Quick reference

Radiological Technology

  • mAs controls photon quantity and receptor dose; kVp controls beam penetrability and subject contrast. A 15% kVp increase roughly doubles receptor dose at constant mAs.

  • Inverse square law: I1/I2 = (d2/d1)^2, so doubling the distance from a point source reduces intensity to one quarter.

  • Magnification = SID/SOD, where SOD = SID minus OID. Reducing object-to-image distance reduces magnification and improves recorded detail.

  • Photoelectric absorption varies roughly with Z^3/E^3 and dominates at low kVp and with high-atomic-number contrast media; Compton scatter depends on electron density and dominates at higher kVp.

  • Grid ratio = lead strip height divided by interspace width. Higher ratios remove more scatter and improve contrast but require more mAs; grid cut-off follows lateral decentring or tube angulation.

  • Typical half-value layer in general radiography is about 2.5 mm Al at 70 kVp; minimum total filtration is 2.5 mm Al equivalent.

  • CT pitch = table travel per gantry rotation divided by total beam collimation. DLP = CTDIvol multiplied by scan length; effective dose is estimated as DLP times a region-specific k-factor.

  • Display mapping: window level sets the CT number shown as mid-grey (brightness); window width sets the range mapped to the grey scale (contrast). Narrowing width raises contrast but lowers latitude.

  • Occupational dose limit is 20 mSv per year averaged over five years, with no more than 50 mSv in any single year; the public limit is 1 mSv per year.

Quick reference

Magnetic Resonance

  • Larmor equation: precessional frequency = gyromagnetic ratio times field strength. For hydrogen-1, 42.58 MHz/T gives 63.9 MHz at 1.5 T and 127.7 MHz at 3 T.

  • TR governs T1 weighting and TE governs T2 weighting: short TR/short TE = T1-weighted, long TR/long TE = T2-weighted, long TR/short TE = proton density.

  • T2* combines true T2 dephasing with static field inhomogeneity. A 180-degree refocusing pulse (spin echo) removes the inhomogeneity component; gradient echo does not.

  • IEC normal operating mode limits whole-body SAR to 2 W/kg averaged over 6 minutes; first-level controlled mode permits 4 W/kg averaged over 15 minutes.

  • Gadolinium chelates shorten both T1 and T2; at low concentration T1 shortening dominates, giving high signal on T1-weighted images. Avoid in severe renal impairment because of nephrogenic systemic fibrosis risk.

  • STIR suppresses fat by inversion timing and is insensitive to field inhomogeneity; spectral (chemical) fat saturation requires good shimming and fails at low field strength.

  • Pixel size = FOV divided by matrix; SNR scales with voxel volume, so doubling the matrix in both directions at fixed FOV cuts SNR by about a factor of four.

  • K-space centre lines encode image contrast and signal-to-noise; peripheral lines encode edges and fine detail. Half-Fourier and echo-planar trajectories exploit this ordering.

  • Chemical shift displacement in hertz equals about 3.5 ppm times 42.58 times field strength, approximately 220 Hz at 1.5 T; the 5-gauss line defines the controlled access area.

Quick reference

Nuclear Medicine

  • Tc-99m has a 6.02-hour half-life and a 140 keV gamma emission and is eluted from a Mo-99/Tc-99m generator (Mo-99 half-life 66 hours); test eluate for Mo-99 breakthrough and aluminium.

  • Decay law: activity = initial activity times exp(-0.693 t / half-life); activity halves with every elapsed half-life.

  • Effective half-life combines physical and biological clearance: 1/Te = 1/Tphysical + 1/Tbiological.

  • PET positron annihilation produces two 511 keV photons emitted about 180 degrees apart; coincidence timing windows are typically a few nanoseconds and random events rise with the square of activity.

  • Counting follows Poisson statistics: standard deviation of a count equals the square root of the count, so quadrupling counts halves the relative uncertainty.

  • A pulse-height analyser window of about plus or minus 10% around the photopeak accepts primary photons while rejecting most scattered events.

  • Common Tc-99m agents: MDP for bone, DTPA or MAG3 for renal function, sulfur colloid for liver and spleen, pertechnetate for thyroid and Meckel diverticulum imaging; I-123 for thyroid imaging and I-131 for therapy.

  • Dose calibrator quality control includes daily constancy measurement, accuracy within plus or minus 10%, and checks that readings are independent of sample volume and geometry.

Quick reference

Radiation Therapy

  • Absorbed dose: 1 Gy = 1 J/kg = 100 rad. Equivalent and effective dose use the sievert, where 1 Sv = 100 rem.

  • BED = n x d x (1 + d/(alpha/beta)), where n is fraction number and d is dose per fraction. Alpha/beta is about 10 Gy for tumours and early effects, about 3 Gy for late effects, and about 1.5 Gy for prostate.

  • Typical curative regimen: 2 Gy per fraction over 30 to 35 fractions, total 60 to 70 Gy; palliative schedules use 8 to 30 Gy in 1 to 10 fractions.

  • Percentage depth dose increases with beam energy, field size and source-to-surface distance; TMR and TPR are independent of SSD and are used for isocentric (SAD) setups.

  • ICRU volumes: GTV is visible disease, CTV adds microscopic spread, ITV adds internal motion, PTV adds set-up margin; PRV is the planning margin applied to an organ at risk.

  • Megavoltage photon beams show skin sparing, with maximum dose at about 1.5 cm for 6 MV and about 3 cm for 18 MV; bolus restores surface dose.

  • Electron beam therapeutic depth in centimetres is approximated by energy in MeV divided by 3 to 4, with rapid dose fall-off beyond that depth.

  • IMRT and VMAT use inverse planning against dose-volume objectives; VMAT modulates the multileaf collimator during continuous gantry rotation, shortening treatment time relative to fixed-field IMRT.

  • Brachytherapy dose falls roughly as the inverse square of distance from the source; the TG-43 formalism uses air-kerma strength, dose-rate constant and anisotropy functions.

Quick reference

Radiological Technology

  • mAs sets photon quantity and receptor dose, kVp sets penetrability.

  • Inverse square law: intensity falls with the square of distance.

  • Grid ratio compares strip height to interspace width.

  • Photoelectric absorption rises with atomic number cubed.

  • Magnification equals SID divided by SOD.

  • CT pitch equals table travel divided by collimation.

  • Window level controls brightness; width controls contrast.

  • Occupational dose limit averages 20 mSv per year.

Quick reference

Magnetic Resonance

  • Larmor frequency equals gyromagnetic ratio times field strength.

  • TR and TE determine T1, T2 or proton density weighting.

  • T2 star decay includes field inhomogeneity dephasing.

  • Ernst angle maximises signal for a given TR and T1.

  • STIR suppresses fat despite field inhomogeneity.

  • SAR rises with field strength squared and flip angle.

  • Gadolinium shortens T1 and brightens T1-weighted images.

  • K-space centre lines govern image contrast.

Quick reference

Nuclear Medicine

  • Tc-99m has a 6-hour half-life and 140 keV gamma.

  • Effective half-life combines physical and biological clearance.

  • Annihilation yields two 511 keV photons at 180 degrees.

  • Energy windows are set near the photopeak.

  • Random coincidences increase with the square of activity.

  • Counting precision improves with the square root of counts.

  • Generators are eluted and tested for breakthrough.

  • FDG uptake reflects regional glucose metabolism.

Quick reference

Radiation Therapy

  • Gray equals one joule per kilogram.

  • BED compares schedules using alpha-beta.

  • PTV includes set-up and motion margins.

  • PDD rises with energy, field size and SSD.

  • Alpha-beta near 10 gray marks early-responding tissue.

  • Bolus builds up dose on the skin surface.

  • Inverse planning optimises fluence maps.

  • Brachytherapy dose falls as inverse square.