CBET study plan
CBET · Give each week a clear focus.
8 weeks · About 7 hours per week
Healthcare Technology and Function
Categorize at least 12 medical device families (physiologic monitors, infusion pumps, ventilators, anesthesia machines, electrosurgical units, defibrillators, dialysis machines, lab analyzers, imaging systems) by the physiologic parameter measured or therapy delivered.
Reconstruct from memory the block-diagram signal path (transducer, amplifier, filter, ADC, processor, display) for three monitoring device types.
Compare the operating principles of ventilators, anesthesia machines, and infusion pumps in a written comparison table.
Identify the dominant patient-safety hazard and the typical failure mode for each device family you list.
Your checkpoint
A one-page device-function matrix covering 12 device families with columns for operating principle, measured parameter, primary hazard, and typical failure mode.
Healthcare Technology and Function & Healthcare Technology Problem Solving
Trace the full equipment lifecycle (specification, incoming inspection, preventive maintenance, calibration, repair, decommission) for two device types from memory.
Execute and document a simulated PM/calibration procedure on two device categories, naming the required test equipment and the tolerance criteria used.
Explain signal acquisition for ECG, SpO2, NIBP, and capnography, and list the common artifact sources for each.
Construct a troubleshooting decision tree for one device family using a structured symptom-to-subsystem method.
Your checkpoint
Two completed PM checklists (with test equipment and tolerances recorded) plus one troubleshooting decision tree.
Healthcare Technology Problem Solving
Apply a structured six-step troubleshooting model (verify, gather information, hypothesize, test, resolve, document) to eight written fault scenarios.
Distinguish operator error, environmental cause, and hardware failure in each scenario before proposing a repair.
Select the appropriate test instrument (DMM, oscilloscope, patient simulator/analyzer, electrical safety analyzer) for a described fault and justify the choice.
Diagnose faults confined to power-supply, sensor, and interconnect subsystems using measured values rather than substitution.
Your checkpoint
Written worked solutions for eight troubleshooting scenarios, each ending with a stated root cause, corrective action, and verification step.
Healthcare Technology Problem Solving & Healthcare Information Technology
Solve at least 15 timed mixed practice scenarios in problem solving and score 80% or better.
Sequence a complete electrical safety test (ground continuity, chassis and patient leakage current, insulation) and interpret results against pass/fail limits.
Describe how a work order flows through the CMMS from open to close, including how findings are documented for traceability.
Justify, for a described fault, whether a device is removed from service, repaired in place, or returned to service.
Your checkpoint
A score report from a 25-item Domain 5 practice set plus an error log that tags every miss by failure category and root cause.
Healthcare Information Technology
Diagram the end-to-end data path from a bedside device through an interface/connectivity engine into the EHR, including wired and wireless segments.
Differentiate HL7 v2 message types (ADT, ORU, ORM) and state what clinical content each carries.
Explain the purpose of DICOM and describe how a modality worklist is populated and used.
Describe medical-device cybersecurity controls (asset inventory, network segmentation, patching, access control) and where each applies.
Assign ownership (IT, BMET, vendor) for eight common device-integration problems and defend each assignment.
Your checkpoint
A labeled network and data-flow diagram plus a two-column ownership table covering eight integration issues.
Healthcare Information Technology & Public (employee, patient, visitor) Safety in the Healthcare Facility
Explain the service-work implications of HIPAA privacy and security rules, including PHI on device storage, disk/media disposal, and remote vendor access.
Apply the hierarchy of controls to five healthcare facility hazards and rank the resulting mitigations.
Describe fire response and evacuation procedures, electrical safety requirements, and medical-gas/oxygen-enriched-atmosphere precautions.
Explain how recalls and hazard alerts are received, triaged, and closed out in a hospital equipment program.
Your checkpoint
A one-page safety and compliance quick-reference card covering HIPAA handling, fire response, electrical safety, and medical-gas hazards.
Public (employee, patient, visitor) Safety in the Healthcare Facility & Anatomy & Physiology & Fundamentals of Electricity and Electronics
Label the major body systems and state, for each, at least one device you service that interfaces with it.
Compute Ohm's law, power, series and parallel resistance, and decibel conversions across 10 practice problems without reference.
Identify resistors, capacitors, inductors, diodes, transistors, and op-amps from schematic symbols and state each component's function in a device circuit.
Select the correct precaution (ESD control, lockout/tagout, PPE, oxygen-safe tools) for a described task.
Your checkpoint
A 20-item combined quiz on anatomy/physiology, electronics fundamentals, and facility safety scored at 80% or above.
Anatomy & Physiology & Fundamentals of Electricity and Electronics & Healthcare Information Technology & Healthcare Technology and Function & Healthcare Technology Problem Solving & Public (employee, patient, visitor) Safety in the Healthcare Facility
Complete two full-length timed practice exams and score 80% or better on each.
Re-study the 10 weakest objectives identified in your cumulative error log.
Reproduce the device-function matrix, key formulas, and safety limits from memory.
Condense all notes into a single formula and concept sheet for final review.
Your checkpoint
Final mock-exam score reports with a dated error log showing every weak objective closed out, plus the completed formula and concept sheet.
Milestones
Week 2: Device-function matrix covers 12 families and two PM checklists are complete with test equipment and tolerances documented.
Week 4: Domain 5 practice set of 25 items is scored at 80% or better and the error log tags every miss by failure category.
Week 6: Network/data-flow diagram and ownership table are finished, and the Domain 3 self-quiz is scored at 75% or better.
Week 8: Two full-length timed practice exams are both scored at 80% or better with no repeated error in the log.