- Cardiac output formula
- Cardiac output equals heart rate multiplied by stroke volume; a typical resting adult value is about five liters per minute.
- Sinoatrial node intrinsic rate
- The SA node, the heart's primary pacemaker, normally fires at 60 to 100 beats per minute; the AV node fires at 40 to 60.
- ECG waves and intervals
- The P wave is atrial depolarization, the QRS complex is ventricular depolarization, and the T wave is ventricular repolarization; a normal PR interval is 0.12 to 0.20 second.
- Mean arterial pressure formula
- MAP approximately equals diastolic pressure plus one third of the pulse pressure, which is systolic minus diastolic; normal MAP is roughly 70 to 100 mmHg.
- Nephron filtration rate
- Each kidney contains about one million nephrons; combined glomerular filtration rate is roughly 125 mL per minute, regulating fluid, electrolytes, and acid-base balance.
- Autonomic nervous system divisions
- The sympathetic division mediates fight-or-flight responses such as tachycardia, while the parasympathetic division, largely via the vagus nerve, slows the heart and promotes digestion.
- Ohm's law
- Voltage equals current multiplied by resistance (V = IR); rearranged, current equals voltage divided by resistance and resistance equals voltage divided by current.
- Electrical power formulas
- Electrical power equals voltage times current (P = VI), and equivalently P = I squared times R or P = V squared divided by R.
- Series versus parallel resistance
- In series, resistances simply add and current is common; in parallel, conductances add, so total resistance is less than the smallest branch and voltage is common.
- Capacitive and inductive reactance
- Capacitive reactance is one divided by two pi f C and decreases as frequency rises; inductive reactance is two pi f L and increases with frequency.
- Root mean square voltage
- For a sine wave, RMS voltage equals peak voltage divided by the square root of two, about 0.707 times peak; 120 V mains corresponds to about 170 V peak.
- Patient leakage current limits
- Under IEC 60601, patient leakage current must not exceed 100 microamps in normal condition or 500 microamps under single-fault condition for patient-connected equipment.
- HL7 version 2 message types
- HL7 v2 messages include ADT for admissions and demographics, ORM for orders, and ORU for observation results; segments are separated by carriage returns and fields by pipe characters.
- DICOM standard purpose
- DICOM defines file formats and network services for medical images, letting modalities, workstations, and PACS store, query, and retrieve studies across different vendors.
- HIPAA security safeguards
- The HIPAA Security Rule requires administrative, physical, and technical safeguards for electronic protected health information, including access control, audit logs, and encryption of transmitted data.
- DHCP versus DNS
- DHCP automatically assigns IP addresses and related network settings to devices, while DNS resolves human-readable host names into numeric IP addresses.
- TCP versus UDP
- TCP is connection-oriented and guarantees ordered delivery with acknowledgments, while UDP is connectionless and does not guarantee delivery, making it suitable for real-time streams.
- Pulse oximeter wavelengths
- Pulse oximeters compare light absorption at approximately 660 nanometers (red) and 940 nanometers (infrared) to compute the ratio of oxygenated to deoxygenated hemoglobin.
- End-tidal CO2 normal range
- Capnography measures exhaled carbon dioxide; end-tidal CO2 in a healthy adult is 35 to 45 mmHg, and a sudden drop may indicate airway disconnection.
- Lung-protective tidal volume
- Ventilators are typically set to deliver 6 to 8 milliliters of tidal volume per kilogram of predicted body weight to reduce ventilator-induced lung injury.
- Defibrillation energy settings
- Biphasic defibrillators typically deliver 120 to 200 joules per shock, while older monophasic devices may use up to 360 joules.
- Electrosurgical return electrode
- Monopolar electrosurgery requires a large-area return electrode on the patient, and return electrode monitoring circuitry detects pad detachment to prevent burns.
- CT Hounsfield unit scale
- Hounsfield units express CT attenuation: water is zero, air is about minus 1000, and dense bone is near plus 1000, allowing tissue contrast to be windowed.
- Half-split troubleshooting method
- The half-split method divides a system into sections and tests the midpoint first, reducing the number of checks needed to isolate a faulty stage.
- Protective earth resistance limit
- Grounding resistance measured from the plug's ground pin to exposed chassis metal should not exceed 0.2 ohm on patient-care equipment.
- Components checked first
- Power cords, fuses, batteries, connectors, and transducers fail most often, so they are verified before deeper circuit-level diagnosis begins.
- Five Whys technique
- Repeatedly asking why a failure occurred, typically five times, drives the analysis from the observed symptom toward the underlying process root cause.
- CMMS work order record
- A computerized maintenance management system records each repair's complaint, findings, parts used, and labor, providing device history needed for regulatory surveys.
- Fire extinguisher classes
- Class A is ordinary combustibles, B flammable liquids, C energized electrical equipment, and D combustible metals; water must never be used on a Class C fire.
- PASS fire extinguisher use
- PASS stands for Pull the pin, Aim at the base of the fire, Squeeze the handle, and Sweep the nozzle side to side across the fire.
- Medical gas cylinder colors
- In the United States, oxygen cylinders are green, nitrous oxide blue, nitrogen black, carbon dioxide gray, and medical air yellow.
- ALARA radiation principle
- ALARA means keeping radiation exposure as low as reasonably achievable using time, distance, and shielding; intensity falls with the square of distance.