ASNT NDT III ET study materials
ASNT NDT III ET study resources

ASNT NDT III ET cheat sheet

ASNT NDT III ET · Keep the key ideas within reach.

Quick reference

NDT General (Basic)

  • ASNT SNT-TC-1A is a recommended practice; a written practice issued by the employer is the controlling document for personnel qualification and certification.

  • A Level III is responsible for method selection, technique development, procedure approval, and interpretation of codes and standards for the method certified.

  • Certification to Level II requires passing a general examination, a method-specific examination, and a practical examination, plus documented experience and vision acuity.

  • Near vision acuity is typically verified at 12 in (30 cm) with Jaeger J1 or equivalent, and color perception must be demonstrated per the written practice.

  • Probability of detection increases with discontinuity size, favorable orientation to the inspection direction, and higher signal-to-noise ratio; it is never 100 percent.

  • A discontinuity is any interruption in material continuity; a defect is a discontinuity that exceeds acceptance criteria or impairs serviceability.

  • Reference standards establish the baseline response; they must be traceable, of the same material family as the part when possible, and verified before and after inspection.

  • The examination re-certification interval is defined by the employer's written practice; typical intervals are 3 to 5 years with periodic performance checks.

Quick reference

Electromagnetic Testing (ET) Method Specific

  • Eddy current testing depends on Faraday induction: a time-varying coil current induces eddy currents in the conductor, and the secondary field from those currents is sensed by the coil as a change in impedance.

  • Standard depth of penetration is delta = 1/sqrt(pifmu*sigma); at one delta the current density is 36.8 percent of the surface value, and effective penetration is about three delta.

  • For a nonmagnetic conductor, delta (mm) is approximately 503*sqrt(rho/f) with resistivity in microhm-cm and frequency in Hz; higher frequency or higher conductivity reduces depth.

  • Fill factor is the ratio of test object cross-sectional area to coil cross-sectional area; higher fill factor increases coupling and signal amplitude.

  • Lift-off changes coil impedance and appears as a phase shift; it is minimized by close coupling or suppressed by choosing a test frequency and phase that separates lift-off from the flaw signal.

  • Absolute coils respond to gradual property changes and drift; differential coils reject common-mode effects such as temperature and lift-off but do not detect gradual changes.

  • Phase analysis separates variables: discontinuity depth, conductivity, permeability, and geometry each produce characteristic phase angles on the impedance plane.

  • Magnetic particle and flux leakage methods apply to ferromagnetic materials; eddy current testing is most effective on nonmagnetic conductors unless permeability is controlled.

  • Narrow-band filtering and phase discrimination improve signal-to-noise ratio, but filtering must not distort or delay the flaw indication beyond acceptable limits.