ARRT Bone Densitometry study materials
ARRT Bone Densitometry study resources

ARRT Bone Densitometry cheat sheet

ARRT Bone Densitometry · Keep the key ideas within reach.

Quick reference

Equipment and Instrumentation

  • Dual-energy separation: a K-edge filter (typically cerium or samarium) splits the spectrum into low-energy (~70 kVp peak) and high-energy (~140 kVp peak) components; differential attenuation permits soft-tissue subtraction and isolation of bone mineral.

  • Pencil-beam geometry: single collimated detector in rectilinear raster scanning - slower, high precision, negligible magnification. Fan-beam geometry: linear detector array with a wide beam - faster, higher spatial resolution, slight magnification that varies with object-to-detector distance.

  • Areal BMD (g/cm²) = bone mineral content (g) ÷ projected area (cm²); because the measurement is projected, a large bone reads higher and a small bone lower at identical true volumetric density.

  • Typical effective dose is roughly 1–10 µSv for a lumbar spine or hip scan (fan beam at the upper end); scatter is low enough that the operator may remain in the scan room without a lead barrier.

  • The QC phantom is an anthropomorphic spine phantom with hydroxyapatite-equivalent inserts; its measured BMD is plotted on a control chart on the manufacturer-recommended schedule (usually daily).

  • A common patient table weight limit is about 300 lb (136 kg); exceeding the stated limit risks table damage and invalid data, so the manufacturer's specification is checked before scanning.

  • DXA is the reference standard for BMD; QCT reports volumetric density in mg/cm³, and calcaneal QUS reports speed of sound and broadband attenuation - QUS values cannot be converted into DXA T-scores.

  • Collimation, detector, or alignment faults present as nonuniform phantom BMD or an unstable control chart; phantom BMD trending beyond tolerance indicates service is required before patient scanning.

Quick reference

Patient Care and Education

  • T-score = (patient BMD − young-adult mean) ÷ young-adult SD. Z-score = (patient BMD − age-matched mean) ÷ age-matched SD.

  • WHO categories by T-score: −1.0 or above = normal; between −1.0 and −2.5 = low bone mass (osteopenia); −2.5 or lower = osteoporosis; −2.5 or lower with a fragility fracture = severe (established) osteoporosis.

  • Diagnosis uses the lowest T-score among lumbar spine, femoral neck, total hip, and 33% radius (the radius only when spine and hip are unavailable or invalid).

  • Z-scores, not T-scores, are reported for premenopausal women, men younger than 50, and children; Z ≤ −2.0 is termed below the expected range for age and prompts secondary workup.

  • Metal hardware (zippers, buttons, coins, keys, snaps) is removed before acquisition; recent barium or iodinated contrast and recent radionuclide administration can invalidate lumbar spine analysis and follow the facility's deferral interval.

  • Pregnancy is a relative contraindication: DXA is deferred unless the clinical indication clearly outweighs the small fetal dose.

  • History needed for interpretation: prior fragility fracture, glucocorticoid use of 3 months or more, parental hip fracture, smoking, alcohol intake, height loss, and previous DXA results.

  • Standing height and weight are measured at each visit; they are required for reference database matching, Z-score derivation, FRAX calculation, and detection of height loss.

Quick reference

Image Acquisition and Analysis

  • Default lumbar spine region of interest is L1–L4; exclude vertebrae that are fractured, contain focal sclerosis or artifact, have vertebroplasty or laminectomy, or show severe degenerative change.

  • A minimum of two evaluable lumbar vertebrae is required; if only one remains, the spine result is non-diagnostic and the hip is used instead.

  • Patient positioning for the spine: supine, centered near L3, with the legs resting on a foam block to reduce lumbar lordosis and open the disc spaces.

  • Hip acquisition: internal rotation of the femur about 15–25° with the foot secured in the positioning device so the femoral neck lies parallel to the table and is measured at full length.

  • Reported hip regions are femoral neck, total hip, and trochanter; Ward's area is displayed but is excluded from diagnosis because of poor reproducibility.

  • When the two hips differ, the lower BMD and lower T-score at the femoral neck or total hip govern diagnosis and monitoring; a single available hip may be used.

  • Forearm site: 33% radius of the non-dominant arm, a cortical site used when the spine and hip are unavailable, invalid, or uninterpretable.

  • Vertebral numbering landmarks: lowest ribs align with T12 and the iliac crest with the L4–L5 interspace; consistent labeling is mandatory for serial comparison.

  • A single vertebra whose BMD differs from its neighbors by more than about 1.0 T-score (roughly 0.1 g/cm²) is inspected for artifact and excluded if the cause is confirmed.

Quick reference

Quality Control and Data Management

  • Precision CV (%) = (standard deviation ÷ mean) × 100, derived from repeat in vivo measurements or repeated phantom scans.

  • Least significant change at 95% confidence = 2.77 × precision CV; a follow-up difference smaller than the LSC lies within measurement error and is not reported as change.

  • ISCD precision assessment: duplicate scans in 30 patients, or 15 patients repositioned between scans, per site and per technologist; recalculate after a change in technologist, software, or equipment.

  • Daily phantom values are plotted on a control chart; investigate drift exceeding about 1.5% from baseline or any point outside ±3 SD before scanning patients.

  • Reference databases: femoral neck T-scores in the U.S. use NHANES III data for non-Hispanic white women aged 20–29; spine reference data come from the manufacturer - the same database version must be used for serial scans.

  • Cross-calibration after equipment replacement or major software change: scan a phantom and/or 10–30 patients on both systems and derive a conversion before comparing serial results.

  • Repeat DXA is generally not indicated sooner than 2 years unless therapy changes or clinical findings warrant it, since typical annual change often falls below the LSC.

  • FRAX estimates 10-year major osteoporotic and hip fracture probability from femoral neck BMD, age, sex, BMI, prior fracture, parental hip fracture, smoking, glucocorticoids, rheumatoid arthritis, alcohol, and secondary osteoporosis.

  • Report documentation: scanner model, software version, technologist, sites analyzed, excluded vertebrae or regions, artifacts, and comparison with prior studies.

Quick reference

Equipment and Instrumentation

  • Dual-energy separation uses a K-edge filter to produce distinct low- and high-energy spectra.

  • Pencil beam: single detector, slow, high precision, negligible magnification.

  • Fan beam: array detector, fast, higher resolution, slight magnification and dose increase.

  • Areal BMD (g/cm²) = bone mineral content ÷ projected bone area, so it is size-dependent.

  • Effective dose for spine or hip DXA is roughly 1–10 µSv.

  • QC phantom is an anthropomorphic spine phantom with hydroxyapatite-equivalent inserts.

  • Typical table weight limit is about 300 lb (136 kg); verify the manufacturer's specification.

  • QCT gives volumetric mg/cm³; calcaneal QUS gives SOS and BUA, not interchangeable with DXA.

  • Phantom BMD drift beyond tolerance indicates a collimation, detector, or alignment problem.

  • BMD is expressed in g/cm² because DXA measures a projected, areal density.

Quick reference

Patient Care and Education

  • T-score compares patient BMD to the young-adult mean; Z-score compares to the age-matched mean.

  • WHO thresholds: normal ≥ −1.0; low bone mass −1.0 to −2.5; osteoporosis ≤ −2.5; severe osteoporosis if a fragility fracture is present.

  • Diagnosis uses the lowest T-score of spine, femoral neck, total hip, or 33% radius.

  • Premenopausal women, men under 50, and children are reported with Z-scores, not T-scores.

  • Metal hardware is removed; recent barium, contrast, or radionuclides can invalidate spine analysis.

  • Pregnancy is a relative contraindication to DXA.

  • History items relevant to interpretation include glucocorticoid use, prior fragility fracture, and family history.

  • Height and weight are recorded at each visit for reference data and FRAX.

  • DXA involves no injection and typical total visit time is 15–30 minutes.

Quick reference

Image Acquisition and Analysis

  • Standard spine ROI is L1–L4; fractured, sclerotic, or artifact-containing vertebrae are excluded.

  • At least two evaluable vertebrae are required for a valid spine result.

  • Legs elevated on a block reduce lumbar lordosis during spine scanning.

  • Femur is internally rotated 15–25° so the femoral neck is measured at full length.

  • Reported hip regions: femoral neck, total hip, trochanter; Ward's area is not diagnostic.

  • Use the lower of the two hips when they disagree.

  • 33% radius of the non-dominant forearm is the alternative site.

  • Lowest ribs = T12; iliac crest = L4–L5 interspace for vertebral numbering.

  • TBS is a texture index from the spine image that adds fracture risk information beyond BMD.

Quick reference

Quality Control and Data Management

  • CV (%) = (SD ÷ mean) × 100 quantifies precision.

  • LSC at 95% confidence = 2.77 × precision CV.

  • ISCD precision protocol: 30 patients with duplicate scans, or 15 patients repositioned.

  • Daily phantom BMD is plotted on a control chart; ±1.5% or ±3 SD triggers investigation.

  • Femoral neck T-scores use NHANES III non-Hispanic white women aged 20–29.

  • Cross-calibration is required after scanner or software change.

  • Repeat scans generally are not done before 2 years unless clinically indicated.

  • FRAX uses femoral neck BMD plus clinical risk factors to give 10-year fracture probability.

  • Reports must document scanner, software version, sites, exclusions, and artifacts.

Quick reference

Equipment and Instrumentation

  • K-edge filter materials commonly used in DXA include cerium and samarium.

  • Pencil-beam scanning produces essentially no magnification of the measured region.

  • Fan-beam magnification increases as the object moves farther from the detector.

  • Bone mineral content is reported in grams; BMD in g/cm²; area in cm².

  • DXA dose is far lower than a chest radiograph (about 100 µSv).

  • The spine phantom is scanned on a fixed schedule to verify scanner stability.

  • Calcaneal QUS predicts fracture but is not used for diagnosis or monitoring by DXA criteria.

  • Table weight and height limits are published by the manufacturer and must be observed.

Quick reference

Patient Care and Education

  • T-score = (patient BMD − young-adult mean BMD) ÷ young-adult SD.

  • Z-score = (patient BMD − age-matched mean BMD) ÷ age-matched SD.

  • Z ≤ −2.0 is below the expected range for age.

  • Exclude metal objects before scanning to avoid focal BMD artifact.

  • Defer spine scanning after recent barium or radionuclide studies per protocol.

  • Document height, weight, and relevant medications such as glucocorticoids.

  • Explain that DXA measures bone density and estimates fracture risk, not current fracture.

  • A 33% radius T-score may be used for diagnosis only when spine and hip are unusable.

  • Premenopausal women should not be diagnosed with osteoporosis from a T-score alone.

Quick reference

Image Acquisition and Analysis

  • Vertebrae with vertebroplasty cement or surgical hardware are excluded from the spine ROI.

  • Degenerative sclerosis and aortic calcification artificially raise spine BMD.

  • The femoral neck should appear long and narrow on a well-positioned hip image.

  • Total hip ROI includes the femoral neck, trochanter, and intertrochanteric region.

  • The 33% radius site is predominantly cortical bone.

  • Inconsistent vertebral labeling causes comparison of different vertebrae over time.

  • A vertebra differing by more than about 1.0 T-score from neighbors is checked for artifact.

  • Low TBS indicates degraded trabecular microarchitecture at a given BMD.

  • Single-energy or mis-centered scans produce non-diagnostic images and require repeat acquisition.

Quick reference

Quality Control and Data Management

  • Precision is expressed as CV in percent or as absolute SD in g/cm².

  • LSC must be exceeded before a BMD change is called significant.

  • Precision values are site-specific and technologist-specific.

  • Control charts use Shewhart-type rules to detect drift and shifts.

  • The same reference database version is maintained for all serial scans.

  • Conversion formulas from cross-calibration are stored with the patient record.

  • Clinical risk factors beyond BMD are captured for FRAX reporting.

  • Software upgrades require repeat phantom scanning and review of prior comparisons.

  • Archived scan files should be retained per facility policy for future re-analysis.