ASNT NDT III LT flashcards
ASNT NDT III LT · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Leak rate formula
- For viscous flow through a capillary, Q = π d^4 ΔP / (128 μ L), where d is diameter, L is length, μ is viscosity, and ΔP is pressure drop.
- Bubble immersion test
- A leak testing method where the component is pressurized and immersed in liquid; leaks are indicated by bubbles. Leak rate is estimated from bubble size and formation rate.
- Helium mass spectrometer
- A sensitive leak detector that measures helium partial pressure; minimum detectable leak typically 1×10^-10 std cm³/s. Requires calibration with a certified leak standard.
- Pressure decay test
- A leak test where the component is pressurized and the pressure drop over time is measured. Leak rate Q = (V ΔP) / (Δt Patm), assuming isothermal ideal gas.
- Vacuum hood test
- A tracer gas leak test where a hood encloses the component and vacuum is applied; leaking tracer gas is detected. Leak rate is derived from signal rise and calibration.
- Ultrasonic leak detection
- Detects high-frequency sound produced by turbulent gas flow through a leak. Effective for pressure differentials >5 psi and leak rates >1×10^-3 std cm³/s.
- Halogen diode detector
- A leak detector sensitive to halogen gases; minimum detectable leak ~1×10^-6 std cm³/s. Sensor can be poisoned by high concentrations of halogen.
- Tracer gas selection
- Tracer gas should have low background concentration, high detection sensitivity, and compatibility with the test component. Helium is common due to low atmospheric concentration.
- Leak standard calibration
- A certified leak standard with known leak rate is used to calibrate leak detectors. Its uncertainty should be ≤1/3 of the tolerance being measured.
- Procedure qualification record
- Documentation that a leak testing procedure has been qualified, including test method, tracer gas, pressure, temperature, equipment, and acceptance criteria.
- Repeatability and reproducibility
- Qualification of LT procedures requires demonstration of repeatability (same operator) and reproducibility (different operators) typically via three runs by two operators.
- Environmental factors in LT
- Temperature, background tracer gas, and humidity can affect leak test results; they must be controlled or compensated during testing.
- Acceptance criteria
- The maximum allowable leak rate specified for a component, typically expressed in std cm³/s or mbar·L/s, based on design requirements.
- Personnel qualification for LT
- Personnel performing leak tests must be qualified to a recognized standard (e.g., SNT-TC-1A) with documented training, experience, and vision acuity.
- Ideal gas law
- PV = nRT, used to convert leak rates to standard conditions (T=273 K, P=1 atm). Essential for comparing leak rates measured at different conditions.
- Viscosity temperature dependence
- For gases, viscosity increases with temperature approximately as μ ∝ T^0.5; this affects leak rate calculations for viscous flow.
- Speed of sound in gas
- c = √(γ R T / M), where γ is specific heat ratio, R is gas constant, T is temperature, and M is molar mass. Used in acoustic leak detection.
- Vacuum unit conversions
- 1 atm = 760 torr = 101325 Pa = 1.01325 bar. 1 std cm³/s = 0.001 Pa·m³/s = 0.001 mbar·L/s.
- Knudsen number
- Kn = λ / L, where λ is mean free path and L is characteristic length. Kn < 0.01 indicates viscous flow; Kn > 1 indicates molecular flow.
- Pirani gauge
- A thermal conductivity vacuum gauge used for rough vacuum measurement (typically 10^-3 to 10 torr). Not suitable for high vacuum.
- Ionization gauge
- A vacuum gauge that ionizes gas molecules and measures ion current; used for high vacuum (10^-3 to 10^-10 torr).
- NDT method selection
- Choice of NDT method (VT, PT, MT, UT, RT, ET, LT) depends on material, defect type, and accessibility. Leak testing is for detecting leaks, not internal defects.
- SNT-TC-1A
- A recommended practice for NDT personnel qualification and certification. Employers write their own practice; certification levels I, II, III with minimum training and experience hours.
- CP-189
- A standard that establishes minimum requirements for qualification and certification of NDT personnel. More prescriptive than SNT-TC-1A.
- Level III responsibilities
- Level III personnel approve procedures, interpret codes, and train/certify Level I and II personnel. They also must pass a basic, method, and practical exam.
- Certification examination
- Requires passing general, specific, and practical examinations. Passing score typically 80%. Records must be maintained by the employer.
- Recertification requirements
- SNT-TC-1A requires recertification every 5 years, including annual vision test (Jaeger 1 or equivalent) and documented continuing experience.
- Vision acuity test
- Annual vision test required for NDT personnel; must demonstrate near vision acuity of Jaeger 1 or equivalent, and color perception as applicable.
- Employer certification records
- Employer must maintain records of training, experience, examination results, and vision tests for each certified individual.
- Leak rate units conversion
- 1 std cm³/s = 0.001 mbar·L/s = 0.1 Pa·L/s = 0.001 Pa·m³/s. Standard conditions are 0°C and 1 atm.
- Molecular flow conductance
- For a long tube, C = (π d^3 / (12 L)) √(2π R T / M), where d is diameter, L is length, R is gas constant, T is temperature, M is molar mass.
- Background tracer gas
- Ambient concentration of tracer gas can cause false signals; must be minimized by ventilation or using a tracer with low atmospheric concentration.