CAMRT MRI Exam flashcards
CAMRT MRI Exam · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Larmor equation
- The resonance frequency of a nucleus equals the gyromagnetic ratio times the magnetic field strength; for hydrogen at 1.5 T it is about 63.9 MHz.
- Zone IV
- The magnet room itself, where the static field is present; access is physically restricted and only MR-safe equipment and screened personnel may enter.
- MR-Conditional
- A device labeling that means safe only under specified conditions such as field strength, spatial gradient, SAR limit, and coil type, not under all conditions.
- SAR
- Specific absorption rate, the RF power absorbed per unit mass in watts per kilogram; whole-body average is limited to 4 W/kg over 6 minutes in normal mode.
- Quench
- Loss of superconductivity causing rapid helium boil-off and oxygen displacement; evacuate the magnet room, ventilate, and call the service engineer.
- T1 weighting
- Produced by a short TR and short TE; tissues with short T1 such as fat appear bright and fluid appears dark.
- T2 weighting
- Produced by a long TR and long TE; tissues with long T2 such as fluid appear bright.
- Proton density
- Produced by a long TR and short TE; contrast depends mainly on the number of protons per unit volume with little T1 or T2 effect.
- FLAIR
- Fluid-attenuated inversion recovery; an inversion pulse with a long TI nulls CSF signal so periventricular lesions stand out on a T2-weighted image.
- STIR
- Short tau inversion recovery; an inversion pulse with a short TI nulls fat signal and is robust to field inhomogeneity, useful at low field.
- Chemical shift artifact
- A bright and dark rim at fat-water interfaces along the frequency-encode direction, caused by the roughly 3.5 ppm frequency difference between fat and water protons.
- Aliasing
- Wrap-around of anatomy outside the FOV into the opposite side of the image in the phase-encode direction; corrected by larger FOV or oversampling.
- Zipper artifact
- A bright line across the image caused by RF interference entering the room through a shield breach or external RF source.
- Susceptibility artifact
- Local field distortion near metal or air-tissue interfaces; worse with gradient echo, longer TE, and higher field, and reduced with spin echo and shorter TE.
- Gadolinium contrast
- A paramagnetic agent that shortens T1 and T2; typical extracellular dose is 0.1 mmol/kg and the effect is greatest on T1-weighted images.
- Nephrogenic systemic fibrosis
- A fibrosing disorder associated with gadolinium exposure in patients with severe renal impairment, more common with linear agents.
- Extravasation management
- Stop the injection, elevate the limb, apply a cold compress, document the volume and site, and arrange follow-up.
- Anaphylactoid reaction
- A severe contrast reaction with urticaria, bronchospasm, or hypotension; stop the injection, call for help, give oxygen, and administer epinephrine per protocol.
- Parallel imaging
- Undersampling phase-encode lines and using coil sensitivity information to reconstruct; it reduces scan time and echo train length at the cost of SNR.
- K-space center
- The central lines of k-space carry contrast information, while the periphery carries spatial detail; the filling order affects contrast and motion sensitivity.
- Echo-planar imaging
- A fast technique that fills all k-space lines after a single excitation; used for diffusion and functional imaging but sensitive to susceptibility distortion.
- Diffusion-weighted imaging
- Uses strong bipolar gradients; the b-value sets diffusion weighting and the ADC map removes T2 shine-through to show restricted diffusion.
- Conus medullaris level
- In adults the spinal cord typically terminates at L1-L2; below that level the cauda equina nerve roots are seen in the thecal sac.
- Couinaud segments
- The liver is divided into eight segments based on portal and hepatic vein anatomy, used to localize lesions for surgical planning.
- SBAR handoff
- Situation, Background, Assessment, Recommendation; a structured format for communicating with the radiologist, nurse, or anesthesia team.
- Closed-loop communication
- The sender states the message, the receiver repeats it back, and the sender confirms, reducing errors in high-stress situations.
- Just culture
- A framework that distinguishes human error from at-risk and reckless behavior, encouraging reporting without blame for honest mistakes.
- PDSA cycle
- Plan-Do-Study-Act; a small-scale test of a change with measurement before wider adoption in quality improvement.
- Peripheral nerve stimulation
- Induced by rapidly changing gradient fields at high dB/dt; the system enforces limits to keep stimulation below threshold.
- RF shield
- A copper enclosure around the magnet room that prevents external RF interference from entering and causing zipper or noise artifacts.
- SNR and NEX
- Signal-to-noise ratio increases with the square root of the number of acquisitions, so quadrupling NEX doubles SNR at four times the scan time.
- Two-identifier check
- Patient identification uses two identifiers such as name and date of birth, checked against the requisition before every study.