Larmor frequency

The rate at which nuclear spins precess in a magnetic field, equal to the gyromagnetic ratio multiplied by field strength. At 1.5 T, hydrogen protons precess at about 63.9 MHz.

Related: Gyromagnetic ratio, Resonance, Flip angle
Gyromagnetic ratio

A constant specific to each nucleus describing the ratio of its magnetic moment to its angular momentum. It determines the Larmor frequency at a given field strength; for hydrogen it is about 42.58 MHz/T.

Related: Larmor frequency, Resonance
Resonance

The condition in which radiofrequency energy matches the Larmor frequency of a spin system, allowing net absorption of energy and tipping of magnetization away from its equilibrium alignment.

Related: Larmor frequency, Flip angle, Radiofrequency pulse
Flip angle

The angle by which net magnetization is rotated away from the longitudinal axis by an RF pulse. It depends on pulse amplitude and duration; 90 degrees creates transverse magnetization and 180 degrees inverts it.

Related: Radiofrequency pulse, T1 relaxation, Spin echo
Radiofrequency pulse

A brief burst of electromagnetic energy at the Larmor frequency applied to excite spins. Its shape and duration determine the flip angle and the slice profile when combined with a gradient.

Related: Flip angle, Slice selection, Resonance
T1 relaxation

The recovery of longitudinal magnetization toward equilibrium after excitation, caused by energy transfer from spins to their surroundings. Tissues with short T1 recover quickly and appear bright on T1-weighted images.

Related: T2 relaxation, Repetition time, Inversion recovery
T2 relaxation

The decay of transverse magnetization caused by interactions between neighbouring spins, leading to loss of phase coherence. It is always shorter than or equal to T2 star and reflects true tissue properties.

Related: T2 star, Spin echo, Echo time
T2 star

The observed decay of transverse magnetization when both spin-spin interactions and magnetic field inhomogeneity contribute. It is shorter than true T2 and is the basis of gradient echo contrast.

Related: T2 relaxation, Gradient echo, Susceptibility artifact
Repetition time

The time between successive excitation pulses in a pulse sequence. Longer values allow more longitudinal recovery, giving stronger T1 weighting when combined with a short echo time.

Related: Echo time, T1 relaxation, Saturation
Echo time

The interval between excitation and the collection of the echo signal. Longer values increase T2 weighting; short values minimize T2 decay and favour proton density or T1 contrast.

Related: Repetition time, Spin echo, T2 relaxation
Spin echo

A pulse sequence in which a 180-degree refocusing pulse reverses dephasing from field inhomogeneity, producing an echo at a chosen time. It reduces susceptibility effects and provides true T2 weighting.

Related: Echo time, T2 relaxation, Gradient echo
Gradient echo

A sequence that refocuses transverse magnetization by reversing a magnetic field gradient rather than using a 180-degree pulse. It is fast, sensitive to field inhomogeneity, and allows short echo and repetition times.

Related: T2 star, Flip angle, Steady-state free precession
Inversion recovery

A preparation method in which a 180-degree pulse inverts longitudinal magnetization before excitation. The inversion time chosen determines fluid suppression, as in FLAIR and STIR sequences.

Related: Inversion time, T1 relaxation, Fluid-attenuated inversion recovery
Inversion time

The delay between the inverting 180-degree pulse and the excitation pulse in an inversion recovery sequence. It is selected so that a chosen tissue crosses zero magnetization and appears dark.

Related: Inversion recovery, Fluid-attenuated inversion recovery, Short tau inversion recovery
Slice selection

Spatial localization achieved by applying a magnetic field gradient during the RF pulse so that only spins in a narrow frequency band are excited. Slice thickness depends on gradient strength and pulse bandwidth.

Related: Radiofrequency pulse, Slice thickness, Frequency encoding
Slice thickness

The width of the excited imaging plane, determined by the RF pulse bandwidth divided by the slice-select gradient strength. Thinner slices improve through-plane resolution but reduce signal-to-noise.

Related: Slice selection, Signal-to-noise ratio, Slice gap
Frequency encoding

Spatial localization along one axis by applying a gradient during readout so that position maps to precessional frequency. A Fourier transform converts the received signal into a spatial profile.

Related: Phase encoding, K-space, Field of view
Phase encoding

Spatial localization by applying a gradient briefly before readout so that position maps to accumulated phase. Each repetition samples one line of k-space, so the number of phase steps sets scan time.

Related: Frequency encoding, K-space, Scan time
K-space

The raw data matrix in which MRI signals are stored before Fourier transformation. Its centre contains low-spatial-frequency contrast information and its periphery contains high-spatial-frequency edge detail.

Related: Fourier transform, Phase encoding, Parallel imaging
Fourier transform

The mathematical operation that converts k-space data from the frequency and phase domain into a displayable image. It is the final reconstruction step in conventional two-dimensional imaging.

Related: K-space, Frequency encoding
Signal-to-noise ratio

The ratio of true signal amplitude to random background noise in an image. It rises with voxel volume, field strength, and number of signal averages, and falls with higher bandwidth.

Related: Slice thickness, Number of signal averages, Receive bandwidth
Number of signal averages

The number of times each k-space line is acquired and averaged. Signal-to-noise ratio improves with the square root of the number of averages, while scan time increases proportionally.

Related: Signal-to-noise ratio, Scan time, K-space
Receive bandwidth

The range of frequencies sampled during readout. A wider bandwidth reduces chemical shift and susceptibility artifacts and shortens minimum echo time, but admits more noise and lowers signal-to-noise ratio.

Related: Signal-to-noise ratio, Chemical shift artifact, Echo time
Chemical shift artifact

Misregistration at fat-water interfaces caused by the small difference in precessional frequency between fat and water protons. It appears as bright and dark bands along the frequency-encoding direction.

Related: Receive bandwidth, Fat suppression, Frequency encoding
Susceptibility artifact

Local signal loss and geometric distortion near materials that alter magnetic field homogeneity, such as metal implants or air-tissue boundaries. It is worse with gradient echo sequences and longer echo times.

Related: Gradient echo, T2 star, Echo time
Aliasing

Wrapping of anatomy from outside the field of view into the opposite side of the image when the sampled object exceeds the field of view. It occurs along the phase-encoding direction and is corrected by oversampling or saturation bands.

Related: Field of view, Phase encoding, Saturation
Field of view

The physical dimensions of the imaged region. It relates to pixel size through the acquisition matrix; a smaller field of view improves spatial resolution but increases aliasing risk and lowers signal-to-noise.

Related: Aliasing, Frequency encoding, Spatial resolution
Spatial resolution

The ability to distinguish small adjacent structures, set by pixel dimensions and slice thickness. Finer resolution requires more encoding steps, which lengthens scan time and reduces signal-to-noise ratio.

Related: Field of view, Slice thickness, Signal-to-noise ratio
Parallel imaging

A reconstruction technique that uses the differing spatial sensitivity of multiple receiver coils to skip phase-encoding lines. It shortens scan time or reduces distortion, at the cost of lower signal-to-noise.

Related: Phase encoding, K-space, Signal-to-noise ratio
Fat suppression

Techniques that reduce signal from adipose tissue, including chemical shift saturation, inversion recovery such as STIR, and Dixon methods. It improves conspicuity of edema and contrast enhancement.

Related: Short tau inversion recovery, Chemical shift artifact, Inversion time
Fluid-attenuated inversion recovery

A T2-weighted sequence using a long inversion time to null cerebrospinal fluid signal. It improves detection of periventricular and cortical lesions that would otherwise be obscured by bright fluid.

Related: Inversion time, Inversion recovery, T2 relaxation
Short tau inversion recovery

An inversion recovery sequence with a short inversion time chosen to null fat signal while preserving fluid brightness. It is highly sensitive to edema and useful when field homogeneity is poor.

Related: Inversion time, Fat suppression, Inversion recovery
Gadolinium contrast agent

A paramagnetic chelate that shortens T1 relaxation of nearby water protons, increasing signal on T1-weighted images. It is used to detect blood-brain barrier breakdown and characterize lesions.

Related: T1 relaxation, Nephrogenic systemic fibrosis, Contrast enhancement
Nephrogenic systemic fibrosis

A rare fibrosing disorder associated with exposure to older linear gadolinium agents in patients with severe renal impairment. Risk assessment of renal function guides contrast use and agent selection.

Related: Gadolinium contrast agent, Contrast enhancement
Contrast enhancement

Increased signal on post-contrast images where gadolinium accumulates, reflecting regional perfusion and blood-brain barrier integrity. Comparison with pre-contrast images confirms true enhancement rather than intrinsic T1 shortening.

Related: Gadolinium contrast agent, T1 relaxation
Specific absorption rate

The rate at which radiofrequency energy is deposited in tissue, expressed in watts per kilogram. It is monitored because excessive heating risks burns, and it rises with field strength and with fast, high-flip-angle sequences.

Related: Radiofrequency pulse, Flip angle, Quench
Quench

The sudden loss of superconductivity in the magnet coil, releasing cryogen gas and collapsing the magnetic field. It requires immediate evacuation of the room and activation of emergency procedures.

Related: Superconducting magnet, Specific absorption rate
Superconducting magnet

The main magnet design in clinical MRI, in which a niobium-titanium coil cooled by liquid helium carries current with no resistance, producing a strong, stable, and homogeneous static field.

Related: Quench, Shielding, Fringe field
Fringe field

The stray static magnetic field extending beyond the scanner room. It is contained by passive or active shielding, and its extent defines the controlled access area and pacemaker exclusion zones.

Related: Shielding, Superconducting magnet, MR safety zone
Shielding

Design measures that confine the magnetic fringe field or block external radiofrequency interference. Passive shielding uses steel; active shielding uses opposing coils; RF shielding uses a copper-lined room.

Related: Fringe field, Radiofrequency interference, Superconducting magnet
MR safety zone

One of four designated areas around an MRI suite, graded from freely accessible to the scanner room itself. Zones three and four are restricted to screened personnel and patients under direct supervision.

Related: Fringe field, Projectile effect
Projectile effect

The forceful attraction of ferromagnetic objects into the magnet bore, which can injure patients or staff and damage equipment. Prevention relies on screening, signage, and strict control of zone four.

Related: MR safety zone, Fringe field
Radiofrequency interference

External electromagnetic noise that degrades MR images when it enters the scan room. It is prevented by a continuous RF shield, filtered penetrations, and careful door and waveguide maintenance.

Related: Shielding, Signal-to-noise ratio