Leak rate formula
For viscous flow through a capillary, Q = π d^4 ΔP / (128 μ L), where d is diameter, L is length, μ is viscosity, and ΔP is pressure drop.
Bubble immersion test
A leak testing method where the component is pressurized and immersed in liquid; leaks are indicated by bubbles. Leak rate is estimated from bubble size and formation rate.
Helium mass spectrometer
A sensitive leak detector that measures helium partial pressure; minimum detectable leak typically 1×10^-10 std cm³/s. Requires calibration with a certified leak standard.
Pressure decay test
A leak test where the component is pressurized and the pressure drop over time is measured. Leak rate Q = (V ΔP) / (Δt Patm), assuming isothermal ideal gas.
Vacuum hood test
A tracer gas leak test where a hood encloses the component and vacuum is applied; leaking tracer gas is detected. Leak rate is derived from signal rise and calibration.
Ultrasonic leak detection
Detects high-frequency sound produced by turbulent gas flow through a leak. Effective for pressure differentials >5 psi and leak rates >1×10^-3 std cm³/s.
Halogen diode detector
A leak detector sensitive to halogen gases; minimum detectable leak ~1×10^-6 std cm³/s. Sensor can be poisoned by high concentrations of halogen.
Tracer gas selection
Tracer gas should have low background concentration, high detection sensitivity, and compatibility with the test component. Helium is common due to low atmospheric concentration.
Leak standard calibration
A certified leak standard with known leak rate is used to calibrate leak detectors. Its uncertainty should be ≤1/3 of the tolerance being measured.
Procedure qualification record
Documentation that a leak testing procedure has been qualified, including test method, tracer gas, pressure, temperature, equipment, and acceptance criteria.
Repeatability and reproducibility
Qualification of LT procedures requires demonstration of repeatability (same operator) and reproducibility (different operators) typically via three runs by two operators.
Environmental factors in LT
Temperature, background tracer gas, and humidity can affect leak test results; they must be controlled or compensated during testing.
Acceptance criteria
The maximum allowable leak rate specified for a component, typically expressed in std cm³/s or mbar·L/s, based on design requirements.
Personnel qualification for LT
Personnel performing leak tests must be qualified to a recognized standard (e.g., SNT-TC-1A) with documented training, experience, and vision acuity.
Ideal gas law
PV = nRT, used to convert leak rates to standard conditions (T=273 K, P=1 atm). Essential for comparing leak rates measured at different conditions.
Viscosity temperature dependence
For gases, viscosity increases with temperature approximately as μ ∝ T^0.5; this affects leak rate calculations for viscous flow.
Speed of sound in gas
c = √(γ R T / M), where γ is specific heat ratio, R is gas constant, T is temperature, and M is molar mass. Used in acoustic leak detection.
Vacuum unit conversions
1 atm = 760 torr = 101325 Pa = 1.01325 bar. 1 std cm³/s = 0.001 Pa·m³/s = 0.001 mbar·L/s.
Knudsen number
Kn = λ / L, where λ is mean free path and L is characteristic length. Kn < 0.01 indicates viscous flow; Kn > 1 indicates molecular flow.
Pirani gauge
A thermal conductivity vacuum gauge used for rough vacuum measurement (typically 10^-3 to 10 torr). Not suitable for high vacuum.
Ionization gauge
A vacuum gauge that ionizes gas molecules and measures ion current; used for high vacuum (10^-3 to 10^-10 torr).
NDT method selection
Choice of NDT method (VT, PT, MT, UT, RT, ET, LT) depends on material, defect type, and accessibility. Leak testing is for detecting leaks, not internal defects.
SNT-TC-1A
A recommended practice for NDT personnel qualification and certification. Employers write their own practice; certification levels I, II, III with minimum training and experience hours.
CP-189
A standard that establishes minimum requirements for qualification and certification of NDT personnel. More prescriptive than SNT-TC-1A.
Level III responsibilities
Level III personnel approve procedures, interpret codes, and train/certify Level I and II personnel. They also must pass a basic, method, and practical exam.
Certification examination
Requires passing general, specific, and practical examinations. Passing score typically 80%. Records must be maintained by the employer.
Recertification requirements
SNT-TC-1A requires recertification every 5 years, including annual vision test (Jaeger 1 or equivalent) and documented continuing experience.
Vision acuity test
Annual vision test required for NDT personnel; must demonstrate near vision acuity of Jaeger 1 or equivalent, and color perception as applicable.
Employer certification records
Employer must maintain records of training, experience, examination results, and vision tests for each certified individual.
Leak rate units conversion
1 std cm³/s = 0.001 mbar·L/s = 0.1 Pa·L/s = 0.001 Pa·m³/s. Standard conditions are 0°C and 1 atm.
Molecular flow conductance
For a long tube, C = (π d^3 / (12 L)) √(2π R T / M), where d is diameter, L is length, R is gas constant, T is temperature, M is molar mass.
Background tracer gas
Ambient concentration of tracer gas can cause false signals; must be minimized by ventilation or using a tracer with low atmospheric concentration.