Bone Densitometry is one of ARRT's postprimary credentials. ARRT states its purpose as recognizing individuals who are qualified to perform the role of a Bone Densitometrist-technologists who create images with specialized X-ray equipment so that clinicians can verify low bone density, osteoporosis, or subtle bone loss before recommending treatment such as medication or lifestyle change. Because it is a postprimary discipline, candidates must already hold ARRT certification and registration in an approved supporting discipline; ARRT notes that a supporting credential may in some cases come from another organization, and that holders of a Nuclear Medicine Technology credential from the NMTCB can submit an online account access form to get started. The examination samples four content areas: equipment and instrumentation, patient care and education, image acquisition and analysis, and quality control and data management. The ARRT Bone Densitometry study guide has 4 sections and about 1,079 words.

01

Equipment and Instrumentation

DXA measures areal bone mineral density (BMD, g/cm²) by passing two photon spectra through tissue and solving the attenuation equations for soft tissue and bone. Systems either switch kVp or filter the beam with a K-edge filter (cerium or samarium) to create low- and high-energy components; the low-energy beam is preferentially absorbed by calcium hydroxyapatite.

279 words, about 1 min
02

Patient Care and Education

Ask about pregnancy, recent contrast or nuclear medicine studies, and metal in the region of interest. DXA is very low dose (roughly 1–10 µSv, comparable to a few days of background radiation); pregnancy is not an absolute contraindication, but scan only when justified.

285 words, about 1 min
03

Image Acquisition and Analysis

Spine: supine, centered on the table, knees flexed over a foam block to flatten lumbar lordosis; scan approximately T12–L5. Hip: supine with the foot secured in the positioning device and the leg in 15–25° internal rotation so the femoral neck lies parallel to the table; include the ischium.

250 words, about 1 min
04

Quality Control and Data Management

Scan the manufacturer's phantom daily , before patients, and plot results on a control chart. Investigate values beyond ±2 SD or a sustained trend.

265 words, about 1 min