA
Annual occupational dose limit

The regulatory ceiling on effective dose to a radiation worker in a year, generally 50 mSv (5 rem) in United States regulation. It applies to declared workers, not to members of the public.

Related: Personnel dosimeter, Effective dose
Personnel dosimeter

A device worn by a radiation worker that records cumulative occupational dose, commonly a thermoluminescent or optically stimulated luminescence badge. Results are reviewed against regulatory and institutional action levels.

Related: Annual occupational dose limit, Pregnancy dose inquiry
Pregnancy dose inquiry

The practice of asking patients of childbearing potential about possible pregnancy before pelvic or abdominal ionizing imaging, so that risk counseling and protocol adjustment can occur. It is not a universal requirement for every examination.

Related: Personnel dosimeter, Fetal dose
Fetal dose

The estimated absorbed dose to a conceptus from a maternal imaging procedure, typically well under 50 mGy for most diagnostic studies. Counseling thresholds for stochastic and deterministic risk are commonly cited around that level.

Related: Pregnancy dose inquiry, Absorbed dose
Photoelectric effect

Interaction in which a photon transfers all its energy to a bound inner-shell electron, which is ejected. Probability rises steeply at low energy and with high atomic number, producing strong contrast between bone and soft tissue.

Related: Compton scattering, K-edge
Compton scattering

Interaction in which a photon transfers part of its energy to a loosely bound electron and is deflected with reduced energy. It is the dominant attenuation process in soft tissue at diagnostic energies and a major source of scatter.

Related: Photoelectric effect, Scatter radiation
K-edge

A sharp increase in photoelectric absorption that occurs when photon energy just exceeds the binding energy of the K-shell electron of an atom. Iodine and barium K-edges are exploited in contrast imaging.

Related: Photoelectric effect, Contrast agent
Linear attenuation coefficient

Probability per unit path length that a photon interacts in a material, dependent on photon energy and material density. It describes exponential attenuation of a narrow beam through matter.

Related: Half-value layer, Compton scattering
Scatter radiation

Radiation deflected from its original path by interactions in the patient or surroundings, degrading image contrast and contributing occupational dose. It is controlled with grids, collimation, and shielding.

Related: Compton scattering, Antiscatter grid
Antiscatter grid

An array of alternating radiopaque strips and radiolucent spacers placed between patient and detector to absorb obliquely traveling scatter while transmitting primary photons. Grid ratio and frequency govern its effectiveness and dose cost.

Related: Scatter radiation, Bucky factor
Bucky factor

The ratio of patient dose with a grid to dose without it for equal detector signal, reflecting the increased exposure needed to compensate for primary attenuation by the grid. It rises with grid ratio.

Related: Antiscatter grid, Exposure index
Heel effect

Variation in x-ray beam intensity across the field, lower on the anode side because photons produced deeper in the target must traverse more anode material. It influences tube orientation for anatomy of differing thickness.

Related: Focal spot, Tube output
Focal spot

The area on the anode where electrons strike and x-rays are produced. Its size limits spatial resolution; small focal spots improve sharpness but restrict tube current because of heat loading.

Related: Heel effect, Modulation transfer function
Modulation transfer function

A measure of how well an imaging system preserves contrast at each spatial frequency, normalized to one at zero frequency. It describes resolution performance independent of noise.

Related: Detective quantum efficiency, Focal spot
Detective quantum efficiency

The squared ratio of output to input signal-to-noise ratio of a detector, indicating how efficiently it converts incident x-ray quanta into image signal. Higher values mean lower dose for equivalent image quality.

Related: Modulation transfer function, Exposure index
Exposure index

A vendor-specific number reported by digital radiography systems that indicates detector exposure level, used to judge whether technique is appropriate. It is not interchangeable across manufacturers.

Related: Detective quantum efficiency, Bucky factor
Tube output

Radiation exposure produced per unit charge at a given tube potential, typically expressed as air kerma per milliampere-second at a reference distance. It depends on target material, filtration, and voltage waveform.

Related: Heel effect, Air kerma
Technetium-99m

A metastable radionuclide with a 6-hour half-life and 140 keV gamma emission, produced from a molybdenum-99 generator. It is the workhorse of nuclear medicine because of its favorable energy and versatile chemistry.

Related: Generator elution, Photopeak
Generator elution

The process of extracting sodium pertechnetate from a molybdenum-99/technetium-99m generator using saline. Elution efficiency and molybdenum breakthrough are quality control concerns.

Related: Technetium-99m, Radionuclide purity
Radionuclide purity

The fraction of total radioactivity in a preparation attributable to the intended radionuclide, excluding contaminating species such as molybdenum-99. It is verified by measuring transmission through lead shielding.

Related: Generator elution, Radiochemical purity
Radiochemical purity

The proportion of radioactivity present in the desired chemical form rather than as free or altered species. It is assessed by chromatography and affects biodistribution and image quality.

Related: Radionuclide purity, Technetium-99m
Photopeak

The peak in a gamma spectrum corresponding to full absorption of a photon in the detector, used to set the energy window. Photomultiplier gain is adjusted so the peak falls at the expected channel.

Related: Energy window, Gamma camera
Energy window

A selected range of pulse heights accepted by the gamma camera, usually centered on the photopeak, that determines which events are counted. Narrower windows reject scatter but reduce sensitivity.

Related: Photopeak, Scatter radiation
Gamma camera

An imaging device using a collimator, scintillation crystal, and photomultiplier tubes to detect and localize gamma emissions from a radiotracer. Spatial resolution and sensitivity trade off through collimator design.

Related: Photopeak, Collimator
Collimator

A lead structure with holes that restricts the gamma camera to accepting only nearly parallel photons, projecting the source distribution onto the crystal. Hole size, length, and septa set resolution and sensitivity.

Related: Gamma camera, Energy window
Positron emission tomography

Imaging modality detecting pairs of 511 keV annihilation photons emitted in opposite directions from positron decay. Coincidence detection localizes the event along a line of response.

Related: Time-of-flight PET, Attenuation correction
Time-of-flight PET

A PET technique that measures small differences in arrival times of annihilation photons to localize the event along the line of response, improving signal-to-noise ratio in large patients.

Related: Positron emission tomography, Attenuation correction
Attenuation correction

Correction of emission data for photon loss within the patient, typically using a transmission or computed tomography map. Without it, deep structures appear artificially low in activity.

Related: Positron emission tomography, Compton scattering
Osteoblast

A bone-forming cell that synthesizes and mineralizes osteoid matrix. Its activity underlies radiotracer uptake on bone scans and explains focal uptake at sites of fracture healing or metastasis.

Related: Osteoclast, Skeletal scintigraphy
Osteoclast

A large multinucleated cell that resorbs bone matrix. Coupled with osteoblast activity, it drives the remodeling that produces increased tracer uptake in metabolically active bone lesions.

Related: Osteoblast, Skeletal scintigraphy
Skeletal scintigraphy

Nuclear medicine imaging of bone using technetium-99m labeled diphosphonates that adsorb to hydroxyapatite, with uptake reflecting regional blood flow and osteoblastic remodeling.

Related: Osteoblast, Technetium-99m
Nephron

The functional unit of the kidney, comprising a glomerulus and tubule that filter plasma and reabsorb or secrete solutes. Glomerular filtration rate and tubular secretion determine renal tracer clearance.

Related: Glomerular filtration rate, Renal scintigraphy
Glomerular filtration rate

The volume of plasma filtered by the kidneys per unit time, a key measure of renal function. It can be estimated from the clearance of a filtration-only radiotracer.

Related: Nephron, Renal scintigraphy
Renal scintigraphy

Functional renal imaging with radiotracers such as technetium-99m mercaptoacetyltriglycine, assessing perfusion, extraction, and excretion. It is useful when anatomy and function diverge.

Related: Nephron, Glomerular filtration rate
Thyroid follicle

The structural unit of the thyroid gland, lined by cells that trap iodide and synthesize thyroid hormones stored as colloid. Its iodide handling allows radionuclide imaging and therapy.

Related: Radioiodine uptake, Thyroid scintigraphy
Radioiodine uptake

Measurement of the fraction of administered radioiodine concentrated by the thyroid at a set time, reflecting gland activity. It helps distinguish hyperthyroid states and guides therapy dosing.

Related: Thyroid follicle, Thyroid scintigraphy
Thyroid scintigraphy

Imaging of the thyroid after administration of a gamma-emitting iodine or pertechnetate tracer to map functional tissue. It identifies autonomous nodules and ectopic or remnant tissue.

Related: Thyroid follicle, Radioiodine uptake